Medical Devices Rules, 2017
Medical Devices Rules, 2017[1]
[31st
January, 2017]
Whereas the draft of the
Medical Devices Rules, 2016 was published, as required under sub-section (1) of
Section 12 and sub-section (1) of Section 33 of the Drugs and Cosmetics Act,
1940 (23 of 1940), in the Gazette of India, Extraordinary, Part II, Section 3,
sub-section (i), vide notification number G.S.R. 983(E), dated the 17th
October, 2016, by the Central Government, after consultation with the Drugs
Technical Advisory Board, inviting objections and suggestions from all persons
likely to be affected thereby, before the expiry of a period of thirty days
from the date on which copies of the said Gazette containing the said
notification were made available to the public;
And whereas, copies of the
Gazette containing the said notification were made available to the public on
the 17th October, 2016;
And whereas, all objections
and suggestions received in response to the said draft notification have been
duly considered by the Central Government;
Now, therefore, in exercise
of the powers conferred by Section 12 and Section 33 of the Drugs and Cosmetics
Act, 1940 (23 of 1940), the Central Government, after consultation with the
Drugs Technical Advisory Board, hereby makes the following rules, namely—
Chapter I PRELIMINARY
Rule - 1. Short title and commencement.
(1)     These rules may be called
the Medical Devices Rules, 2017.
(2)     These rules shall, unless
specified otherwise, come into force with effect from 1st day of January, 2018.
Rule - 2. Application.
These rules shall be
applicable in respect of,—
(i)       substances used for in
vitro diagnosis and surgical dressings, surgical bandages, surgical staples,
surgical sutures, ligatures, blood and blood component collection bag with or
without anticoagulant covered under sub-clause (i);
(ii)      substances including
mechanical contraceptives (condoms, intrauterine devices, tubal rings),
disinfectants and insecticides notified under sub-clause (ii); and
(iii)     devices notified from time
to time under sub-clause (iv),
of clause (b) of Section 3
of the Drugs and Cosmetics Act, 1940 (23 of 1940);
Rule - 3. Definitions.
In these rules, unless the
context otherwise requires,—
(a)      “academic clinical study”
means a clinical study conducted for academic purpose on a medical device for
the approved or a new intended use, new material of construction, new improved
design or new population;
(b)      “Act” means the Drugs and
Cosmetics Act, 1940 (23 of 1940);
(c)      “active diagnostic medical
device” means any active medical device used, whether alone or in combination
with other medical devices, to supply information for detecting, diagnosing or
monitoring, or to provide support in the treatment of, any physiological
condition, state of health, illness or congenital deformity;
(d)      “active medical device”
means a medical device, the operation of which depends on a source of
electrical energy or any other source of energy other than the energy generated
by human or animal body or gravity;
(e)      “active therapeutic medical
device” means any active medical device used, whether alone or in combination
with any other medical device, to support, modify, replace or restore
biological functions or structures, with a view to the treatment or alleviation
of any illness, injury or handicap;
(f)       “authorised agent” means a
person including any firm or organisation who has been appointed by an overseas
manufacturer through a power of attorney to undertake import of medical device
in India;
(g)      “body orifice” means any
natural opening in a human body including the external surface of any eyeball,
or any permanent artificial opening, such as a stoma or permanent tracheotomy;
(h)     “Central Licensing
Authority” means the Drugs Controller General of India appointed by the Central
Government;
(i)       “central medical devices
testing laboratory” means a medical devices laboratory established or
designated by the Central Government under Rule 19 and shall be deemed to be a
Central Drug Laboratory established for the purpose of Section 6 of the Act;
(j)       “change in the constitution
of a licensee” in relation to,—
(i)       a firm means change from
proprietorship to partnership including Limited Liability Partnership or vice
versa;
(ii)      a company means—
(A)     its conversion from a
private to a public company, or from a public to a private company; or
(B)     any change in the ownership
of shares of more than fifty per cent of the voting capital in the body
corporate or in case of a body corporate not having a share capital, any change
in its membership; and where the managing agent, being a body corporate is a
subsidiary of another body corporate, includes a change in the constitution of
that other body corporate within the meaning of this clause;
(k)      “clinical evidence” means,
in relation to,—
(i)       an in vitro diagnostic
medical device, is all the information derived from specimen collected from
human that supports the scientific validity and performance for its intended
use;
(ii)      a medical device, the
clinical data and the clinical evaluation report that supports the scientific
validity and performance for its intended use;
(l)       “clinical investigation”
means the systematic study of an investigational medical device in or on human
participants to assess its safety, performance or effectiveness;
(m)    “clinical investigation
plan” means a document which contains the information about the rationale, aims
and objective, design and the proposed analysis, conduct, methodology including
performance, management, adverse event, withdrawal and statistical
consideration and record keeping pertaining to clinical investigation;
(n)     “clinical performance
evaluation” means the systematic performance study of a new in vitro diagnostic
medical device on a specimen collected from human participants to assess its
performance;
(o)      “clinical research
organisation” means any entity to whom a sponsor may transfer or delegate one
or more of its functions and duties regarding conduct of clinical investigation
or clinical performance evaluation;
(p)      “conformity assessment”
means the systematic examination of evidence generated and procedures
undertaken, by the manufacturer to determine that a medical device is safe and
performs as intended by the manufacturer and therefore conforms to the
essential principles of safety and performance for medical devices;
(q)      “controlling officer” means
the officer designated under Rule 10;
(r)      “custom made medical
device” means a medical device made specifically in accordance with a written
prescription of a registered medical practitioner, specialised in the relevant
area, under his responsibility for the sole use of a particular patient, but
does not include a mass production of such device;
(s)      “Ethics Committee” means
the committee referred to in Rule 50;
(t)       “Form” means forms
specified in Appendix to these rules;
(u)     “Good Clinical Practices
Guidelines” means Good Clinical Practices Guidelines issued by Central Drugs
Standards Control Organisation, Directorate General of Health Services,
Ministry of Health and Family Welfare, Government of India;
(v)      “intended use” means the
use for which the medical device is intended according to the data supplied by
the manufacturer on the labelling or in the document containing instructions
for use [2][or
electronic instructions for use] of such device or in promotional material
relating to such device, which is as per approval obtained from the Central
Licensing Authority;
(w)     “invasive device” means a
device which, in whole or part, penetrates inside the body, either through a
body orifice or through the surface of the body;
(x)      “investigational medical
device” in relation to a medical device, other than in vitro diagnostic medical
device, means a medical device specified in clause (zb),—
(i)       which does not have its
predicate device as defined in clause (zm); or
(ii)      which is licensed under
sub-rule (4) or sub-rule (6) of Rule 20, sub-rule (1) of Rule 25, or sub-rule
(1) of Rule 36 and claims for new intended use or new population or new
material or major design change;
and is being assessed for
safety or performance or effectiveness in a clinical investigation.
(y)      “licence” means a licence
granted by the State Licensing Authority or the Central Licensing Authority in
Form MD-5, Form MD-6, Form MD-9, Form MD-10, Form MD-15, Form MD-17 or Form
MD-19 as the case may be;
(z)      “loan licence” means a licence
issued for manufacturing a medical device by the State Licensing Authority or
the Central Licensing Authority, as the case may be, to a person who intends to
utilise the manufacturing site of other licensee for manufacturing the same
medical device as manufactured by the licensee at that site;
(za)
 “long-term use” means intended continuous
use of a medical device for more than thirty days;
(zb)
 “medical device” means—
(A)     substances used for in
vitro diagnosis and surgical dressings, surgical bandages, surgical staples,
surgical sutures, ligatures, blood and blood component collection bag with or
without anticoagulant covered under sub-clause (i),
(B)     substances including
mechanical contraceptives (condoms, intrauterine devices, tubal rings),
disinfectants and insecticides notified in the Official Gazette under
sub-clause (ii),
(C)     devices notified from time
to time under sub-clause (iv),
of clause (b) of Section 3
of the Act;
Explanation.—For the
purpose of these rules, substances used for in vitro diagnosis shall be
referred as in vitro diagnostic medical device.
(zc)
 “medical device grouping” means a set of
devices having same or similar intended uses or commonality of technology
allowing them to be classified in a group not reflecting specific characteristics;
(zd)
 “Medical Device Officer” means an officer
appointed or designated by the Central Government or the State Government, as
the case may be, under sub-rule (2) of Rule 18;
(ze)
 “medical devices testing laboratory”
means any institute, organisation registered under sub-rule (3) of Rule 83 for
carrying out testing or evaluation of any medical device on behalf of a
licensee for manufacture for sale;
(zf)
  “Medical Device Testing Officer” means
an officer appointed or designated by the Central Government under sub-rule (1)
of Rule 18;
(zg)
 “near patient testing” means any
investigation carried out in a clinical setting or at the patient's home for
which the result is available without reference to a laboratory and rapidly
enough to affect immediate patient management;
(zh)
 “new in vitro diagnostic medical device”
means any medical device as referred to in sub-clause (A) of clause (zb) used
for in vitro diagnosis that has not been approved for manufacture for sale or
for import by the Central Licensing Authority and is being tested to establish
its performance for relevant analyte or other parameter related thereto
including details of technology and procedure required;
(zi)
  “notified” means notified in the
Official Gazette by the Central Government;
(zj)
  “Notified Body” means a body corporate
or other legal entity, registered under Rule 13 as a body competent to carry
out the audit of manufacturing site, assessment, and verification of specified
category of medical devices for establishing conformity with standards;
(zk)
 “performance evaluation” in relation to
in vitro diagnostic medical device means any systematic investigation by which
data is assessed and analysed to establish or verify performance of the in
vitro diagnostic medical device for its intended use;
(zl)
  “Post Marketing Surveillance” means
systematic process to collect and analyse information gained from medical
device that have been placed in the market;
(zm)
“predicate device” means a device, first time and first of its kind, approved for
manufacture for sale or for import by the Central Licensing Authority and has
the similar intended use, material of construction, and design characteristics
as the device which is proposed for licence in India;
(zn)
 “Quality Management System” means requirements
for manufacturing of medical devices as specified in the Fifth Schedule;
(zo)
 “reagent” means a chemical, biological or
immunological component, solution or preparation intended by the manufacturer
to be used as in vitro diagnostic medical device;
(zp)
 “recall” means any action taken by its
manufacturer or authorised agent or supplier to remove the medical device from
the market or to retrieve the medical device from any person to whom it has
been supplied, because the medical device,—
(a)      is hazardous to health; or
(b)      fails to conform to any
claim made by its manufacturer relating to its quality, safety or efficacy; or
(c)      does not meet the
requirements of the Act and these rules;
(zq)
 “serious adverse event” means an untoward
medical occurrence that leads to,—
(i)       a death; or
(ii)      a serious deterioration in
the health of the subject that either—
(A)     resulted in a life
threatening illness or injury; or
(B)     resulted in a permanent
impairment of a body structure or a body function; or
(C)     required in-patient hospitalisation
or prolongation of existing hospitalisation; or
(D)     resulted in medical or
surgical intervention to prevent life threatening illness or injury or
permanent impairment to a body structure or a body function; or
(iii)     foetal distress, foetal
death or a congenital abnormality or birth defect;
(zr)
 “short-term use” means intended
continuous use of a medical device for not less than sixty minutes but not more
than thirty days;
(zs)
 “specimen receptacle” means a device,
whether vacuum type or not, specifically intended by its manufacturer for the
primary containment of specimens derived from human or animal body;
(zt)
 “sponsor” includes a person, an
investigator, a company or an institution or an organisation responsible for
the initiation and management of a clinical investigation or clinical
performance evaluation in India;
(zu)
 “State Licensing Authority” means the
authority designated by the State Government under sub-rule (2) of Rule 8;
(zv)
 “transient use” means a device intended
for continuous use for less than sixty minutes;
(zw)
 “transmissible agent”, for the purpose of
classification of in vitro diagnostic medical device, means an agent capable of
being transmitted to a person, which causes communicable, infectious or
contagious disease;
(zx)
 words and expressions used but not
defined in these rules, shall have the meanings respectively assigned to them
in the Act and the Drugs and Cosmetics Rules, 1945.
Chapter II REGULATION OF
MEDICAL DEVICE
Rule - 4. Classification of medical
devices.
(1)     Medical devices other than
in vitro diagnostic medical devices shall be classified on the basis of
parameters specified in Part I of the First Schedule, in the following classes,
namely—
(i)       low risk — Class A;
(ii)      low moderate risk — Class
B;
(iii)     moderate high risk — Class
C;
(iv)    high risk — Class D.
(2)     In vitro diagnostic medical
devices shall be classified on the basis of parameters specified in Part II of
the First Schedule, in the following classes, namely—
(i)       low risk — Class A;
(ii)      low moderate risk — Class
B;
(iii)     moderate high risk — Class
C;
(iv)    high risk — Class D.
(3)     The Central Licensing
Authority shall, classify medical devices referred to in Rule 2, based on the
intended use of the device and other parameters specified in the First
Schedule.
(4)     Based on the classification
referred to in sub-rule (3), class wise list of medical devices shall be
published on the website of the Central Drugs Standard Control Organisation:
Provided that the Central
Licensing Authority may, from time to time, make additions or deletions in such
list of medical devices or modify the class of any medical device.
Rule - 5. Medical device grouping.
Any person who intends to
apply for grant of licence in respect of medical devices for,—
(i)       import;
(ii)      manufacture for sale or for
distribution; and
(iii)     sale, stock, exhibit or
offer for sale,
may group all or any
medical device in accordance with the guidelines to be issued from time to time
by the Ministry of Health and Family Welfare in the Central Government, by
taking into consideration the technological changes or development in the field
of medical devices and in vitro diagnostic medical devices.
Rule - 6. Essential principles for
manufacturing medical devices.
Medical device manufacturer
shall follow the essential principles of safety and performance of medical
devices as may be specified in the guidelines issued by the Ministry of Health
and Family Welfare in the Central Government, from time to time keeping in view
the contemporary scientific and technological knowledge and development:
Provided that the
guidelines to be so specified shall be in conformity with the provisions of the
Act and these rules.
Rule - 7. Product standards for medical
device.
(1)     The medical device shall
conform to the standards laid down by the Bureau of Indian Standards
established under Section 3 of the Bureau of Indian Standards Act, 1985 (63 of
1985) or as may be notified by the Ministry of Health and Family Welfare in the
Central Government, from time to time.
(2)     Where no relevant Standard
of any medical device has been laid down under sub-rule (1), such device shall
conform to the standard laid down by the International Organisation for
Standardisation (ISO) or the International Electro Technical Commission (IEC),
or by any other pharmacopoeial standards.
(3)     In case of the standards
which have not been specified under sub-rule (1) and sub-rule (2), the device
shall conform to the validated manufacturer's standards.
Chapter III AUTHORITIES,
OFFICERS AND BODIES
Rule - 8. Licensing Authorities.
(1)     The Central Licensing
Authority shall be the competent authority for enforcement of these rules in
matters relating to—
(i)       import of all Classes of
medical devices;
(ii)      manufacture of Class C and
Class D medical devices;
(iii)     clinical investigation and
approval of investigational medical devices;
(iv)    clinical performance
evaluation and approval of new in vitro diagnostic medical devices; and
(v)      coordination with the State
Licensing Authorities.
(2)     The State Drugs Controller,
by whatever name called, shall be the State Licensing Authority and shall be
the competent authority for enforcement of these rules in matters relating to—
(i)       manufacture for sale or
distribution of Class A or Class B medical devices;
(ii)      sale, stock, exhibit or
offer for sale or distribution of medical devices of all classes.
Rule - 9. Delegation of powers of Licensing
Authorities.
(1)     The Central Licensing
Authority, may with the prior approval of the Central Government, by an order
in writing, delegate all or any of its powers to any other officer of the
Central Drugs Standard Control Organisation not below the rank of Assistant
Drugs Controller.
(2)     The officer to whom the
powers have been delegated under sub-rule (1) shall exercise the powers of the
Central Licensing Authority under its name and seal.
(3)     The State Licensing
Authority, may, with the prior approval of the State Government, by an order in
writing, delegate all or any of its powers to any officer under its control.
(4)     The officer to whom the
powers have been delegated under sub-rule (3) shall exercise the powers of the
State Licensing Authority under its name and seal.
Rule - 10. Controlling officer.
Any officer not below the
rank of Assistant Drugs Controller, by whatever name called, shall be the
controlling officer to supervise and give instructions to any officer
subordinate to such controlling officer to exercise powers and functions under
these rules for areas and purposes specified, by an order, of the Drugs
Controller General of India or the Drugs Controller, by whatever name called,
of the State concerned.
Rule - 11. National Accreditation Body.
(1)     The Central Government may,
by notification, designate such institute, firm or a Government aided or
Government organisation, which fulfils the criteria specified from time to time
by the Government, as the National Accreditation Body:
Provided that the National
Accreditation Board for Certification Bodies under the Quality Council of
India, registered under the Societies Registration Act, 1860 (21 of 1860) set
up by the Ministry of Commerce and Industry in the Government of India shall
act as the National Accreditation Body for the purposes of accrediting Notified
Bodies referred to in Rule 13, till such time any other body for the purpose is
notified, with immediate effect.
(2)     The National Accreditation
Body shall have the required number of competent persons for proper performance
of its functions.
(3)     The designated National
Accreditation Body referred to in sub-rule (1) shall be responsible for
carrying out the assessment of such entities who may apply for accreditation to
become a Notified Body for the purpose of these rules.
(4)     The National Accreditation
Body referred to in sub-rule (1), shall, after carrying out the assessment of
the entity which applied for accreditation, issue a certificate to such entity
in respect of specified categories of standards for which such entity has been
assessed and found qualified:
Provided that where the
entity has been found not possessing the requisite qualification and other
requirements, the National Accreditation Body, shall reject the application.
(5)     The National Accreditation
Body shall not act as a Notified Body.
Rule - 12. Functions of National
Accreditation Body.
The National Accreditation
Body shall,—
(a)      lay down the conformity
assessment activities for accreditation of Notified Bodies and lay down
standards for such accreditation;
(b)      prepare norms and
procedures for accreditation of Notified Body;
(c)      audit a Notified Body
periodically for assessing conformance with these rules and the norms laid down
by it.
Rule - 13. Notified Body.
(1)     Any institute, organisation
or body corporate may seek accreditation, after notification of these rules, as
a Notified Body by applying to the National Accreditation Body referred to in
Rule 11 in such form and manner as may be determined by the National
Accreditation Body from time to time.
(2)     The Notified Body
accredited under sub-rule (1) shall be competent to carry out audit of
manufacturing sites of Class A [3][(other
than non-sterile and non-measuring)] and Class B medical devices to verify
conformance with the Quality Management System and other applicable standards
as specified under these rules in respect of such medical devices as and when
so advised by the State Licensing Authority.
(3)     Any Notified Body
accredited under sub-rule (1) shall, if it intends to carry out audit of a
manufacturing site of Class A [4][(other
than non-sterile and non-measuring)] or Class B of medical devices in
accordance with sub-rule (2), register with the Central Licensing Authority.
(4)     Any Notified Body under
sub-rule (3), with an experience of at least two years, may apply to the
Central Licensing Authority for registration as a Notified Body for carrying
out audit of Class C and Class D medical devices, provided it has personnel
with requisite qualification and experience.
(5)     With effect from the 1st
day of the July, 2017, the Notified Body accredited in accordance with sub-rule
(3) may make an application to the Central Licensing Authority for registration
in Form MD-1 through online portal accompanied with a fee specified in the
Second Schedule along with documents as specified in Part I of the Third
Schedule.
(6)     The Central Licensing
Authority, on being satisfied, shall register the Notified Body and issue a
registration certificate in Form MD-2.
(7)     The Registration
Certificate shall remain valid in perpetuity, unless, it is suspended or
cancelled, provided the registration certificate holder deposits a registration
retention fee as specified in the Second Schedule every five years from the
date of its issue.
(8)     If the registration
certificate holder fails to pay the required registration certificate retention
fee on or before due date as referred to in sub-rule (7), the registration
certificate holder shall, in addition to the retention fee, be liable to pay a late
fee calculated at the rate of two per cent of the registration certificate
retention fee for every month or part thereof within ninety days, and in the
event of non-payment of such fee during that period, the registration
certificate shall be deemed to have been cancelled.
(9)     The Notified Body shall
perform the functions as specified in Part II of the Third Schedule.
(10)   The Central Licensing
Authority, may, in cases where the requirement specified for registration of
Notified Body have not been complied with, reject the application and shall
inform the applicant of the reasons for such rejection.
(11)   An applicant who is
aggrieved by the decision of the Central Licensing Authority under sub-rule
(10), may file an appeal within forty-five days from the date of receipt of
such rejection before the Central Government, which may after such enquiry and
after giving an opportunity of being heard to the appellant, dispose of the
appeal within a period of sixty days.
Rule - 14. Duties of Notified Body.
A registered Notified Body,
referred to in Rule 13, shall carry out its duties and functions, in respect of
Class A [5][(other
than non-sterile and non-measuring)] or Class B medical devices as specified in
Part II of the Third Schedule.
Rule - 15. Procedure to be adopted by
Notified Body.
A registered Notified Body
shall carry out its duties and functions either by itself or by any other
qualified person on its behalf as per specified procedure as detailed in Part
II of the Third Schedule.
Rule - 16. Fees to be charged by Notified
Body.
A registered Notified Body
may charge fee from the applicant for the services rendered by it as may be
determined by the Central Government.
Rule - 17. Suspension and cancellation of
registration certificate of Notified Body.
(1)     The Central Licensing
Authority may, after giving an opportunity to show cause as to why such an
order should not be passed, by an order in writing stating the reasons
therefor, cancel the registration of a Notified Body or suspend it for such
period as the Central Licensing Authority thinks fit, if in its opinion, the
Notified Body has failed to comply with any provision of these rules.
(2)     A registered Notified Body
whose registration has been suspended or cancelled under sub-rule (1) may,
within thirty days of the receipt of a copy of the order by it, prefer an
appeal to the Central Government and the Central Government may, after giving
the Notified Body an opportunity of being heard, confirm, reverse or modify
such order.
(3)     The registration of a
Notified Body with the Central Licensing Authority shall be deemed to have been
cancelled with effect from the expiry of the date of the validity of its
accreditation by a National Accreditation Body.
Rule - 18. Medical Device Testing Officer
and Medical Device Officer.
(1)     The Central Government may
designate a Government Analyst appointed under Section 20 of the Act as Medical
Device Testing Officer.
(2)     The Central Government or,
as the case may be, the State Government, may designate an Inspector appointed
under Section 21 of the Act as Medical Device Officer.
(3)     The Medical Device Testing
Officer and Medical Device Officer designated under sub-rule (1) and sub-rule
(2) respectively, while exercising powers and duties under the Act and these
rules, shall be deemed to have been appointed as the Government Analyst and the
Inspector, respectively.
Rule - 19. Central medical device testing
laboratory.
(1)     The Central Government may,
by notification, establish Central medical devices testing laboratory for the
purpose of,—
(a)      testing and evaluation of
medical devices; or
(b)      functioning as an appellate
laboratory; or
(c)      to carry out any other
function as may be specifically assigned to it.
(2)     Without prejudice to
sub-rule (1), the Central Government may also designate any laboratory having
facility for carrying out test and evaluation of medical devices as central
medical devices testing laboratory for the purposes specified in sub-rule (1):
Provided that no medical
devices testing laboratory, shall be so designated unless it has been duly
accredited by the National Accreditation Body for Testing and Calibration
Laboratories:
[6][Provided further that the
testing laboratories of State Governments and Central Government shall be
exempted from the requirement of the accreditation by the National Accreditation
Board for Testing and Calibration Laboratories for a period of two years from
the date of coming into force of the Medical Devices (Fifth Amendment) Rules,
2019.]
NOTIFICATIONS
Noti. No. S.O. 2237(E),
dated 1-6-2018.—In
pursuance of the powers conferred by sub-rule (2) of Rule 19 of the Medical
Devices Rules, 2017, the Central Government hereby designates the laboratories
specified in column (2) of the Table below having facilities for carrying out
test and evaluation of medical devices, as Central Medical Device Testing
Laboratory for the purposes of—
(a)      testing and evaluation;
(b)      functioning as an appellate
laboratory; and
(c)      to carry out any other
function as may be specifically assigned to it by the Central Government, in
relation to the medical devices specified in column (3) of the said Table.
TABLE
| 
   Serial Number  | 
  
   Name of Laboratory  | 
  
   Category of medical device  | 
 
| 
   (1)  | 
  
   (2)  | 
  
   (3)  | 
 
| 
   1.  | 
  
   The National Institute of Biologicals, Noida  | 
  
   In-Vitro Diagnostics for human Immunodeficiency
  virus, Hepatitis B Surface Antigen and Hepatitis C Virus, Blood Grouping
  sera, Glucose Test Strip, Fully Automated Analyser Based Glucose Reagent  | 
 
| 
   2.  | 
  
   The Central Drugs Testing laboratory, Chennai  | 
  
   Condoms  | 
 
| 
   3.  | 
  
   The Central Drugs Laboratory, Kolkata  | 
  
   Surgical Dressings, Surgical Cotton, Surgical
  Bandages, Disinfectant  | 
 
| 
   4.  | 
  
   The Regional Drugs Testing Laboratory (RDTL),
  Guwahati  | 
  
   Disposable Hypodermic Syringes, Disposable
  Hypodermic Needle, Disposable Perfusion Sets, I.V. Cannulae  | 
 
| 
   5.  | 
  
   The Central Drugs Testing Laboratory, Mumbai  | 
  
   Intra Uterine Devices (IUD) and Falope Rings  | 
 
2. The laboratories
designated above are duly accredited by the National Accreditation Body for
Certification Laboratories.
3. This Order shall come
into force on and from the date of its publication in the Official Gazette.
[7][Chapter IIIA REGISTRATION OF CERTAIN MEDICAL DEVICES
Rule - 19A.
(1)     This Chapter shall be applicable
to all devices notified under clause (b) of Section 3 of the Act except the
medical devices and devices specified in the Annexure of Eighth Schedule of
these rules.
(2)     The Medical devices
referred in sub-rule (1) shall be registered with the Central Licensing
Authority through an identified online portal established by the Central Drugs
Standard Control Organisation for this purpose:
Provided that registration
under this Chapter shall be on voluntary basis for a period of eighteen months
from the commencement of this Chapter there after it shall be mandatory.
Rule - 19B.
(1)     The manufacturer of a
medical device shall upload the information specified in sub-rule (2) relating
to that medical device for registration on the “Online System for Medical
Devices” established by the Central Drugs Standard Control Organisation for
this purpose.
(2)     The manufacturer shall
upload,—
(i)       name & address of the
company or firm or any other entity manufacturing the medical device along with
name and address of manufacturing site of medical device,
(ii)      Details of medical device
| 
   Generic Name  | 
  
   Mo-del No.  | 
  
   Inten-ded Use  | 
  
   Class of Medi-cal device  | 
  
   Material of Construc-tion  | 
  
   Dimen-sion (if any)  | 
  
   Shelf Life  | 
  
   Sterile or Non Sterile  | 
  
   Brand Name (if registe-red under the Trade Marks
  Act, 1999)  | 
 
| 
      | 
  
(iii)     certificate of compliance
with respect to ISO 13485 standard accredited by National Accreditation Board
for Certification Bodies or International Accreditation Forum in respect of
such medical device:
[8][Provided that in case the
applicant submits, on or before the 28th February, 2022, an undertaking that
applicant shall obtain the ISO 13485 certificate on or before the 31st May,
2022 in lieu of certificate of compliance as referred in clause (iii) of sub-rule
(2) of Rule 19B, a provisional registration number shall be generated which
will remain valid up to the 31st May, 2022 or the date on which the applicant
obtained such ISO certificate whichever is earlier. The said generated
provisional registration number shall be valid for all purposes.
Explanation.—For the
removal of doubt, it is hereby declared that in case of such ISO 13485
certificate not obtained before the 31st May, 2022 as per undertaking referred
in the Proviso by the applicant the provisional registration shall be deemed to
have been cancelled for all purposes without any notice.]
(iv)    undertaking duly signed by
the manufacturer stating that the information furnished by the applicant is
true and authentic.
Rule - 19C.
After furnishing of the
above information on the “Online System for Medical Devices” established by
Central Drugs Standard Control Organisation for this purpose by the
applicant's, registration number will be generated. Manufacturer [9][may,
if so desired, mention the registration number or provisional registration
number, as the case may be, for a period up to the 31st May, 2022, thereafter
it shall be mandatory for all registration holders] on the label of the medical
device.
Rule - 19D.
(1)     Any person who imports any
medical device referred in Rule 19A shall upload the following information
relating to that medical device for registration on the “Online System for
Medical Devices” established by the Central Drugs Standard Control Organisation
for this purpose.
(2)     The importer shall upload,—
(i)       name of the company or firm
or any other entity importing the medical device and specification and
standards of that medical device,
(ii)      Details of medical device
| 
   Generic Name  | 
  
   Mo-del No.  | 
  
   Inten-ded Use  | 
  
   Class of Medi-cal device  | 
  
   Material of Construc-tion  | 
  
   Dimen-sion (if any)  | 
  
   Shelf Life  | 
  
   Sterile or Non Sterile  | 
  
   Brand Name (if registe-red under the Trade Marks
  Act, 1999)  | 
 
| 
      | 
  
(iii)     certificate of compliance
with respect to ISO 13485 standard accredited by National Accreditation Board
for Certification Bodies or International Accreditation Forum in respect of
such medical device:
[10]Provided that in case the
applicant submits, on or before the 28th February, 2022, an undertaking that
applicant shall obtain the ISO 13485 certificate on or before the 31st May,
2022 in lieu of certificate of compliance as referred in clause (iii) of sub-rule
(2) of Rule 19D, a provisional registration number shall be generated which
will remain valid up to the 31st May, 2022 or the date on which the applicant
obtained such ISO certificate whichever is earlier. The said generated
provisional registration number shall be valid for all purposes.
Explanation.—For the
removal of doubt, it is hereby declared that in case of such ISO 13485
certificate not obtained before the 31st May, 2022 as per undertaking referred
in the Proviso by the applicant the provisional registration shall be deemed to
have been cancelled for all purposes without any notice.]
(iv)    Free sale certificate from
country of origin.
(v)      undertaking duly signed by
the importer stating that the information furnished by the applicant is true
and authentic.
Rule - 19E.
After furnishing of the
above information on the “Online System for Medical Devices” established by the
Central Drugs Standard Control Organisation for this purpose by the
applicant's, registration number will be generated. Importer [11][may,
if so desired, mention the registration number or provisional registration
number, as the case may be, for a period up to the 31st May, 2022, thereafter
it shall be mandatory for all registration holders] on the label of the medical
device.
Rule - 19F.
Central Licensing Authority
may verify the documents at any point of time and investigate quality or safety
related failure or complaints. The Central Licensing Authority may, after
giving the registrant an opportunity to show cause as to why such an order should
not be passed, by an order in writing stating the reasons therefor, cancel the
registration number or suspend it for such period as the Central Licensing
Authority thinks fit either wholly or in respect of any of the medical devices
to which it relates, if in its opinion, the registrant has failed to comply
with any provision of these rules.]
[12][Chapter IIIB REGISTRATION OF CLASS A (NON-STERILE AND
NON-MEASURING) MEDICAL DEVICES
Rule - 19G. Application of this Chapter.
(1)     This Chapter shall be
applicable to all non-sterile and non-measuring devices classified as Class A
medical devices as per the First Schedule (herein in this chapter referred to
as Class A non-sterile and non-measuring medical device).
(2)     The medical devices
referred to in sub-rule (1) shall be registered through an identified online
portal established for the purpose.]
Rule - 19H. [Uploading of information for
registration.
(1)     The manufacturer of a Class
A non-sterile and non-measuring medical device shall upload the information
specified in sub-rule (2) relating to that medical device for registration on
the Online System for Medical Devices.
(2)     The manufacturer shall
upload the following in the Online System for Medical Devices, namely—
(i)       name and address of the
manufacturing site;
(ii)      details of Class A
non-sterile and non-measuring medical devices to be provided:
| 
   Generic name  | 
  
   Brand Name (if registered under the Trade Marks
  Act, 1999)  | 
  
   Model No (if any)  | 
  
   Intended use  | 
  
   Material of construction  | 
  
   Dimension (if applicable)  | 
  
   Shelf life (if applicable)  | 
 
| 
   (1)  | 
  
   (2)  | 
  
   (3)  | 
  
   (4)  | 
  
   (5)  | 
  
   (6)  | 
  
   (7)  | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
 
(iii)     an undertaking from the
manufacturer stating that the proposed device is a Class A non-sterile and
non-measuring medical device, as per the First Schedule;
(iv)    the manufacturer shall
self-certify that the product is conforming to the essential principles
checklist of safety and performance of such devices;
(v)      the manufacturer shall
self-certify to comply with the standards specified in these rules; and
(vi)    an undertaking duly signed
by the manufacturer stating that the information furnished by the applicant is
true and authentic.][13] 
Rule - 19I. [Registration number.
The registration number for
a Class A non-sterile and non-measuring medical device shall be generated after
furnishing of the information in accordance with Rule 19H on the Online System
for Medical Devices established for this purpose.][14] 
Rule - 19J. [Import of Class A non-sterile
and non-measuring medical device.
(1)     Any person who intends to
import any Class A non-sterile and non-measuring medical device shall upload
the information in sub-rule (2) relating to that medical device for
registration on the Online System for Medical Devices.
(2)     The importer shall upload
the following in the Online System for Medical Devices, namely—
(i)       name and address of
importer and the name and address of the manufacturing site;
(ii)      details of Class A
non-sterile and non-measuring medical devices to be provided:
| 
   Generic name  | 
  
   Brand Name (if registered under the Trade Marks
  Act, 1999)  | 
  
   Model No (if any)  | 
  
   Intended use  | 
  
   Material of construction  | 
  
   Dimension (if applicable)  | 
  
   Shelf life (if applicable)  | 
 
| 
   (1)  | 
  
   (2)  | 
  
   (3)  | 
  
   (4)  | 
  
   (5)  | 
  
   (6)  | 
  
   (7)  | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
 
(iii)     an undertaking from the
importer stating that the proposed device is Class A non-sterile and non-measuring
medical device, as per the First Schedule;
(iv)    the importer shall
self-certify that the product is conforming to the essential principles
checklist of safety and performance of such devices;
(v)      the importer shall
self-certify to comply with the standards specified in these rules;
(vi)    self-attested copy of the
overseas manufacturing site or establishment or plant registration, by whatever
name called, in the country of origin issued by the competent authority or Free
Sale Certificate issued by the National Regulatory Authority; and
(vii)   an undertaking duly signed
by the importer stating that the information furnished by the applicant is true
and authentic.][15]
Rule - 19K. [Registration number for
import.
The registration number for
import of a Class A non-sterile and non-measuring medical device shall be
generated after furnishing of the information in accordance with Rule 19J on
the Online System for Medical Devices established for this purpose.][16] 
Rule - 19L. [Maintainance of records.
(1)     The manufacturer or, as the
case may be, importer shall maintain the records relating to manufacturing or
importing along with its sales or distribution.
(2)     The manufacturer or, as the
case may be, importer shall produce the records, labels, instructions for use,
on request by Licensing Authorities.
(3)     The Licensing Authorities
may verify the records and documents referred to in sub-rule (2) at any point
of time and investigate quality or safety related failures or complaints.][17] 
Rule - 19M. [Cancellation or suspension of registration.
(1)     The State Licensing
Authority or the Central Licensing Authority, as the case may be, may, after
giving the registrant an opportunity to show cause as to why such an order
should not be passed, by an order in writing stating the reasons thereof,
cancel the registration number generated under the provisions of Rule 19I or
Rule 19K, or suspend it for such period as the Licensing Authority thinks fit,
either wholly or in respect of any of the medical devices to which it relates,
if in its opinion, the registrant has failed to comply with any of the
provisions of the rules under this Chapter.
(2)     Any person who is aggrieved
by an order passed by the State Licensing Authority or the Central Licensing
Authority, as the case may be, may, within forty-five days of the receipt of a
copy of such order, prefer an appeal to the State Government or the Central
Government, as the case may be, and the State Government or the Central
Government, shall after giving the said appellant an opportunity of being heard,
confirm, reverse or modify such order.][18] 
Chapter IV MANUFACTURE OF
MEDICAL DEVICES FOR SALE OR FOR DISTRIBUTION
Rule - 20. Application for manufacture for
sale or for distribution of Class A [(other than non-sterile and
non-measuring)][19]
or Class B medical device.
(1)     Any person who intends to
manufacture a Class A [20][(other
than non-sterile and non-measuring)] or Class B medical device including in
vitro diagnostic medical device shall make an application for grant of licence
or loan licence to manufacture for sale or for distribution to the State
Licensing Authority.
(2)     The application under
sub-rule (1) shall be made through an identified online portal of the Ministry
of Health and Family Welfare in the Central Government in Form MD-3 for licence
or in Form MD-4 for loan licence accompanied with a fee, as specified in the
Second Schedule along with respective documents as specified in Part II of the
Fourth Schedule.
(3)     The application made under
sub-rule (1), shall, amongst others, be accompanied with an undertaking to the
effect that the requirements of Quality Management System as specified in the
Fifth Schedule have been complied with.
(4)     The State Licensing
Authority shall, after scrutiny of documents and on being satisfied that the
requirements of these rules have been complied with, grant a licence to
manufacture Class A [21][(other
than non-sterile and non-measuring)] medical devices in Form MD-5 or loan
licence in Form MD-6, as the case may be, or if not satisfied, reject the
application for reasons to be recorded in writing, within forty five days from
the date, the application is made under sub-rule (1).
Provided that—
(i)       no audit of the
manufacturing site shall be necessary prior to grant of licence or loan licence
to manufacture for sale or for distribution of Class A [22][(other
than non-sterile and non-measuring)] medical device; and
(ii)      the required audit of such
manufacturing site by the registered Notified Body in the manner as specified
in the Third Schedule shall be carried out within one hundred and twenty days
from the date on which the licence was granted by the State Licensing
Authority.
(5)     Manufacturing site of the
applicant, in respect of Class B device, shall conform to the requirements of
Quality Management System as specified under the Fifth Schedule and applicable
standards as specified under these rules and such conformance shall be verified
through an audit by a Notified Body as referred under Rule 13 before grant of
licence.
(6)     In case of application for
grant of licence or loan licence to manufacture for sale or for distribution of
Class B medical devices—
(i)       the audit of the
manufacturing site shall be carried out within ninety days from the date of
application by the registered Notified Body in the manner specified in the
Third Schedule;
(ii)      the Notified Body shall
furnish its report to the State Licensing Authority within thirty days of the
completion of audit;
(iii)     the State Licensing
Authority shall, after scrutiny of documents, audit report as referred to in
clause (ii) and on being satisfied that the requirements of these rules have
been complied with, grant a licence to manufacture Class B medical devices in
Form MD-5 or loan licence in Form MD-6, as the case may be, or if not
satisfied, reject the application for reasons to be recorded in writing, within
a period of twenty days from the date of receipt of the report of audit by the
Notified Body.
(7)     If the application for
grant of licence or loan licence to manufacture for sale or for distribution is
rejected under sub-rule (4) or sub-rule (6), the aggrieved person may file an
appeal before the State Government within forty-five days from the date of
receipt of such rejection, which may, after such enquiry and after giving an
opportunity of being heard to the appellant, be disposed of within a period of
sixty days.
(8)     Where the Central Licensing
Authority or the State Licensing Authority has reason to believe or it has been
alleged or suspected that the medical device does not conform to the standards
of quality, or the provisions of the Fifth Schedule are not complied with, the
State licensing Authority, in case of Class A [23][(other
than non-sterile and non-measuring)] or Class B medical device, or the Central
Licensing Authority, in case of any Class of medical device, may direct a team
of officers referred to in Rule 23 to cause inspection of licensed
manufacturing site.
Rule - 21. Application for manufacturing
Class C or Class D devices.
(1)     An application shall be
made to the Central Licensing Authority through an identified online portal of
the Central Government for licence or loan licence to manufacture for sale or
for distribution, as the case may be, of Class C or Class D medical device in
Form MD-7 or Form MD-8, respectively.
(2)     The application in Form
MD-7 or Form MD-8 referred to in sub-rule (1) relating to Class C or Class D
medical device, as the case may be, shall be accompanied with a fee as
specified in the Second Schedule along with documents as specified in clause
(ii) of Part II of the Fourth Schedule.
(3)     The Central Licensing
Authority may, wherever required, in case of Class C or Class D medical
devices, use the services of any expert in the relevant field for scrutiny of
application and other technical documents.
(4)     The scrutiny referred to in
sub-rule (3) shall be completed by the Central Licensing Authority within a
period of forty-five days from the date of online submission of application.
(5)     In case, where the
documents are found to be complete and in order, the Central Licensing
Authority shall cause an inspection of the manufacturing site carried out under
Rule 23 by a team of officers accompanied by such experts, as may be considered
necessary.
(6)     The Central Licensing
Authority may, where required, avail the services of a Notified Body referred
to in sub-rule (4) of Rule 13 for inspecting the manufacturing site of Class C
and Class D medical devices.
(7)     In case, where the
documents furnished with the application referred to in sub-rule (1) are not
found to be complete and in order, the Central Licensing Authority shall reject
the application and inform the applicant of the reasons for such rejection
electronically:
Provided that where
deficiencies that can be rectified, are pointed out by the Central Licensing
Authority within the stipulated period, the period referred to in sub-rule (4)
shall reckon from the date these deficiencies have been removed.
Rule - 22. Requirements for grant of
manufacturing licence or loan licence.
While making application
for grant of licence or loan licence under Rule 20 or Rule 21, the applicant
shall meet the following requirements, namely—
(i)     the manufacturing site
shall comply with the requirements of the Quality Management System as
specified under the Fifth Schedule;
(ii)    appoint competent technical
staff under whose direction and supervision the manufacturing activity of a
medical device shall be undertaken and such staff shall possess the following
educational qualification and experience,—
(a)      degree in engineering in
relevant branch or in pharmacy or in science in appropriate subject from a
recognised University and shall have experience of not less than two years in
manufacturing or testing of medical devices; or
(b)      diploma in engineering (in
relevant branch) or in pharmacy from a recognised institute and shall have the
experience of not less than four years in manufacturing or testing of medical
devices;
(iii)     appoint competent technical
staff with degree or diploma in engineering (in relevant branch) or in pharmacy
or in science in relevant subject and having experience of not less than two
years in testing of medical devices under whose direction and supervision, the
testing activity of a medical device shall be undertaken.
Rule - 23. Inspection for grant of licence
or loan licence for Class C or Class D medical device.
(1)     Before grant of licence to
manufacture for sale or for distribution in respect of Class C or D medical
device, the manufacturing site shall be inspected within a period of sixty days
from the date of application by a team comprising not less than two Medical
Device Officers which may include any officer senior to the Medical Device
Officer with or without an expert, or a Notified Body referred to in sub-rule
(4) of Rule 13:
Provided that no inspection
of a medical device manufacturing site for grant of loan licence to manufacture
such medical device shall be required to be carried out if the manufacturing
site is already licensed to manufacture such medical device for sale or for
distribution.
(2)     The composition of the
inspection team referred to in sub-rule (1) shall be determined by the
controlling officer and no inspection shall be carried out without prior
approval of the controlling officer.
Rule - 24. Inspection report.
After completion of
inspection as referred to in Rule 23, the inspection team shall forward a
descriptive report containing findings on each aspect of inspection along with
the recommendations to the Central Licensing Authority, through online portal
of the Ministry of Health and Family Welfare in the Central Government and
forward a copy of the same to the applicant.
Rule - 25. Grant of licence or loan licence
to manufacture for sale or for distribution.
(1)     If the Central Licensing
Authority, after receipt of the report as referred to in Rule 24, and such
further enquiry, if any, as may be considered necessary, is satisfied that the
requirements of these rules have been complied, that Authority shall grant a licence
in Form MD-9, or loan licence in Form MD-10 or may reject the application for
reasons to be recorded in writing, within a period of forty-five days from the
date the inspection report has been received.
(2)     If the application for
grant of licence or loan licence to manufacture for sale or for distribution is
rejected under sub-rule (1), the aggrieved person may file an appeal before the
Central Government within forty-five days from the date of receipt of such
rejection, which may, after such enquiry and after giving an opportunity of
being heard to the appellant, be disposed of within a period of sixty days.
(3)     In case, a licensee or loan
licensee intends to manufacture additional medical devices in the licensed
manufacturing site, the manufacturer shall make an application for grant of
permission to manufacture such medical devices to the Central Licensing
Authority or State Licensing Authority, as the case may be, along with the fee
as specified in the Second Schedule and the documents as referred to in Rule 20
or Rule 21, as the case may be.
(4)     In case of investigational
medical device or new in vitro diagnostic medical device, the applicant shall
obtain prior permission in Form MD-27 or Form MD-29 from the Central Licensing
Authority and no licence to manufacture any class of such medical device shall
be granted without such permission.
Rule - 26. Conditions for manufacturing
licence or loan licence.
After grant of licence or
loan licence in Form MD-5, Form MD-6, Form MD-9 or MD-10, as the case may be,
the licence holder shall comply with the following conditions, namely—
(i)       licence shall be produced
when requested by the Medical Device Officer or any other senior officer under
the control of Central Licensing Authority or State Licensing Authority, as the
case may be;
(ii)      the licence holder shall
inform the State Licensing Authority or the Central Licensing Authority, as the
case may be, of the occurrence of any suspected unexpected serious adverse
event and action taken thereon including any recall within fifteen days of such
event coming to the notice of licence holder;
(iii)     the licence holder shall
obtain prior approval from the Central Licensing Authority or the State
Licensing Authority, as the case may be, before any major change as specified
in the Sixth Schedule is carried out and the Central Licensing Authority or the
State Licensing Authority, as the case may be, shall indicate its approval or
rejection within forty five days and in case where no communication is received
within the stipulated time from such Authority, such change shall be deemed to
have been approved;
(iv)    the licence holder shall
inform any minor change as specified in the Sixth Schedule to the State
Licensing Authority or Central Licensing Authority, as the case may be, within
a period of thirty days after such minor change take place;
(v)      the licence holder shall
carry out test of each batch of product manufactured prior to its release for
compliance with specifications either in his own laboratory or in any other
laboratory registered under sub-rule (3) of Rule 83;
(vi)    the licence holder shall,
on being informed by the Central Licensing Authority or State Licensing
Authority, as the case may be, that any part of any lot of the medical device
has been found not conforming with the provisions specified under the Act and these
rules, and on being directed so to do by such licensing authority, withdraw the
remainder of that lot from sale and, so far as may, in the particular
circumstances of the case, be practicable, recall the issues already made from
that lot;
(vii)   the licence holder shall
maintain an audit or inspection book in Form MD-11 to enable the Notified Body
or Medical Device Officer to record his observations and non-conformity, if
any;
(viii)  the licence holder shall
maintain at least one unit of sample from each batch of invasive medical device
and in vitro diagnostic medical device manufactured for reference purpose for a
period of one hundred and eighty days after the date of expiry of such batch;
(ix)    the licence holder shall
maintain records of manufacturing and sales which shall be open to inspection
by a Medical Device Officer;
(x)      the medical device, when
offered for sale, shall be accompanied by either its package insert or user
manual, wherever applicable;
(xi)    the manufacturing or
testing activity of medical device shall be undertaken only under the direction
and supervision of the competent technical staff;
(xii)   if the manufacturer has
stopped manufacturing activity or closed the manufacturing site for a period of
thirty days or more, the same shall be intimated to the Central Licensing
Authority or the State Licensing Authority, as the case may be.
Rule - 27. Change in constitution.
In case of change in
constitution of a licensee, after grant of licence under sub-rule (4) of Rule
20 or sub-rule (6) of Rule 20 or sub-rule (1) of Rule 25, as the case may be,
the manufacturer inform the Central Licensing Authority or the State Licensing
Authority, as the case may be, within forty-five days and shall shall make an
application under sub-rule (1) of Rule 20 or sub-rule (1) of Rule 21, as the
case may be, for grant of licence within a period of one hundred eighty days
from the date of such change in constitution:
Provided that the existing
licence shall be deemed to be valid till such time, a fresh licence is issued
or application is rejected by the State Licensing Authority or the Central
Licensing Authority, as the case may be:
Provided further that if
the application is rejected, the manufacturer may appeal to the Central
Government or the State Government, as the case may be, within a period of
sixty days.
Rule - 28. Unannounced inspection by State
Licensing Authority.
The State Licensing
Authority shall, in cases where licence has been granted for manufacturing
Class A and Class B medical devices under Rule 20, cause an inspection of the
manufacturing site to be carried out by a Medical Device Officer on a random
basis and such inspection shall not be less than two per cent of the total
audits carried out by Notified Bodies within that State for that class of
medical device.
Rule - 29. Validity of licence.
(1)     A licence or loan licence
issued in Form MD-5, Form MD-6, Form MD-9 or Form MD-10 shall remain valid in
perpetuity, subject to payment of licence retention fee as specified in the
Second Schedule before completion of the period of five years from the date of
its issue, unless, it is suspended or cancelled by State Licensing Authority or
the Central Licensing Authority, as the case may be.
(2)     If the licence holder fails
to pay the required licence retention fee on or before due date as referred to
in sub-rule (1), the licence holder shall, in addition to the licence retention
fee, be liable to pay a late fee calculated at the rate of two per cent of the
licence retention fee for every month or part thereof within one hundred and
eighty days and in the event of non-payment of such fee during that period, the
licence shall be deemed to have been cancelled.
Rule - 30. Suspension and cancellation of
licence.
(1)     Where the licensee
contravenes any provision of the Act and these rules, the State Licensing
Authority or the Central Licensing Authority, as the case may be, shall, after
giving the licensee an opportunity to show cause as to why such an order should
not be passed, shall by an order and for reasons to be recorded in writing,
suspend it for such period as it considers necessary either wholly or in
respect of any of the medical device or cancel the licence or loan licence.
(2)     A licensee whose licence or
loan licence has been suspended or cancelled by the State Licensing Authority
or the Central Licensing Authority, as the case may be, under sub-rule (1), may
within forty-five days of the receipt of a copy of the order by such authority,
prefer an appeal to the State Government or the Central Government, as the case
may be, and the State Government or the Central Government, shall after giving
the licensee an opportunity of being heard, confirm, reverse or modify such
order.
(3)     The State Licensing Authority
or the Central Licensing Authority, as the case may be, may revoke suspension
order issued under sub-rule (2) for reasons to be recorded in writing.
(4)     Orders of suspension issued
or revoked; or cancellation of licence shall be duly published on the concerned
websites of the State Licensing Authority or the Central Licensing Authority,
as the case may be.
Rule - 31. Test licence to manufacture for
test, evaluation, clinical investigations, etc.
(1)     Small quantity of Class
A [24][(other
than non-sterile and non-measuring)] or Class B or Class C or Class D of
medical devices may be manufactured for the purpose of clinical investigations,
test, evaluation, examination, demonstration or training for which an
application shall be made in Form MD-12 to the Central Licensing Authority and
shall be accompanied with a fee as specified in the Second Schedule.
(2)     The application made under
sub-rule (1) shall also be accompanied with the following documents, namely—
(a)      brief description of the
medical device including intended use, material of construction, design and an
undertaking stating that the required facilities including equipment,
instruments, and personnel have been provided to manufacture such medical
devices;
(b)      list of equipment,
instruments;
(c)      list of qualified personnel;
(d)      copy of manufacturing
licence issued under these rules, if any;
(e)      approval letter authorising
to undertake research and development activities issued by any Government
organisation, if any.
(3)     The Central Licensing
Authority, after enquiry, if any, as may be considered necessary, on being
satisfied that the requirements of these rules have been complied, shall grant
a test licence in Form MD-13, or may reject the application for reasons to be
recorded in writing, within a period of thirty days from the date the
application is made under sub-rule (1).
(4)     The licensee shall maintain
a record of the details of quantity of the product manufactured under test
licence.
(5)     A licence granted under
sub-rule (3) shall, unless cancelled earlier, remain in force for a period of
three years from the date of its issuance.
Rule - 32. Conditions of test licence to
manufacture for test, evaluation, clinical investigations, etc.
A licence in Form MD-13
under Rule 31 shall be subject to the following conditions, namely—
(a)      the licensee shall use the
medical device manufactured under licence granted under sub-rule (3) of Rule 31
exclusively for the purpose of clinical investigations, test, evaluation,
examination, demonstration or training at the place specified in the licence;
(b)      the licensee shall allow
any Medical Device Officer to enter, with or without notice, the premises where
the medical device are manufactured and to satisfy himself that only clinical
investigations, test, evaluation, examination, demonstration or training is
being conducted on such device;
(c)      the licensee shall maintain
a record of the quantity of medical device manufactured, tested and stocked and
its disposition.
Rule - 33. Cancellation of test licence to
manufacture for test, evaluation, clinical investigations, etc.
(1)     Where any licensee under
Rule 31 contravenes any provision of these rules, the Central Licensing
Authority, shall, issue a show cause notice to such licensee asking, as to why
an order should not be made to cancel the licence.
(2)     The Central Licensing
Authority shall, after giving an opportunity to the licensee to explain in
writing licensee's defence, pass an order for cancellation or otherwise and
record the reasons therefor in the said order.
(3)     A licensee, whose licence
has been cancelled, may appeal to the Central Government within forty-five days
from the date of the order.
Chapter V IMPORT OF MEDICAL
DEVICES
Rule - 34. Application for grant of import
licence.
(1)     An authorised agent having licence
to manufacture for sale or distribution or wholesale licence for sale or
distribution [25][or
registration certificate in Form MD-42] under these rules, shall make an
application for grant of import licence for medical device to the Central
Licensing Authority through an identified online portal of the Ministry of
Health and Family Welfare in the Central Government in Form MD-14 for obtaining
a licence.
(2)     The application under
sub-rule (1) shall be accompanied with the fee as specified in the Second Schedule
along with respective documents as specified in the Fourth Schedule:
Provided that any change in
the documents submitted at the time of application and prior to grant of
licence shall be informed, in writing, to the Central Licensing Authority.
(3)     Where the Central Licensing
Authority, has reason to believe that the quality of the medical device is
compromised, and decides to subject it to evaluation, test or examination, the
authorised agent shall pay a fee for such evaluation, test or examination, to
the testing laboratory as specified by the Central Licensing Authority.
(4)     Any subsequent application
for,—
(i)       grant of licence for
additional manufacturing site for the same medical device by the same
authorised agent shall be accompanied with a fee and documents as referred in
sub-rule (2);
(ii)      licence for additional
medical device manufactured at the same manufacturing site shall be made by the
same authorised agent accompanied with fee as specified in the Second Schedule
and respective documents as specified in the Fourth Schedule.
Rule - 35. Inspection of overseas
manufacturing site.
(1)     On receipt of an
application under sub-rule (1) of Rule 34, the Central Licensing Authority, may
cause an inspection of the overseas manufacturing site either by itself or by any
other person or body to whom the power has been delegated for the purpose.
(2)     The applicant shall be
liable to pay a fee as specified under the Second Schedule in respect of
expenditure required in connection with the visit to the overseas manufacturing
site under sub-rule (1).
Rule - 36. Grant of import licence.
(1)     After examination of
documents furnished with the application under sub-rule (1) of Rule 34 and on
the basis of the inspection report, if inspection has been carried out, the
Central Licensing Authority may, on being satisfied, grant licence in Form
MD-15 or, may reject such application for which reasons shall be recorded in
writing, within a period of nine months from the date of application.
(2)     In the event of rejection,
the applicant may appeal to the Central Government within a period of
forty-five days and that Government, may, after such enquiry into the matter,
as considered necessary, pass orders in relation thereto within a period of
ninety days from the date of appeal.
(3)     Where, a free sale certificate
has already been issued in respect of any medical device by the national
regulatory authority or other competent authority of any of the countries
namely, Australia, Canada, Japan, European Union Countries, [26][United
Kingdom or the United States of America], a licence shall be granted under
sub-rule (1) to the applicant without carrying out clinical investigation.
(4)     Where a medical device is
imported from countries other than those referred to in sub-rule (3), the
licence in case of Class C and Class D medical devices may be granted after its
safety and effectiveness has been established through clinical investigation in
India as specified under provisions of Chapter VII of these rules.
(5)     Where a medical device, is
imported from countries other than those referred to in sub-rule (3), the
licence in case of Class A [27][(other
than non-sterile and non-measuring)] or Class B medical devices may be granted
after its safety and performance has been established through published safety
and performance data or through clinical investigation in the country of origin
and a free sale certificate from the country of origin is furnished.
(6)     In case of investigational
medical device or new in vitro diagnostic medical device, the applicant shall
obtain prior permission in Form MD-27 or in Form MD-29 from the Central
Licensing Authority and no licence to import any class of such medical device
shall be granted without such permission.
Rule - 37. Validity of licence.
A licence granted under
sub-rule (1) of Rule 36 shall remain valid in perpetuity, unless, it has been
cancelled or surrendered, provided the authorised agent deposits the licence
retention fee with the Central Licensing Authority as specified in the Second
Schedule for each overseas manufacturing site and for each licensed medical
device after completion of every five years from the date of its issue:
Provided that the Central
Licensing Authority may permit to deposit the licence retention fee after due
date but before expiry of ninety days with a late fee calculated at the rate of
two per cent per mensem:
Provided further that if
the licensee fails to deposit the licence retention fee within the above
stipulated period, the licence shall be deemed to have been cancelled.
Rule - 38. Conditions to be complied with
by licence holder.
(1)     The licensee shall comply
with the following conditions, namely—
(i)       licence shall be produced
when requested by the Medical Device Officer or any other senior officer under
the control of Central Licensing Authority or the State Licensing Authority, as
the case may be;
(ii)      the licensee shall inform
the licensing authority forthwith and, in all circumstances, within a period of
fifteen days of any administrative action taken on account of any adverse
reaction, such as market withdrawal, regulatory restrictions, cancellation of
authorisation or declaration of the medical device as not of standards quality
by the regulatory authority of the country of origin or by any regulatory
authority of any other country, where the medical device is marketed, sold or
distributed;
(iii)     authorised agent in cases
referred in clause (ii), shall stop immediately the despatch and marketing of
the medical device referred in that clause;
(iv)    the Central Licensing
Authority, after due consideration of the information as referred in clause
(ii), may issue directions to the licensee in respect of marketing, sale or distribution
of the medical device including withdrawal of medical device from the Indian
market within a period as may be specified by the Central Licensing Authority;
(v)      the authorised agent shall
obtain prior approval from the Central Licensing Authority before any major
change, as specified in the Sixth Schedule, is carried out and the Central
Licensing Authority shall indicate its approval or rejection within sixty days;
(vi)    in case, no communication
of approval or rejection as referred to in clause (v) is received within the
stipulated time from the Central Licensing Authority, such change shall be
deemed to have been approved;
(vii)   licensee shall inform, any
minor change as specified in the Sixth Schedule to the Central Licensing
Authority within a period of thirty days, after such minor change took place;
(viii)  authorised agent shall
inform the Central Licensing Authority in writing within a period of thirty
days in the event of any change in the constitution of the overseas
manufacturer or the authorised agent;
(ix)    the consignment of medical
device shall be accompanied by an invoice or statement showing the name and
quantity of the medical device;
(x)      the licensee shall supply
the medical device for sale or offer it for sale along with its package insert
or user manual, wherever applicable.
(2)     Where the Central Licensing
Authority is satisfied that any medical device is not in conformity with the
provisions of the Act and these rules, it may issue directions that the entire
batch of such medical device may not be sold or offered for sale or may be
recalled from the market including hospitals, if any, where it has been
stocked:
Provided that where the
Central Licensing Authority considers it necessary or expedient, more than one
batch or all batches of such medical device may be directed to be recalled.
Rule - 39. Fresh application in case of
change in constitution.
In case of change in
constitution of a licensee, after grant of licence under sub-rule (1) of Rule
36, an application shall be made under sub-rule (1) of Rule 34 for grant of
licence within a period of one hundred and eighty days from the date of such
change in constitution:
Provided that the existing
licence shall be deemed to be valid till such time, the fresh licence is issued
or application is rejected by the Central Licensing Authority.
Explanation.—For the
purpose of this rule, the licensee shall include overseas manufacturer who
executed the power of attorney in favour of authorised agent.
Rule - 40. Test licence for import for
test, evaluation, clinical investigations, etc.
(1)     Notwithstanding anything
contained in these rules, any medical device or in vitro diagnostic medical
device may be imported for the purpose of clinical investigations or test or
evaluation or demonstration or training.
(2)     The person who desires to
import medical device under sub-rule (1), shall apply for an import licence for
test, evaluation or demonstration or training to the Central Licensing
Authority in Form MD-16, accompanied by such fee as specified in the Second
Schedule.
(3)     On receipt of the
application under sub-rule (2), the Central Licensing Authority shall
determine, the quantity of the medical devices, after taking into account the
requirement for clinical investigation, approved clinical investigation plan,
and information and documents submitted by the applicant.
Rule - 41. Grant of test licence for import
for test, evaluation, clinical investigations, etc.
(1)     If the Central Licensing
Authority, after such enquiry, if any, is satisfied that the requirements of
these rules have been complied, the said authority shall grant a test licence
in Form MD-17, or may reject the application for reasons to be recorded in writing,
within a period of thirty days from the date the application under sub-rule (2)
of Rule 40.
(2)     The medical device for
which a test licence has been granted under sub-rule (1), shall be used
exclusively for purposes of clinical investigation, test, evaluation,
demonstration or training, as the case may be, and such clinical investigations
or test or evaluation or training, shall be conducted at a place specified in
such test licence:
Provided that in cases
where the medical device is required to be taken to any place other than the
ones mentioned in the test licence, the Central Licensing Authority shall be
informed in writing before doing so.
(3)     The holder of the test
licence shall maintain record of the activities undertaken including the name
of manufacturer, quantity imported and date of import.
(4)     The consignment of medical
device shall be accompanied by an invoice or statement showing the name and
quantity of medical device.
(5)     A licence in Form MD-17
shall, unless cancelled earlier, be in force for a period of three years from
the date of its issue.
(6)     The medical devices
including in vitro diagnostic medical device referred to in sub-rule (2) that
are not used, may be permitted to be exported or destroyed under intimation to
the Central Licensing Authority.
(7)     Where any licensee under
sub-rule (1) contravenes any provision of these rules, the Central Licensing
Authority, shall, issue a show cause notice to such licensee asking, as to why
an order should not be made to cancel the licence.
(8)     The Central Licensing
Authority shall after giving an opportunity to the licensee to explain, in
writing, licensee's defence, pass an order for cancellation or otherwise and
record the reasons therefor in the said order.
(9)     A licensee, whose licence
has been cancelled under sub-rule (8), may appeal to the Central Government
within forty-five days from the date of such order.
Rule - 42. Import of investigational
medical device by Government hospital or statutory medical institution for
treatment of patient.
(1)     Small quantity of
investigational medical device, the import of which is not allowed, but
approved in the country of origin, may be allowed to be imported by the Central
Licensing Authority for treatment of a patient suffering from a life threatening
disease or disease causing serious permanent disability or disease requiring
therapy for unmet medical need, on an application made by a Medical Officer
through the medical superintendent of a Government hospital or a statutory
medical institution in Form MD-18 and such application shall be accompanied by
documents required and the fee as specified in the Second Schedule.
(2)     On receipt of an
application under sub-rule (1), the Central Licensing Authority shall, on being
satisfied about the information and the documents enclosed with the
application, grant import licence for treatment of patient in Form MD-19.
(3)     The medical device for
which the licence is granted under sub-rule (2), shall, be used exclusively for
the purpose of treatment of the patient referred to in sub-rule (1).
(4)     The holder of licence shall
maintain record of the name of the manufacturer, quantity imported and used,
date of import, name and address of the patient and diagnosis.
(5)     The holder of the licence
shall allow the medical device officer authorised by the Central Licensing
Authority in this behalf to enter, with or without prior notice, the premises
where the medical devices are stocked and to inspect the premises and relevant
records and investigate the manner in which the medical device is being used
and to take, if required, samples thereof.
(6)     The quantity considered
necessary shall be determined by the Central Licensing Authority after taking
into account the recommendation of the hospital concerned for treatment of
patient suffering from a life threatening disease or disease causing serious
permanent disability or disease requiring therapy for unmet medical need.
(7)     Where the Central Licensing
Authority is satisfied, it may, in exceptional and special circumstances, allow
import of larger quantity of medical devices for use by the patient.
(8)     The consignment of medical
device shall be accompanied by an invoice or a statement showing the name and
quantity of medical device.
Rule - 43. Import of medical device for
personal use.
(1)   Small quantity of medical
device, the import of which is otherwise prohibited under Section 10 of the
Act, may be imported for personal use subject to the following conditions,
namely—
(i)       the medical device shall
form part of a personal baggage of a passenger and be intended for the
exclusive use of such passenger;
(ii)      the medical device shall be
declared as personal baggage of the passenger to the customs authorities, if
they so direct;
(iii)     the quantity of any single
medical device so imported shall not exceed the quantity specified by the
registered medical practitioner;
(iv)    the medical device has been
prescribed by a registered medical practitioner; and
(v)      the medical device so
imported shall be accompanied with an invoice or a statement showing the name
and quantity of medical device.
(2)     Small quantity of medical
device, the import of which is otherwise prohibited under Section 10 of the
Act, and which is not forming a part of bona fide personal baggage, may be
imported for personal use, on an application made by the applicant in Form
MD-20 and such application shall be accompanied by documents confirming that
the device is for bona fide personal use and a prescription to that effect by a
registered medical practitioner.
(3)     On receipt of an
application under sub-rule (2), the Central Licensing Authority shall, on being
satisfied about the information and the documents enclosed with the
application, grant permission in Form MD-21 or may reject the application for
reasons to be recorded in writing within a period of seven days from the date
of application under sub-rule (2).
(4)     Medical devices as referred
to in sub-rule (2) shall be subject to the following conditions, namely—
(i)       the medical device shall be
declared to the Customs Authorities if they so direct;
(ii)      the consignment of the
medical device so imported shall be accompanied with an invoice or statement
showing the name and quantity of medical device.
Chapter VI LABELLING OF
MEDICAL DEVICES
Rule - 44. Labelling of medical devices.
The following particulars
shall be printed in indelible ink on the label, on the shelf pack of the
medical device or on the outer cover of the medical device and on every outer
covering in which the medical device is packed, namely—
(a)      name of the medical device;
(b)      the details necessary for
the user to identify the device and its use;
(c)      the name of manufacturer
and address of manufacturing premises where the device has been manufactured;
(d)      the correct statement about
the net quantity in terms of weight, measure, volume, number of units, as the
case may be, and the number of the devices contained in the package expressed
in metric system;
(e)      the month and year of
manufacture and expiry (alternately the label shall bear the shelf-life of the
product):
Provided that in case of
sterile devices, the date of sterilisation may be given as date of manufacture
of the device:
Provided further that where
the device is made up of stable materials such as stainless steel or titanium,
and supplied non-sterile or in case of medical equipment or instruments or
apparatus, the date of expiry may not be necessary.
Explanation.—For the
purposes of this clause, the date of expiry shall be in terms of the month and
the year and it shall mean that the medical device is recommended till the last
day of the month and the date of expiry shall be preceded by the words “Expiry
date” or “Shelf Life”;
(f)       to provide, wherever
required, an indication that the device contains medicinal or biological
substance;
(g)      to provide, a distinctive
batch number or lot number preceded by the word “Lot No.” or “Lot” or “Batch
No.” or “B. No.”;
(h)     to indicate, wherever
required, any special storage or handling conditions applicable to the device;
(i)       to indicate, if the device
is supplied as a sterile product, its sterile state and the sterilisation
method;
(j)       to give, if considered
relevant, warnings or precautions to draw the attention of the user of medical
device;
(k)      to label the device appropriately,
if the device is intended for single use;
(l)       to overprint on the label
of the device, the words “Physician's Sample—Not to be sold”, if a medical
device is intended for distribution to the medical professional as a free
sample;
(m)    to provide, except for
imported devices, the manufacturing licence number by preceding the words
“Manufacturing Licence Number” or “Mfg. Lic. No.” or “M. L”;
(n)     to provide on the label, in
case of imported devices, by way of stickering, where such details are not
already printed, the import licence number, name and address of the importer,
address of the actual manufacturing premises and the date of manufacture:
Provided that the label may
bear symbols recognised by the Bureau of Indian Standards or International Organisation
for Standardisation (ISO) in lieu of the text and the device safety is not
compromised by a lack of understanding on the part of the user, in case the
meaning of the symbol is not obvious to the device user;
(o)      in case of small sized
medical devices on which information cannot be printed legibly, shall include
the information necessary for product identification and safety such as
information covered by clauses (a), (b), (c), (d), (e), (g), (k), and (m) shall
be included.
Rule - 45. Exemption of labelling
requirements for export of medical devices.
The labels on packages or
container of devices for export shall be adopted to meet the specific
requirements of law of the country to which the device is to be exported, but
the following particulars shall appear in a conspicuous manner on the label of
the inner most pack or shelf pack of the medical device in which the device is
packed and every other outer covering in which the container is packed—
(a)      name of the device;
(b)      the distinctive batch
number or lot number or serial number preceded by the word “Lot No.” or “Lot”
or “Batch No.” or “B. No.” or “Serial No.”;
(c)      date of expiry, if any;
(d)      the name and address of
manufacturer and address of actual premises where the device has been
manufactured;
(e)      licence number preceded by
letters “Licence No. or Lic. No.”;
(f)       internationally recognised
symbols in lieu of text, wherever required:
Provided that where a
device is required by the consignee not to be labeled with the name and address
of manufacturer, the label on the package or container shall bear a code number
as approved by the Central Licensing Authority and the code number shall bear
the name of the State or Union Territory, in abbreviation, followed by the word
“Device” and “manufacturing licence number”:
Provided further that where
a device is required by the consignee not to be labeled with the code number
also, the label on the packages or container shall bear a special code number,
as requested by the consignee, and approved by the Central Licensing Authority.
Rule - 46. [Unique device identification of
medical device.
With effect from such date
as the Central Government may, by order specify, every medical device approved
for manufacture for sale or distribution or import, shall bear a unique device identification
in the manner as may be specified in such order.][28] 
Rule - 47. Shelf life of medical devices.
The shelf-life of the
medical devices, shall be determined keeping in view the technical parameters
and shall ordinarily not exceed sixty months from the date of manufacture to be
reckoned from month to month (i.e. January to January), except in cases where
satisfactory evidence is produced by the manufacturer to justify a shelf-life
of more than sixty months of a device to the satisfaction of the Central
Licensing Authority:
Provided that any medical
device, whose total shelf-life claim is less than ninety days, shall not be
allowed to be imported by the licensing authority if it has less than forty per
cent residual shelf-life on the date of import:
Provided further that any
medical device, whose total shelf-life claim is between ninety days and one
year, shall not be allowed to be imported by the licensing authority if it has
less than fifty per cent residual shelf-life on the date of import:
Provided also that any
medical device, whose total shelf-life claim is more than one year, shall not
be allowed to be imported by the licensing authority if it has less than sixty
per cent residual shelf-life on the date of import.
Rule - 48. Labelling medical device or a
new in vitro diagnostic medical device for purpose of test, evaluation,
clinical investigations, etc.
Any medical device or new
in vitro diagnostic medical device imported or manufactured, for the purpose of
clinical investigation or clinical performance evaluation, test, evaluation,
demonstration and training, shall be kept in containers bearing labels,
indicating the name of the product or code number, batch or lot number, serial
number wherever applicable, date of manufacture, use before date, storage
conditions, name and address of the manufacturer, and the purpose for which it
has been manufactured.
Chapter VII CLINICAL
INVESTIGATION OF MEDICAL DEVICE AND CLINICAL PERFORMANCE EVALUATION OF NEW IN
VITRO DIAGNOSTIC MEDICAL DEVICE
Rule - 49. Conduct of clinical
investigation.
No person or sponsor shall
conduct any clinical investigation in respect of investigational medical device
in human participants except in accordance with these rules and in accordance
with the permission granted by the Central Licensing Authority.
Rule - 50. Application of Rule 122-DD of
Drugs and Cosmetics Rules, 1945 with regard to Ethics Committee.
(1)     The Ethics Committee
constituted under Rule 122-DD of the Drugs and Cosmetics Rules, 1945 shall
perform the functions and duties under these rules and shall be deemed to be
constituted under these rules.
(2)     The provisions of Ethics
Committee provided in Rule 122-DD of the Drugs and Cosmetics Rules, 1945 shall,
except where specifically provided under these rules, be applicable mutatis
mutandis, for the purpose of clinical investigation and clinical performance
evaluation under this Chapter.
Rule - 51. Application for grant of
permission to conduct clinical investigation.
(1)     An application for grant of
permission to conduct clinical investigation for investigational medical device
shall be made to the Central Licensing Authority in Form MD-22 by a sponsor and
shall be accompanied with information specified in the Seventh Schedule.
(2)     An application for grant of
permission to conduct,—
(a)      a pilot clinical
investigation on an investigational medical device as referred to in sub-rule
(1) shall be accompanied with a fee as specified in the Second Schedule along
with information as specified in the Seventh Schedule.
Explanation.—For the purposes
of these rules, the pilot clinical investigation means clinical investigation
to be carried out for the first time in human participants;
(b)      a pivotal clinical
investigation on an investigational medical device shall be made on the basis
of data emerging from pilot clinical investigation, accompanied with a fee as
specified in the Second Schedule:
Provided that no fee shall
be payable by any institute, organisation, hospital run or funded by the
Central Government or the State Government, as the case may be, for conduct of
clinical investigation.
(3)     No permission for conduct
of academic clinical study on licensed medical device shall be required,
where,—
(a)      the Ethics Committee
approves such a study; and
(b)      the data generated during
the study shall not be used to furnish to the Central Licensing Authority to
manufacture or to import for marketing any investigational medical device in
the country.
(4)     The Central Licensing
Authority may, in public interest, abbreviate, defer, or waive the requirement
of animal data or clinical data for conducting clinical investigation for
reasons to be recorded in writing before granting permission to conduct
clinical investigation.
(5)     Medical devices requiring
clinical investigation but claiming substantial equivalence to a predicate device
shall not be marketed unless the Central Licensing Authority has approved it.
Explanation 1.—For the
purposes of this sub-rule, a device shall be deemed to be substantially
equivalent in comparison to a predicate device, if it has—
(i)       the same intended use and
technological characteristics; or
(ii)      same intended use and
different technological characteristics, and demonstrate that the device is as
safe and effective as the predicate device.
Explanation 2.—A claim of
substantial equivalence does not mean that the proposed medical device and
predicate device are identical. Substantial equivalence shall be established
with respect to intended use, design, energy used or delivered, materials,
chemical composition, manufacturing process, performance, safety, effectiveness,
labelling, bio-compatibility, standards, and other characteristics, as
applicable.
Rule - 52. Permission to conduct clinical
investigation.
The Central Licensing
Authority, after such further enquiry, if any, as considered necessary, may,—
(i)       if satisfied, that the
requirements of these rules have been complied with, grant permission to
conduct clinical investigation for an investigational medical device in Form
MD-23;
(ii)      if not satisfied with the
requirements as referred to in sub-clause (i), reject the application, for
reasons to be recorded in writing,
within a period of ninety
days, from the date of application made under sub-rule (1) of Rule 51.
Rule - 53. Conditions for permission.
After grant of permission
referred to in Rule 52, the following conditions shall be complied with by the
applicant, namely—
(i)       clinical investigation
shall be initiated after approval of clinical investigation plan by the
registered Ethics Committee;
(ii)      clinical investigation
shall be conducted in accordance with the approved clinical investigation plan,
Good Clinical Practices Guidelines issued by the Central Drugs Standard Control
Organisation and provisions of the Seventh Schedule;
(iii)     clinical investigation
shall be registered with the Clinical Trial Registry of India before enrolling
the first participant for such clinical investigation;
(iv)    annual status report of
each clinical investigation, as to whether it is ongoing, completed or
terminated, shall be submitted to the Central Licensing Authority by the
sponsor, and, in case of termination of any clinical investigation, the
detailed reasons for the same shall be communicated to the Central Licensing
Authority within thirty days of such termination;
(v)      information about any
report of suspected unexpected serious adverse event occurring during clinical
investigation on the subject shall, after due analysis, be submitted by the
sponsor to the Central Licensing Authority within fourteen days of the
knowledge of its occurrence as specified in the Seventh Schedule and in
compliance with the procedure specified in these rules;
(vi)    in case of an injury or
death during clinical investigation of a subject of a clinical investigation,
the applicant shall provide complete medical management or compensation in
accordance with these rules;
(vii)   the premises of the sponsor
including their employees, subsidiaries and branches, their agents, contractors
and sub-contractors and clinical investigation sites shall be open for
inspection by officers of the Central Licensing Authority who may be accompanied
by officers of the State Licensing Authority or outside experts, to verify
compliance of the requirements of these rules for conduct of clinical
investigation;
(viii)  the clinical investigation
shall be initiated by enrolling first participant within a period of one year
from the date of grant of permission, failing which prior permission from the
Central Licensing Authority shall be required to initiate clinical
investigation;
(ix)    the Central Licensing
Authority may impose or exempt any condition while granting permission in
respect of specific clinical investigations, if considered necessary, regarding
the objective, design, subject population, subject eligibility, assessment,
conduct and treatment of clinical investigation.
Rule - 54. Suspension, cancellation, etc.
of permission.
(1)     If any person to whom
permission has been granted under Rule 52 fails to comply with any of the
conditions of permission or any of the provisions of the Act or these rules,
the Central Licensing Authority may,—
(a)      issue warning letter giving
details of deficiency found; or
(b)      debar the investigator or
sponsor including their employees, subsidiaries and branches, their agents,
contractors and sub-contractors to conduct any clinical investigation for such
period as it thinks fit; or
(c)      suspend the permission for
such period as it thinks fit or cancel either wholly or partly the permission.
(2)     Any person who is aggrieved
by the order passed under sub-rule (1), may file an appeal within thirty days
of the receipt of such order before the Central Government, which may, after
such enquiry and after giving an opportunity of being heard to the appellant,
dispose of the appeal within a period of sixty days.
Rule - 55. Medical management and
compensation related to clinical investigation.
(1)     Where any participant is
injured on account of participation in clinical investigation, the sponsor
permitted under Rule 52 shall provide medical management to that participant.
(2)     Where an injury is caused
to the participant in a clinical investigation of any investigational medical
device and such injury is attributable to the use of investigational medical
device, the sponsor permitted under Rule 52 shall provide to that participant,
medical management and such compensation in the manner as specified under Rule
122-DAB of the Drugs and Cosmetics Rules, 1945.
(3)     Where death of a
participant is related to clinical investigation and is attributable to the use
of an investigational medical device, the sponsor, permitted under Rule 52
shall provide to the legal heir of that participant, such compensation, in such
manner as specified under Rule 122-DAB of the Drugs and Cosmetics Rules, 1945.
Rule - 56. Powers of search and seizure,
etc.
The Medical Devices Officer
may enter any premises related to clinical investigation or clinical
performance evaluation, with or without an expert, with prior approval of the
Central Licensing Authority, with or without prior notice, to inspect the
facilities, search and seize, record, data, documents, books, and medical
devices including investigational medical devices or new in vitro diagnostic
medical device.
Rule - 57. Maintenance of record.
Every person, sponsor,
clinical research organisation, any other organisation or investigator
conducting a clinical investigation or his agent holding a permission under
this Chapter shall maintain such data, record, registers and other documents
for a period of seven years after completion of such investigation and shall
furnish such information as may be required by the Central Licensing Authority
or any other officer authorised by it in this behalf under Rule 56.
Rule - 58. Disclosure of name, address,
etc. of persons involved in clinical investigation or clinical performance
evaluation.
Every person, sponsor,
clinical research organisation, any other organisation or investigator
conducting a clinical investigation or clinical performance evaluation or any
agent authorised by any of them, as the case may be, shall, if so required,
disclose to the Medical Device Officer or any other officer authorised by the
Central Licensing Authority, the names, addresses and other particulars of
persons involved in clinical investigation.
Rule - 59. Permission to conduct clinical
performance evaluation for new in vitro diagnostic medical device.
(1)     No person or sponsor shall
conduct any clinical performance evaluation in respect of a new in vitro
diagnostic medical device on any specimen, including blood or tissue derived
from human body except under, and in accordance with, the permission granted by
the Central Licensing Authority subject to such conditions and in such form and
manner as specified in these rules.
(2)     An application for grant of
permission to conduct, clinical performance evaluation of new in vitro
diagnostic medical device shall be made to the Central Licensing Authority in
Form MD-24 by the sponsor and shall be accompanied with a fee as specified in
the Second Schedule along with information specified in sub-rule (3) duly
signed by the sponsor in India:
Provided that no fee shall
be required to be paid by the institutes, organisation, hospitals, run by the
Central Government or the State Government, involved in conduct of clinical
performance evaluation of new in vitro diagnostic medical devices.
(3)     The information required
under sub-rule (2) shall contain the following, namely—
(i)       approval from an Ethics
Committee, which is registered with the Central Licensing Authority, as
specified in Appendix VIII of the Schedule Y of the Drugs and Cosmetics Rules,
1945 and referred to in the Seventh Schedule;
(ii)      source and quantity of
samples which shall be used during evaluation;
(iii)     device description
including specification of raw material and finished product, data allowing
identification of the device in question, proposed instruction for use [29][or
electronic instructions for use], labels and regulatory status in other
countries, if any;
(iv)    in house performance
evaluation data used to establish stability, specificity, sensitivity,
repeatability and reproducibility;
(v)      clinical performance
evaluation plan stating in particular the purpose, scientific, technical or
medical grounds and scope of evaluation;
(vi)    Case Report Form as
specified in Table 6 of the Seventh Schedule;
(vii)   undertaking by
investigators as specified in Table 9 of the Seventh Schedule;
(viii)  the list of laboratories or
other institutions taking part in the evaluation study;
(ix)    the scheduled duration for
evaluation and, in case of devices for self-testing, the location and number of
lay persons involved;
(x)      an undertaking that the
device in question conforms to the requirements of these rules, apart from
aspects covered by evaluation and apart from those specifically itemised in the
undertaking, and that every precaution has been taken to protect the health and
safety of the patient, user and other persons.
(xi)    performance evaluation
report from a laboratory designated under sub-rule (1) of Rule 19.
(4)     The Central Licensing
Authority may, in public interest, abbreviate, defer, or waive the requirements
of conducting clinical performance evaluation for reasons to be recorded in
writing for grant of permission to conduct clinical performance evaluation.
(5)     If the Central Licensing
Authority, after such further enquiry, if any, as may be considered necessary,
is satisfied that the requirements of these rules have been complied, may grant
permission to conduct clinical performance evaluation for a new in vitro
diagnostic medical device in Form MD-25 or may reject the application, for
reasons to be recorded in writing, within a period of ninety days from the date
of application.
Rule - 60. Conditions for permission to
conduct of clinical performance evaluation.
After grant of permission
referred to in sub-rule (5) of Rule 59, the following conditions shall be
complied with by the applicant,—
(i)       clinical performance
evaluation shall be conducted in accordance with the approved clinical
performance evaluation plan and Good Clinical Practices Guidelines;
(ii)      clinical performance
evaluation shall be initiated after approval of clinical investigation plan by
the registered Ethics Committee;
(iii)     clinical performance
evaluation shall be registered with the Clinical Trial Registry of India before
enrolling the first participant for such clinical performance evaluation;
(iv)    annual status report of
each clinical performance evaluation, as to whether it is ongoing, completed or
terminated, shall be submitted to the Central Licensing Authority by the
sponsor, and in case of termination of any clinical performance evaluation, the
detailed reasons for the same shall be communicated to the Central Licensing
Authority within thirty days of the date of termination;
(v)      the laboratories or other
institutions taking part in the evaluation study or the sponsor including their
employees, subsidiaries and branches, their agents, contractors and
sub-contractors, and clinical investigation sites shall be open for inspection
by officers of the Central Licensing Authority authorised in this behalf who
may be accompanied by officers of State Licensing Authority or outside experts
under these rules to verify compliance of the requirements of these rules for
conduct of clinical performance evaluation;
(vi)    the clinical performance
evaluation shall be initiated within a period of one year from the date of
grant of permission, failing which prior permission from the Central Licensing
Authority shall be required to initiate such clinical performance evaluation;
(vii)   the Central Licensing
Authority may impose or exempt any condition while granting permission in
respect of specific clinical performance evaluation, if considered necessary,
regarding the objective, design, subject population, subject eligibility,
assessment, conduct and treatment of clinical performance evaluation.
Rule - 61. Suspension or cancellation of
permission.
(1)     If any person to whom
permission has been granted under sub-rule (5) of Rule 59 fails to comply with
any of the conditions of permission, the Central Licensing Authority may,
suspend the permission for such period as it thinks fit or cancel either wholly
or partly.
(2)     Any person who is aggrieved
by the order passed under sub-rule (1), may file an appeal within thirty days
before the Central Government, which may, after such enquiry and after giving
an opportunity of being heard to the appellant, dispose of the appeal within a
period of sixty days.
Rule - 62. Medical management.
Where any participant is
injured on account of his participation in the clinical performance evaluation,
the sponsor permitted under sub-rule (5) of Rule 59 shall provide medical
management to that participant.
Chapter VIII IMPORT OR
MANUFACTURE MEDICAL DEVICE WHICH DOES NOT HAVE PREDICATE DEVICE
Rule - 63. Permission to import or
manufacture medical device which does not have its predicate device.
(1)     Save as otherwise provided
in these rules, for import or manufacture of medical device which does not have
predicate medical device, an application for grant of permission for such
medical device after completion of its clinical investigation under Chapter VII
shall be made to the Central Licensing Authority in Form MD-26 either by an
authorised agent in case of import or a manufacturer, as the case may be, which
shall be accompanied with fee as specified in the Second Schedule along with
information specified in Part IV of the Fourth Schedule:
Provided that the medical
device which does not have predicate medical device indicated in life
threatening, serious diseases or diseases of special relevance to the Indian
health scenario, national emergencies, extreme urgency, epidemic and medical
devices indicated for conditions, diseases for which there is no therapy, the
animal data or clinical data requirements may be abbreviated, deferred or
omitted, as deemed appropriate by the Central Licensing Authority:
Provided further that in
respect of investigational medical device of Class A, data on clinical
investigation may not be required, except in cases, where depending on the
nature of the medical device, the Central Licensing Authority, for reason to be
recorded in writing, considers such data necessary:
Provided also that subject
to other provisions of these rules, in case of medical device of which drugs
are also a part, the submission of requirements relating to animal toxicology,
reproduction studies, teratogenic studies, perinatal studies, mutagenicity and
carcinogenicity may be relaxed in case of drugs already approved and marketed
in India and supported by adequate published evidence regarding safety of the
drug:
Provided also that, the
results of clinical investigation may not be required to be submitted where the
investigational medical device is approved by the regulatory authorities of
either the United Kingdom or the United States of America or Australia or
Canada or Japan and the said device has been marketed for at least two years in
that country and the Central Licensing Authority is satisfied with the data of
safety, performance and pharmacovigilance of the device, and,—
(a)      there is no evidence or
theoretical possibility, on the basis of existing knowledge, of any difference
in the behavior and performance in Indian population;
(b)      the applicant has given an
undertaking in writing to conduct post marketing clinical investigation with
the objective of safety and performance of such investigational medical device
as per protocol approved by the Central Licensing Authority.
(2)     The Central Licensing
Authority, after being satisfied with the information furnished along with
application under sub-rule (1), may grant permission to import or manufacture
medical device which does not have predicate medical device in Form MD-27, or
may reject the application for reasons to be recorded in writing, within a
period of one hundred and twenty days or such extended period, not exceeding a
further period of thirty days, from the date of application:
Provided that the Central
Licensing Authority shall, where the information is inadequate with regard to
the requirements as referred to in sub-rule (1), intimate the applicant in
writing within the said period, for reasons to be recorded in writing, the
conditions which shall be satisfied before considering the permission:
Provided further that if
the applicant has not furnished the required information sought by the Central
Licensing Authority within ninety days from the date of intimation and the said
Authority is satisfied that the information sought was possible to be furnished
within the said period, it may reject the application for reasons to be
recorded in writing.
(3)     If the applicant does not
receive permission or if the application is rejected within the specified
period as referred to in sub-rule (2), the applicant may appeal to the Central
Government and that Government may, after such enquiry, as it considers
necessary, pass such orders in relation thereto as it thinks fit within a
period of sixty days from the date of appeal.
Rule - 64. Permission to import or
manufacture new in vitro diagnostic medical device.
(1)     An application for grant of
permission to import or manufacture a new in vitro diagnostic medical device
may be made to the Central Licensing Authority in Form MD-28 either by an
authorised agent in case of import or a manufacturer himself, as the case may
be, and shall be accompanied with fee as specified in the Second Schedule along
with information specified in Part IV of the Fourth Schedule:
Provided that the new in
vitro diagnostic medical device used for diagnosis of life threatening, serious
diseases or diseases of special relevance to the Indian health scenario,
national emergencies, extreme urgency, epidemic and diagnostic medical devices
used for diagnosis of conditions, diseases for which there is no diagnostic
medical device available in the country, the clinical data requirements may be
abbreviated, deferred or omitted, as deemed appropriate by the Central
Licensing Authority:
Provided further that for
new in vitro diagnostic medical device classified under Class A, data on
clinical performance evaluation may not be necessary, except in cases, where
the Central Licensing Authority, for reasons to be recorded in writing,
considers it necessary depending on the nature of the medical device.
(2)     The Central Licensing
Authority, may, after being satisfied with the information furnished along with
application under sub-rule (1), grant permission to import or manufacture new
in vitro diagnostic medical device in Form MD-29 or may reject the application,
for reasons to be recorded in writing, within a period of ninety days or such
extended period, not exceeding a further period of thirty days, from the date
of application:
Provided that the Central
Licensing Authority shall, where the information is inadequate with regard to
the requirements as referred to in sub-rule (1), intimate the applicant in
writing within the said period, for reasons to be recorded in writing, the
conditions which shall be satisfied before considering permission:
Provided further that if
the applicant has not furnished the required information sought by the Central
Licensing Authority within ninety days from the date of intimation and the said
Authority is satisfied that the information sought was possible to be furnished
within the said period, it may reject the application for reasons to be
recorded in writing.
Rule - 65. Condition of permission to
import or manufacture medical device which does not have its predicate device
and new in vitro diagnostic medical device.
A Permission under Rules 63
in Form MD-27 and Rule 64 in Form MD-29 shall be subject to the following
conditions, namely—
(a)      the medical device shall
conform to the specifications submitted along with the application;
(b)      the permission holder of
Form MD-27 shall submit the Periodic Safety Update Report to the Central
Licensing Authority from the date of launch in the market and such report shall
be submitted every six months for first two years followed by submission of the
said report annually for the two more successive years;
(c)      the permission holder shall
inform the date of launch of medical device in the market to the Central
Licensing Authority;
(d)      the permission holder of
Form MD-27 shall submit the suspected unexpected serious adverse event within
fifteen days of the awareness of the event to the Central Licensing Authority.
Chapter IX DUTIES OF MEDICAL
DEVICE OFFICER, MEDICAL DEVICE TESTING OFFICER AND NOTIFIED BODY
Rule - 66. Duties of Medical Device Testing
Officer.
The Medical Device Testing
Officer shall cause the sample of medical device or portion thereof tested or
evaluated as may be sent in a sealed package by the Medical Device Officer or
any other person under the provisions of Chapters IV, V, VII and XI of these
rules, and shall furnish the report of the result of the test or evaluation in
accordance with these rules.
Rule - 67. Test or evaluation of sample
under sub-section (4) of Section 25 of the Act.
(1)     The sample of medical
device for test or evaluation under sub-section (4) of Section 25 of the Act
shall be sent by registered post in the outer cover addressed to the Director
of central medical device testing laboratory in a sealed packet with a
memorandum in Form MD-30.
(2)     The packet as well as the
outer cover shall be marked with a distinguishing number.
(3)     A copy of the memorandum in
Form MD-30 and a specimen impression of the seal used to seal the packet shall
be separately sent by registered post to the Director of central medical device
testing laboratory.
(4)     After test or evaluation,
the result of the test or evaluation shall be sent forthwith to the sender in
Form MD-31.
Rule - 68. Procedure to be adopted by
medical device testing officer on receipt of sample.
(1)     On receipt of the sealed
package of medical device or portion thereof, from a Medical Device Officer or
any other person for test or evaluation, the Medical Device Testing Officer
shall compare the seals on the packet or on portion thereof with the specimen
impression received separately and shall note the condition of the seals on the
packet or on portion thereof.
(2)     After completion of test or
evaluation, the Medical Device Testing Officer shall forthwith furnish a report
to the Medical Device Officer in triplicate in Form MD-32 of the result of the
test or evaluation along with full protocols of the test or evaluation applied.
Rule - 69. Application for test or
evaluation of medical device.
For the purpose of these
rules, an application from a purchaser for test or evaluation of a medical
device or portion of medical device under Section 26 of the Act shall be made
in Form MD-33 and the report of such test or evaluation of the medical device
which is prepared on such application shall be supplied to the applicant in
Form MD-32.
Rule - 70. Duties of Medical Device
Officer.
Subject to the instructions
of the Central Licensing Authority or State Licensing Authority, as the case
may be, it shall be the duty of Medical Device Officer to,—
(i)       Inspect, not less than once
in a year, all manufacturing sites licensed by the Central Licensing Authority
or State Licensing Authority, as the case may be, within the area assigned to
him;
(ii)      conform that the conditions
of licence are being observed;
(iii)     take samples of medical
device manufactured or imported for sale, or stocked or exhibited for sale in
respect of which the Medical Device Officer has reason to suspect contravention
of the provisions of the Act or these rules and send them for test or
evaluation:
Provided that in case of
large sized medical device, wherein the opinion of the Medical Device Officer
drawing samples of such a device may not be physically practical, such large
sized medical device shall be inspected at the place where these are kept by
the Medical Device Officer with or without expert and evaluated or tested by
the Medical Device Testing Officer, for any suspect contravention, after
approval of the Central Licensing Authority or the State Licensing Authority,
as the case may be;
(iv)    maintain a record of all
inspections undertaken, drawing of samples, seizure of stocks and action taken
by Medical Device Officer in exercise and performance of duties and to furnish
copies of such record to the Central Licensing Authority or the State Licensing
Authority, as the case may be;
(v)      make such enquiries and
inspections as may be necessary to detect the manufacture or sale of medical
device in contravention of any provision of the Act and these rules;
(vi)    investigate any complaint
made in writing relating to medical device to the Medical Device Officer or any
other senior officer in accordance with the direction of the controlling
officer;
(vii)   institute prosecution in
relation to contravention of the provisions of the Act and these rules;
(viii)  review technical dossier of
medical device furnished with the application under these rules or any other
duties assigned by the Central Licensing Authority or State Licensing Authority,
as the case may be, related to these rules.
Rule - 71. Prohibition of disclosure of
information.
Except for the purpose of
official business or when required by a Court, a Medical Device Officer or
Medical Device Testing Officer shall not, without the previous sanction, in
writing, of his official superior, disclose to any person any information
acquired while exercising such official duties.
Rule - 72. Form of order not to dispose of
stock.
An order in writing by a
Medical Device Officer under clause (c) of sub-section (1) of Section 22 of the
Act requiring a person not to dispose of any stock in his possession shall be
in Form MD-34.
Rule - 73. Prohibition of sale.
No person in possession of
a medical device in respect of which a Medical Device Officer has made an order
under clause (c) of sub-section (1) of Section 22 of the Act shall, in
contravention of that order, sell or otherwise dispose of any stock of such
medical device.
Rule - 74. Form of receipt for seized
medical devices, record, register, documents or any other material objects.
A receipt by a Medical
Device Officer for the stock of any medical device or for any record, register,
document or any other material object seized under clause (c) or clause (cc) of
sub-section (1) of Section 22 of the Act shall be in Form MD-35.
Rule - 75. Manner of certifying copies of
seized documents.
The Medical Device Officer
shall return the document, seized under Section 22 of the Act, within a period
of twenty days from the date of such seizure, to the person from whom they were
recovered or produced, after copies thereof or extracts therefrom have been
signed by the concerned Medical Device Officer and the person from whom they
were recovered or produced.
Rule - 76. Purpose for which samples have
been taken.
When a Medical Device
Officer takes a sample of a medical device other than medical device specified
in proviso to clause (iii) of Rule 70 for the purpose of test or evaluation,
the Medical Device Officer shall inform such purpose in writing in Form MD-36
to the person from whom the sample has been taken and shall tender the fair
price thereof under a written acknowledgement.
Rule - 77. Form of receipt for samples of
medical devices where fair price tendered is refused.
Where the fair price
tendered under sub-section (1) of Section 23 of the Act for sample of medical
device or portion thereof taken for the purposes of test or evaluation has been
refused by the person from whom such sample has been taken, the Medical Device
Officer shall tender a receipt thereof to such person in Form MD-37.
Rule - 78. Procedure for dispatch of sample
to medical device testing officer.
(1)     The sample of medical
device or portion thereof sent by the Medical Device Officer to the Medical
Device Testing Officer for test or evaluation under sub-section (4) of Section
23 of the Act shall be sent by registered post or by courier or by hand in a
sealed packet, enclosed with a memorandum in Form MD-38, in an outer cover
addressed to the Medical Device Testing Officer.
(2)     A copy of the memorandum
and a specimen impression of the seal used to seal the packet shall be sent to
the Medical Device Testing Officer separately by registered post or handed over
by hand and a copy of the memorandum shall be endorsed to the manufacturer.
Rule - 79. Confiscation of medical devices,
implements, machinery, etc.
(1)     Where any person has been
convicted for contravening any provisions of the Act or any (sic of) these
rules, the stock of medical device in respect of which the contravention has
been made, shall be liable to confiscation.
(2)     Where any person has been
convicted for manufacturing any medical device which is misbranded, adulterated
or spurious for sale, stocking or exhibiting for sale or distribution without a
valid licence, any implements or machinery used in such manufacture, sale or
distribution and any receptable, package or covering in which such medical
device is contained and the animals, vehicles, vessels or other conveyances
used in carrying such medical device shall be liable to confiscation.
Rule - 80. Procedure for disposal of
confiscated medical device.
(1)     The Court may refer the
confiscated medical device to the Medical Device Officer concerned for report
as to whether they are of standard quality or contravene the provisions of the
Act or the rules in any respect.
(2)     If the Medical Device
Officer, on the basis of Medical Device Testing Officer's report, finds the
confiscated medical device to be not of standard quality or to contravene any
of the provisions of the Act or rules made thereunder, the Medical Device
Officer shall, with the approval of the Central Licensing Authority or State
Licensing Authority, as the case may be, report to the Court accordingly and
the Court shall thereupon order destruction of such medical devices, which
shall take place under the supervision of the Medical Device Officer in the
presence of such authority, if any, as may be directed by the Court:
Provided that the convicted
person shall be liable to bear the cost of destruction of seized articles.
(3)     If the Medical Device
Officer finds that the confiscated medical devices are of standard quality and
do not contravene the provisions of the Act or the rules made thereunder, the
Medical Device Officer shall, after keeping the Central Licensing Authority or
the State Licensing Authority, as the case may be, informed, report to the
Court accordingly.
(4)     The Court may return the
confiscated devices to the rightful owner, and in case, the ownership is not
established, the same may be given to a hospital or a dispensary maintained or
supported by the Government or to a charitable institution.
Chapter X REGISTRATION OF
LABORATORY FOR CARRYING OUT TEST OR EVALUATION
Rule - 81. Application for registration of
medical device testing laboratory.
(1)     An application for grant of
registration of a medical device testing laboratory to carry out testing or
evaluation of a medical device on behalf of a manufacturer shall be made to the
Central Licensing Authority through online portal of the Central Government in
Form MD-39 accompanied with a fee as specified in the Second Schedule along
with the information specified in sub-rule (2).
(2)     The application made under
sub-rule (1) shall be accompanied with the following information, namely—
(i)       constitution of the medical
device testing laboratory;
(ii)      premises showing location
and area of the different sections;
(iii)     qualification, experience
of technical staff employed for testing and the person in-charge of testing;
(iv)    list of equipment; and
(v)      valid accreditation
certificate issued by the National Accreditation Body for Testing and
Calibration Laboratories or any other similar body as may be notified by the
Central Government.
Rule - 82. Conditions for registration of
medical device testing laboratory.
The following conditions
shall be complied with by the applicant before grant of registration, namely—
(i)       the premises where the test
or evaluation shall be carried out shall be well lighted and properly
ventilated except where the nature of tests of any medical device warrants
otherwise, and wherever necessary, the premises shall be air conditioned so as
to maintain the accuracy and functioning of laboratory instruments or to enable
the performance of special tests such as sterility tests, microbiological
tests, etc;
(ii)      the applicant shall provide
adequate space having regard to the nature and number of samples of medical
devices proposed to be tested and evaluated:
Provided that the approving
authority shall determine from time to time whether the space provided
continues to be adequate;
(iii)     if it is intended to carry out
tests requiring the use of animals, the applicant shall provide for an animal
house and comply with the following requirements—
(a)      the animal house shall be
adequate in area, well lighted and properly ventilated and the animals
undergoing tests shall be kept in air conditioned area;
(b)      the animals shall be
suitably housed in hygienic surroundings and necessary provisions made for
removal of excreta and foul smell;
(c)      the applicant shall provide
for suitable arrangements for preparation of animal feed;
(d)      the applicant shall provide
for suitable arrangements for quarantining of all animals immediately on their
arrival in the institution;
(e)      the animals shall be
periodically examined for their physical fitness;
(f)       the applicant shall provide
for isolation of sick animals as well as animals under test;
(g)      the applicant shall ensure
compliance with the requirements of the Prevention of Cruelty to Animals Act,
1960 (59 of 1960);
(h)     the applicant shall make
proper arrangements for disposal of the carcasses of animals in a manner as not
to cause hazard to public health.
Rule - 83. Registration of medical device
testing laboratory.
(1)     Before grant of registration
to any medical device testing laboratory by the Central Licensing Authority,
the premises shall be inspected by the Medical Device Officer appointed by the
Central Government with or without an expert in the concerned field for
adequacy and suitability.
(2)     The Medical Device Officer,
after completion of the inspection, shall forward a detailed descriptive report
giving findings on each aspect of inspection along with recommendations to the
Central Licensing Authority with a copy to the applicant.
(3)     If on receipt of the
application and the report referred to in sub-rule (2), the Central Licensing
Authority, is satisfied that the applicant is in a position to fulfill the
requirements laid down in these rules, the Central Licensing Authority may
grant registration in Form MD-40 or if not satisfied, may, reject the
application, for reasons to be recorded in writing, within a period of
forty-five days from the date of application.
(4)     The applicant shall provide
and maintain suitable equipment having regard to the nature and number of
samples of medical devices intended to be tested which shall be adequate in the
opinion of the Central Licensing Authority.
(5)     The testing and evaluation
of medical devices shall be under active direction of a person whose qualification
and experience is considered adequate and who shall be held responsible for
reports of test or evaluation issued.
(6)     The applicant shall provide
standards recognised under the provisions of the Act and these rules and such
standards of reference as may be required in connection with the testing or
evaluation of the devices for the testing of which approval has been applied
for.
Rule - 84. Validity of registration.
A registration granted
under sub-rule (3) of Rule 83 in Form MD-40, shall remain valid in perpetuity,
unless, it is suspended or cancelled, provided the registration holder deposits
a registration retention fee to the Central Licensing Authority as specified in
the Second Schedule after completion of every five years from the date of its
issue:
Provided, that the Central
Licensing Authority may permit to deposit the registration retention fee after
due date but before the expiry of six months with a late fee at the rate of two
per cent per mensem or part thereof:
Provided further that, if
the registration holder fails to deposit the registration retention fee within
the above stipulated period, the registration shall be deemed to have been
cancelled for all purposes.
Rule - 85. Conditions of registration.
A registration granted
under sub-rule (3) of Rule 83 in Form MD-40, shall be subject to the following
conditions, namely—
(i)       the registration
certificate shall be kept on the approved premises and shall be produced at the
request of the medical device officer;
(ii)      the person holding
registration certificate shall provide and maintain necessary qualified staff,
adequate premises and equipment;
(iii)     the person holding
registration certificate shall provide proper facilities for storage so as to
preserve the properties of samples picked up for testing;
(iv)    the person holding
registration certificate shall maintain records of tests for evaluation and
performance carried out on all samples of medical devices and the results
thereof together with protocols of tests and the reports showing readings and
calculations and such records shall be retained, in case of substances for
which an expiry date is assigned, for a period of two years beyond the expiry
date, and in the case of other substances, for a period of six years;
(v)      the person holding
registration certificate shall allow the medical device officer appointed under
this Act to enter, with or without prior notice, the premises where the testing
is carried out and to inspect the premises and the equipment used for test and
the testing procedures employed;
(vi)    The person holding
registration certificate shall allow the medical device officer to inspect
records maintained and shall make available such information as may be required
for ascertaining whether the provisions of the Act and these rules have been
complied with;
(vii)   the registration holder
shall inform forthwith, any change of existing expert staff or person in-charge
of the testing or evaluation to the Central Licensing Authority for its
approval;
(viii)  in case, any sample of a
medical device is found on test, to be not of standard quality, the person
in-charge of the registered medical device testing laboratory shall furnish a
copy of the test or evaluation report on the sample with the protocols of tests
applied to the Central Licensing Authority;
(ix)    the person holding
registration certificate shall maintain an inspection book to enable the
Medical Device Officer to record non-compliance with the provisions of the Act
and these rules;
(x)      the registered medical
device testing laboratory shall inform to the Central Licensing Authority in
writing in the event of any change in its constitution and where such change in
the constitution takes place, the current registration shall be deemed to be
valid for a maximum period of ninety days from the date on which the change
took place unless, in the meantime, a fresh approval has been taken from the
Central Licensing Authority with the changed constitution;
Rule - 86. Suspension and cancellation of
registration.
(1)     Where any registered
medical device testing laboratory fails to comply with any of the conditions of
approval, or any provisions of the Act and these rules, the Central Licensing
Authority, may issue a show cause notice for suspension or cancellation of the
registration of the said medical device testing laboratory.
(2)     On receipt of the show
cause notice under sub-rule (1), the registered medical device testing
laboratory shall, furnish its reply in writing, within fifteen days of the
receipt of such show cause notice.
(3)     After considering the reply
of the registered medical device testing laboratory furnished under sub-rule
(2), the Central Licensing Authority may pass an order in writing for
suspension or cancellation of the registration of the said medical device
testing laboratory registered under sub-rule (3) of Rule 83.
(4)     While passing orders under
sub-rule (3), the Central Licensing Authority may suspend or cancel the
registration wholly or partly in respect of medical device and its variant for
testing for such period as may be specified in the order.
(5)     An applicant, who is
aggrieved by an order of suspension or cancellation of registration under
sub-rule (3), may file an appeal within thirty days from the date of receipt of
such order before the Central Government, which may, after such enquiry and
after giving an opportunity of being heard, dispose of the appeal within a
period of sixty days.
Chapter XI SALE OF MEDICAL
DEVICES
Rule - 87. Provisions for sale of medical
devices.
(1)     Subject to the provisions
of these rules, Part VI relating to “Sale of Drugs Other than Homeopathic
Medicines” of the Drugs and Cosmetics Rules, 1945, shall be applicable mutatis
mutandis in respect of sale of medical devices.
[30][(1A) Any person not
holding licence under sub-rule (1) and intends to sell medical devices
exclusively as referred to in clause (zb) of Rule 3, shall obtain registration
certificate as provided in these rules.]
(2)     The licence granted or
renewed under Part VI of the Drugs and Cosmetics Rules, 1945 for sale of drugs,
prior to commencement of these rules, shall be deemed to continue to be valid
for the purpose of sale of medical devices.
Rule - 87A. [Registration certificate to
sell, stock, exhibit or offer for sale or distribute a medical device including
in vitro diagnostic medical device.
(1)     The State Licencing
Authority shall appoint Licensing Authorities for the purpose of issuing
registration certificate under this Part for such areas as may be specified.
(2)     Any person who intends to
sell, stock, exhibit or offer for sale or distribute a medical device including
in vitro diagnostic medical device, shall make an application in Form MD-41 to
the State Licensing Authority for grant of registration certificate to sell,
stock, exhibit or offer for sale or distribution.
(3)     The application made under
sub-rule (2) shall be accompanied with
(i)       a fees specified in Second
Schedule;
(ii)      self certificate of
compliance with respect to Good Distribution Compliance;
(iii)     details of the applicant or
firm including its constitution, along with identification proof, such as,
Aadhar card or PAN card;
(iv)    documentary evidence in
respect of ownership or occupancy on rental of the premises;
(v)      details of competent
technical staff, under whose direction and supervision the sales activity of
medical device shall be undertaken, who shall possess the following educational
qualification and experience, namely—
(a)      hold a degree from a
recognised University/Institution; or
(b)      is a registered pharmacist;
or
(c)      has passed intermediate
examination or its equivalent examination from a recognised Board with one-year
experience in dealing with sale of medical devices;
(vi)    brief description on other
activities carried out by applicant, namely, storage of drugs, medical items,
food products, stationeries, etc., or any other activities carried out by the
applicant in the said premises; and
(vii)   an undertaking to the
effect that the storage requirements to sell, stock, exhibit or offer for sale
or distribute a medical device will be complied with.
(4)     The State Licensing
Authority shall, after scrutiny of documents and on being satisfied that the
requirements of these rules have been complied with, grant a registration
certificate in Form MD-42, or if not satisfied, reject the application for
reasons to be recorded in writing, within ten days from the date the
application is made under sub-rule (2).
(5)     If the application for
grant of registration certificate to sell, stock, exhibit or offer for sale or
distribute a medical device is rejected under sub-rule (4), the aggrieved
person may prefer an appeal before the State Government within forty-five days
from the date of receipt of such rejection, which may, after such enquiry and
after giving an opportunity of being heard to the appellant, dispose it within
a period of sixty days from the date of receipt of such appeal.][31] 
Rule - 87B. [Conditions of registration
certificate to sell, stock, exhibit or offer for sale or distribute a medical
device including in vitro diagnostic medical device.
(1)     The registration
certificate granted under Rule 87A shall be displayed at a prominent place in
the premises visible to the public.
(2)     The registration
certificate holder shall provide adequate space and proper storage condition
for storage of the medical devices.
(3)     The registration
certificate holder shall maintain requisite temperature and lighting as per
requirements of such medical devices.
(4)     The medical devices shall
be purchased only from importer or licensed manufacturer or registered or
licensed entity.
(5)     Separate records, in the
form of invoice or register or electronic details including software of
purchases and sales of medical devices showing the names and quantities of such
medical devices, names and addresses of the manufacturers or importers, batch
number or lot number and expiry date (if applicable) shall be maintained.
(6)     The records referred to in
sub-rule (5) shall be open to inspection by a Medical Device Officer appointed
under the sub-rule (2) of Rule 18, who may, if necessary, make enquiries about
purchases and sale of the medical devices and may also take samples for
testing.
(7)     All registers and records
mentioned under these rules, shall be preserved for a period of not less than
two years from the last entry, therein.
(8)     The registration certificate
holder shall maintain an inspection book in Form MD-43 to enable the Medical
Devices Officer to record his observations and defects noticed.][32] 
Rule - 87C. [Validity of registration
certificate.
(1)     A registration certificate
issued in Form MD-42, shall remain valid in perpetuity, subject to payment of
registration certificate retention fee as specified in the Second Schedule,
before completion of the period of five years from the date of its issue,
unless, it is suspended or cancelled by State Licensing Authority:
Provided that, if the
registration certificate holder fails to pay the required registration
certificate retention fee on or before due date, the registration certificate
holder shall, in addition to the registration certificate retention fee, be
liable to pay a late fee calculated at the rate of two per cent of the
registration certificate retention fee for every month or part thereof within
six months:
Provided further that in
the event of non-payment of such fee within the period referred to in the first
proviso, the registration certificate shall be deemed to have been cancelled.][33] 
Rule - 87D. [Suspension and cancellation of
Registration Certificate.
(1)     Where the registration
certificate holder contravenes any provision of the Act or these rules, the
State Licensing Authority, shall, after giving the registration certificate
holder an opportunity to show cause as to why such an order should not be
passed, by an order and for reasons to be recorded in writing, suspend it for
such period as it considers necessary either wholly or in respect of any of the
medical device or, as the case may be, cancel the registration certificate.
(2)     A registration certificate
holder whose registration certificate has been suspended or cancelled by the
State Licensing Authority under sub-rule (1), may within forty-five days of the
receipt of a copy of the order by such authority, prefer an appeal to the State
Government and the State Government, shall after giving the registration holder
an opportunity of being heard, confirm, reverse or modify such order, with
reasons to be recorded in writing.][34] 
Rule - 88. Supply of medical device to
hospitals against delivery challan.
(1)     Notwithstanding anything
contained in the Drugs and Cosmetics Rules, 1945, any person having a valid
licence to sell, stock, exhibit or offer for sale or distribute by retail or
wholesale [35][or
registration certificate in Form MD-42], may, supply invasive medical devices
to be implanted through surgical intervention to a hospital for its patient
against a delivery challan:
Provided that in respect of
supplies made against delivery challan of such medical devices, the licensee
shall ensure that specified storage conditions are met.
(2)     A cash or credit memo shall
be generated for such medical devices supplied under sub-rule (1), used in the
surgical intervention and record of the same shall be maintained by the
licensee as per condition of licence.
Rule - 89. Recall of medical device.
(1)     If a manufacturer or
authorised agent, as the case may be, considers or has reasons to believe that
a medical device, which has been imported, manufactured, sold or distributed,
is likely to pose risk to the health of a user or patient during its use and
therefore may be unsafe, such manufacturer or authorised agent shall
immediately initiate procedures to withdraw the medical device in question from
the market and patients, indicating reasons for its withdrawal and inform the
competent authority the details thereof.
(2)     A manufacturer or authorised
agent, as the case may be, shall immediately inform the competent authority and
cooperate with them, if there are reasons to believe that a medical device
which has been placed in the market, may be unsafe for the patients.
(3)     The manufacturer or importer
or authorised agent, as the case may be, shall inform the competent authority
of the action taken to prevent risk to the patient and shall not prevent or
discourage any person from cooperating, in accordance with the provisions of
the Act and these rules, with the competent authorities, where this may
prevent, reduce or eliminate a risk arising due to use of such medical device.
Chapter XII MISCELLANEOUS
Rule - 90. Exemption from provisions
related to medical devices.
(1)     The medical devices
specified in the Eighth Schedule shall be exempt from the provisions of these
rules to the extent and subject to the conditions specified in that Schedule.
(2)     The Central Government may,
by notification, from time to time, amend or modify the entries in the Eighth Schedule.
Rule - 91. Export of medical devices.
Where a person intends to
export any medical device, manufactured in India, and for that purpose,
requests a certificate in the nature of free sale certificate or a certificate
about quality, safety and performance in relation to that medical device as
required by the authority concerned of the importing country, such person, may
apply to the Central Licensing Authority [36][for
Class C and Class D medical devices and State Licensing Authority for Class A
and Class B medical devices] for the purpose along with a fee as specified in
the Second Schedule and the said authority shall, if the requirements are
fulfilled, issue a certificate to the applicant.
Rule - 92. Rejection of application.
If any document submitted
by an applicant for grant of licence for import or manufacture or test licence
or permit for personal use or permission to import or manufacture
investigational medical device or new in vitro diagnostic medical device or
permission to conduct of clinical investigation or clinical performance
evaluation is found to be misleading, or fake, or fabricated, the application,
after giving an opportunity to the applicant of being heard, shall be summarily
rejected.
Rule - 93. Debarment of applicant.
(1)     Whoever himself or, any
other person on his behalf, or applicant is found to be guilty of submitting
misleading, or fake, or fabricated documents, may, after giving him an
opportunity to show cause as to why such an order should not be made, in
writing, stating the reasons thereof, be debarred by the Central Licensing
Authority or the State Licensing Authority, as the case may be, for such period
as it may deem proper.
(2)     Where an applicant is
aggrieved by an order made by the Central Licensing Authority or the State Licensing
Authority, as the case may be, under sub-rule (1), such applicant may, within
thirty days of the receipt of the order, make an appeal to the Central
Government or the State Government, as the case may be, and that Government
may, after such enquiry as it considers necessary, and after affording an
opportunity of being heard, make such order as it may deem proper.
Rule - 94. Mode of payment of fee.
(1)     The fees prescribed under
these rules, in case of application made to the Central Licensing Authority,
shall be paid through challan or by electronic mode, in the Bank of Baroda,
Kasturba Gandhi Marg, New Delhi-110001 or any other branch of Bank of Baroda,
or any other bank, notified by the Ministry of Health and Family Welfare in the
Central Government, to be credited under the Head of Account “0210- Medical and
Public Health, 04-Public Health, 104-Fees and Fines.
(2)     Where the fee specified is
payable to the State Licensing Authority, the same shall be paid through a
challan or by electronic mode as may be specified by the State Government
concerned.
Rule - 95. Digitalisation of form.
The forms prescribed under
these rules may be suitably modified for conversion into digital forms by the
Central Drugs Standard Control Organisation and such modification shall not
require any amendment in these rules.
Rule - 96. Overriding effect.
The provisions of these
rules shall have effect, notwithstanding anything inconsistent therewith
contained in the Drugs and Cosmetics Rules, 1945.
Rule - 97. Savings.
Notwithstanding the
non-applicability of the Drugs and Cosmetics Rules, 1945, for the substances
and devices referred to in Rule 2,—
(i)       the licence or registration
certificate, issued under the provisions of the Act and the Drugs and Cosmetics
Rules, 1945, prior to commencement of these rules, shall be deemed to be valid
till its expiry or for a period of eighteen months from the date these rules are
notified, whichever is later, under the corresponding provisions of these
rules;
(ii)      new drug approval, or
things done or any action taken or purported to have been done or taken,
including any rule, notification, inspection, order or notice made or issued or
any appointment or declaration made or any operation undertaken or any
direction given or any proceedings taken or any penalty, punishment, forfeiture
or fine imposed under the Drugs and Cosmetics Rules, 1945 shall, be deemed to
have been done or taken under the corresponding provisions of these rules and
shall always remain valid for all purposes.
FIRST
SCHEDULE
[See Rule
4]
Parameters
for classification of medical devices and in vitro diagnostic medical devices
Part I Parameters for classification of medical devices other
than in vitro diagnostic medical devices
Basic Principles for
classification:
(i)       Application of the
classification provisions shall be governed by the intended purpose of the
device.
(ii)      If the device is intended
to be used in combination with another device, the classification rules shall
apply separately to each of the devices. Accessories are classified in their
own right separately from the device with which they are used.
(iii)     Software, which drives a
device or influences the use of a device, falls automatically in the same
class.
(iv)    If the device is not
intended to be used solely or principally in a specific part of the body, it
must be considered and classified on the basis of the most critical specified
use.
(v)      If several rules apply to
the same device, based on the performance specified for the device by the
manufacturer, the strictest rules resulting in the higher classification shall
apply.
(1)     Parameters for
classification of medical devices:
(i)       Non-invasive medical
devices which come into contact with injured skin.
(a)      A non-invasive medical
device which comes into contact with injured skin shall be assigned to Class A,
if it is intended to be used as a mechanical barrier, for compression or for
absorption of exudates only, for wounds which have not breached the dermis and
can heal by primary intention;
(b)      Subject to clause (c), a
non-invasive medical device which comes into contact with injured skin shall be
assigned to Class B, if it is intended to be used principally with wounds which
have breached the dermis, or is principally intended for the management of the
microenvironment of a wound;
(c)      a non-invasive medical
device which comes into contact with injured skin shall be assigned to Class C,
if it is intended to be used principally with wounds which have breached the
dermis and cannot heal by primary intention.
(ii)      Non-invasive medical
devices for channeling or storing substances.
(a)      Subject to clauses (b) and
(c), a non-invasive medical device shall be assigned to Class A, if it is
intended for channeling or storing body liquids or tissues or liquids or gases
for the purpose of eventual infusion, administration or introduction into a
human body;
(b)      A non-invasive medical
device referred to in clause (a) shall be assigned to Class B, if it is
intended to be connected to an active medical device which is in Class B, C or
D or for channeling blood or storing or channeling other body liquids or
storing organs, parts of organs or body tissues:
Provided, that the
circumstances when a non-invasive medical device is connected to an active
medical device include circumstances where the safety and performance of the
active medical device is influenced by the non-invasive medical device, or vice
versa; or
(c)      A non-invasive medical
device referred to in clause (a) shall be assigned to Class C, if it is a blood
bag that does not incorporate a medicinal product.
(iii)     Non-invasive medical
devices for modifying compositions of substances.
(a)      Subject to clause (b), a
non-invasive medical device shall be assigned to Class C, if it is intended for
modifying the biological or the chemical composition of blood or other body
liquids or other liquids intended for infusion into the body.
(b)      A non-invasive medical
device as referred to in clause (a) shall be assigned to Class B, if the
intended modification is carried out by filtration, centrifuging or any
exchange of gas or of heat.
(iv)    Other non-invasive medical
devices.
A non-invasive medical
device to which sub-paragraphs (i), (ii) and (iii) do not apply shall be
assigned to Class A, if it does not come into contact with a person or comes
into contact with intact skin only.
(v)      Invasive (body orifice)
medical devices for transient use.
(a)      Subject to clause (b), an
invasive (body orifice) medical device shall be assigned to Class A, if,—
(1)     it is intended for
transient use; and
(2)     it is not intended to be
connected to an active medical device; or
(3)     it is intended to be
connected to a Class A medical device only.
(b)      An invasive (body orifice)
medical device referred to in clause (a) shall be assigned to Class B, if,—
(1)     it is intended for use on
the external surface of an eyeball; or
(2)     it is liable to be absorbed
by the mucous membrane.
(vi)    Invasive (body orifice)
medical devices for short-term use.
(a)      Subject to clause (b), an
invasive (body orifice) medical device shall be assigned to Class B, if,—
(1)     it is intended for
short-term use; and
(2)     it is not intended to be
connected to an active medical device; or
(3)     it is intended to be
connected to a Class A medical device only.
(b)      An invasive (body orifice)
medical device referred to in clause (a) shall be assigned to Class A, if,—
(1)     it is intended for use in
an oral cavity as far as the pharynx or in an ear canal up to the ear drum or in
a nasal cavity; and
(2)     it is not liable to be
absorbed by the mucous membrane.
(vii)   Invasive (body orifice)
medical devices for long-term use.
(a)      Subject to clause (b), an
invasive (body orifice) medical device shall be assigned to Class C, if it is
intended for long-term use and, not intended to be connected to an active
medical device or it is to be connected to a Class A medical device only.
(b)      An invasive (body orifice)
medical device referred to in clause (a) shall be assigned to Class B, if,—
(1)     it is intended for use in
an oral cavity as far as the pharynx or in an ear canal up to the ear drum or
in a nasal cavity; and
(2)     it is not liable to be
absorbed by the mucous membrane.
(viii)  Invasive (body orifice)
medical devices for connection to active medical devices.
An invasive (body orifice)
medical device shall be assigned to Class B, regardless of the duration of its
use, if it is intended to be connected to an active medical device which is in
Class B, C or D.
(ix)    Surgically invasive medical
devices for transient use.
(a)      Subject to clauses (b) to
(g), a surgically invasive medical device intended for transient use shall be
assigned to Class B.
(b)      Subject to clauses (c) to
(g), a transient use surgically invasive medical device shall be assigned to
Class A, if it is a reusable surgical instrument.
(c)      A transient use surgically
invasive medical device shall be assigned to the same class as the active
medical device to which it is intended to be connected.
(d)      A transient use surgically
invasive medical device shall be assigned to Class C, if it is intended for the
supply of energy in the form of ionising radiation.
(e)      A transient use surgically
invasive medical device shall be assigned to Class C, if it is intended to have
a biological effect or to be wholly or mainly absorbed by the human body.
(f)       A transient use surgically
invasive medical device shall be assigned to Class C, if it is intended for the
administration of any medicinal product by means of a delivery system and such
administration is done in a manner that is potentially hazardous.
(g)      A transient use surgically
invasive medical device shall be assigned to Class D, if it is intended to be
used specifically in direct contact with the central nervous system or for the
diagnosis, monitoring or correction of a defect of the heart or of the central
circulatory system through direct contact with these parts of the body.
(x)      Surgically invasive medical
devices for short-term use.
(a)      Subject to clauses (b), (d)
and (e), a surgically invasive medical device intended for short-term use shall
be assigned to Class B.
(b)      Subject to clause (c), a
short-term use surgically invasive medical device shall be assigned to Class C,
if it is intended to undergo a chemical change in the body.
(c)      A short-term use surgically
invasive medical device referred to in clause (b) shall be assigned to Class B,
if it is intended to be placed into any tooth.
(d)      A short-term use surgically
invasive medical device shall be assigned to Class C, if it is intended for the
administration of any medicinal product or the supply of energy in the form of
ionising radiation.
(e)      A short-term use surgically
invasive medical device shall be assigned to Class D, if it is intended to have
a biological effect or to be wholly or mainly absorbed by the human body or to
be used specifically in direct contact with the central nervous system or for
the diagnosis, monitoring or correction of a defect of the heart or of the
central circulatory system through direct contact with these parts of the body.
(xi)    Implantable medical devices
and surgically invasive medical devices for long-term use.
(a)      Subject to clauses (b), (c)
and (d), an implantable medical device or a surgically invasive medical device
intended for long-term use shall be assigned to Class C.
(b)      A long-term use medical
device shall be assigned to Class B, if it is intended to be placed into any
tooth.
(c)      A long-term use medical
device shall be assigned to Class D, if it is intended,—
(1)     to be used in direct
contact with the heart, the central circulatory system or the central nervous
system;
(2)     to be life supporting or
life sustaining;
(3)     to be an active medical
device;
(4)     to be wholly or mainly
absorbed by the human body;
(5)     for the administration of
any medicinal product; or
(6)     to be a breast implant.
(d)      Subject to clause (b), a
long-term use medical device shall be assigned to Class D, if it is intended to
undergo chemical change in the body.
(xii)   Active therapeutic medical
devices for administering or exchanging energy.
(a)      Subject to clause (b), an
active therapeutic medical device shall be assigned to Class B, if it is
intended for the administration or exchange of energy to or with a human body.
(b)      An active therapeutic
medical device referred to in (a) shall be assigned to Class C, if the administration
or exchange of energy may be done in a potentially hazardous way (such as
through the emission of ionising radiation), taking into account the nature,
density and site of application of the energy and the type of technology
involved.
(c)      An active therapeutic
medical device shall be assigned to Class C, if it is intended for the control
or monitoring, or to be used to directly influence the performance, of a Class
C active therapeutic device.
(xiii)  Active diagnostic medical
devices.
(a)      Subject to clauses (b) and
(c), an active diagnostic medical device shall be assigned to Class B, if it is
intended,—
(1)     to be used to supply energy
which will be absorbed by the human body;
(2)     to be used to capture any
image of the in vivo distribution of radiopharmaceuticals; or
(3)     for the direct diagnosis or
monitoring of vital physiological processes.
(b)      An active diagnostic
medical device referred to in sub-clause (1) of clause (a) shall be assigned to
Class A, if it is intended to be used solely to illuminate a patient's body
with light in the visible or near infrared spectrum.
(c)      An active diagnostic
medical device referred to in clause (a) shall be assigned to Class C, if it is
intended specifically for,—
(1)     the monitoring of vital
physiological parameters, where the nature of any variation is such that it
could result in immediate danger to the patient (such as any variation in
cardiac performance, respiration or activity of the central nervous system); or
(2)     diagnosing in a clinical
situation where the patient is in immediate danger.
(d)      An active diagnostic
medical device shall be assigned to Class C, if it is intended for the emission
of ionising radiation and to be used in diagnostic or interventional radiology.
(e)      An active diagnostic
medical device shall be assigned to Class C, if it is intended for the control
or monitoring, or to be used to directly influence the performance, of any
active diagnostic medical device referred to in clause (d).
(f)       Subject to clause (g), an
active medical device shall be assigned to Class B, if it is intended for the
administration, or removal of, any medicinal product, body liquid or other
substance to or from a human body.
(g)      An active medical device
referred to in clause (f) shall be assigned to Class C, if the administration
or removal of the medicinal product, body liquid or other substance is done in
a manner that is potentially hazardous, taking into account,
(1)     the nature of the medicinal
product, body liquid or substance;
(2)     the part of the body
concerned; and
(3)     the mode and route of the
administration or removal.
(xiv)  Other active medical
devices.
An active medical device to
which provisions of sub-paragraphs (xii) and (xiii) do not apply shall be
assigned to Class A.
(xv)   Medical devices
incorporating medicinal products.
(a)      Subject to clause (b), a
medical device shall be assigned to Class D, if it incorporates as an integral
part a substance which,—
(1)     if used separately, may be
considered to be a medicinal product; and
(2)     is liable to act on a human
body with an action ancillary to that of the medical device.
(b)      A medical device referred
to in clause (a) shall be assigned to Class B, if the incorporated substance is
a medicinal product exempted from the licensing requirements of the Drugs and
Cosmetics Act, 1940 (23 of 1940) and the rules made thereunder.
(xvi)  Medical devices
incorporating animal or human cells, tissues or derivatives.
(a)      Subject to clause (b), a
medical device shall be assigned to Class D, if it is manufactured from or
incorporates,—
(1)     cells, tissues or
derivatives of cells or tissues, or any combination thereof, of animal or human
origin, which are or have been rendered non-viable; or
(2)     cells, tissues or
derivatives of cells or tissues, or any combination thereof, of microbial or
recombinant origin.
(b)      A medical device referred
to in clause (a) shall be assigned to Class A, if it is manufactured from or
incorporates non-viable animal tissues, or their derivatives, that come in
contact with intact skin only.
(xvii) Medical devices for
sterilisation or disinfection.
(a)      Subject to clause (b), a
medical device shall be assigned to Class C, if it is intended to be used
specifically for,—
(1)     the sterilisation of any
other medical device;
(2)     the end point disinfection
of any other medical device; or
(3)     the disinfection, cleaning,
rinsing or hydration of contact lenses.
(b)      A medical device shall be
assigned to Class B, if it is intended for the disinfection of any other
medical device before the latter is sterilised or undergoes end point
disinfection:
Provided, that “end-point
disinfection” means the disinfection of a medical device immediately before its
use by or on a patient.
(xviii)   Medical devices for
contraceptive use.
(a)      Subject to clause (b), a
medical device intended to be used for contraception or the prevention of the
transmission of any sexually transmitted disease shall be assigned to Class C.
(b)      A medical device referred
to in clause (a) shall be assigned to Class D, if it is an implantable medical
device or an invasive medical device intended for long-term use.
Part II Parameters for classification
for in vitro diagnostic medical devices
(1)     Basic principles for
classification of in vitro diagnostic medical devices:
(a)      Application of the
classification provisions shall be governed by the intended purpose of the
devices.
(b)      If the device is intended
to be used in combination with another device, the classification rules shall
apply separately to each of the devices. Accessories are classified in their
own right separately from the device with which they are used.
(c)      Software, which drives a
device or influences the use of a device, falls automatically in the same
class.
(d)      Stand-alone software, which
are not incorporated into the medical device itself and provide an analysis
based on the results from the analyser, shall be classified in to the same
category that of the in vitro diagnostic medical device where it controls or
influences the intended output of a separate in vitro diagnostic medical
device.
(e)      Subject to the clauses (c)
and (d), software that is not incorporated in an in vitro diagnostic medical
device, shall be classified using the classification provisions as specified in
Paragraph 2.
(f)       Calibrators intended to be
used with a reagent should be treated in the same class as the in vitro
diagnostic medical device reagent.
(g)      If several rules apply to
the same device, based on the performance specified for the device by the
manufacturer, the stringent rules resulting in the higher classification shall
apply.
(2)     The parameters for
classification of in vitro diagnostic medical devices as follows—
(i)       In vitro diagnostic medical
devices for detecting transmissible agents, etc.:
(a)      An in vitro diagnostic
medical device shall be assigned to Class D, if it is intended to be used for
detecting the presence of, or exposure to, a transmissible agent that,—
(1)     is in any blood, blood
component, blood derivative, cell, tissue or organ, in order to assess the
suitability of the blood, blood component, blood derivative, cell, tissue or
organ, as the case may be, for transfusion or transplantation; or
(2)     causes a life-threatening
disease with a high risk of propagation.
(b)      An in vitro diagnostic
medical device shall be assigned to Class C, if it is intended for use in,—
(1)     detecting the presence of,
or exposure to, a sexually transmitted agent;
(2)     detecting the presence in
cerebrospinal fluid or blood of an infectious agent with a risk of limited
propagation (for example, Cryptococcus neoformans or Neisseria meningitidis);
(3)     detecting the presence of
an infectious agent, where there is a significant risk that an erroneous result
will cause death or severe disability to the individual or foetus being tested
(for example, a diagnostic assay for Chlamydia pneumoniae, Cytomegalovirus or
Methicillin-resistant Staphylococcus aureus);
(4)     pre-natal screening of
women in order to determine their immune status towards transmissible agents
such as immune status tests for Rubella or Toxoplasmosis;
(5)     determining infective
disease status or immune status, where there is a risk that an erroneous result
will lead to a patient management decision resulting in an imminent
life-threatening situation for the patient being tested (for example,
Cytomegalovirus, Enterovirus or Herpes simplex virus in transplant patients);
(6)     screening for disease
stages, for the selection of patients for selective therapy and management, or
in the diagnosis of cancer;
(7)     human genetic testing, such
as the testing for cystic fibrosis or Huntington's disease;
(8)     monitoring levels of
medicinal products, substances or biological components, where there is a risk
that an erroneous result will lead to a patient management decision resulting
in an immediate life-threatening situation for the patient being tested (for
example, cardiac markers, cyclosporin or prothrombin time testing);
(9)     management of patients
suffering from a life-threatening infectious disease such as viral load of
Human immunodeficiency virus or Hepatitis C virus, or genotyping and sub-typing
Hepatitis C virus or Human immunodeficiency virus);or
(10)   screening for congenital
disorders in the foetus such as Down's syndrome or spina bifida.
(ii)      In vitro diagnostic medical
devices for blood grouping or tissue typing:
(a)      Subject to clause (b), an
in vitro diagnostic medical device shall be assigned to Class C, if it is
intended to be used for blood grouping or tissue typing to ensure the
immunological compatibility of any blood, blood component, blood derivative,
cell, tissue or organ that is intended for transfusion or transplantation, as
the case may be.
(b)      An in vitro diagnostic
medical device referred to in clause (a) shall be assigned to Class D, if it is
intended to be used for blood grouping or tissue typing according to the ABO
system, the, the Duffy system, the Kell system, the Kidd system, the rhesus
system (for example, HLA, Anti-Duffy, Anti-Kidd).
(iii)     In vitro diagnostic medical
devices for self-testing:
(a)      Subject to clause (b), an
in vitro diagnostic medical device shall be assigned to Class C, if it is
intended to be used for self-testing.
(b)      An in vitro diagnostic
medical device referred to in clause (a) shall be assigned to Class B, if it is
intended to be used to obtain,—
(1)     test results that are not
for the determination of a medically-critical status; or
(2)     preliminary test results
which require confirmation by appropriate laboratory tests.
(iv)    In vitro diagnostic medical
devices for near-patient testing:
An in vitro diagnostic
medical device shall be assigned to Class C, if it is to be used for
near-patient testing in a blood gas analysis or a blood glucose determination.
Illustration: Anticoagulant
monitoring, diabetes management, and testing for C-reactive protein and
Helicobacter pylori.
(v)      In vitro diagnostic medical
devices used in in vitro diagnostic procedures:
An in vitro diagnostic
medical device shall be assigned to Class A:
(1)     if it is a reagent or an
article which possesses any specific characteristic that is intended by its
product owner to make it suitable for an in vitro diagnostic procedure related
to a specific examination;
(2)     an instrument intended
specifically to be used for an in vitro diagnostic procedure; or
(3)     a specimen receptacle.
(vi)    Other in vitro diagnostic
medical devices:
(a)      An in vitro diagnostic
medical device shall be assigned to Class B, if sub-paragraphs (i) to (v) of
Paragraph 2 do not apply to it; or
(b)      It is a substance or device
used for the assessment of the performance of an analytical procedure or a part
thereof, without a quantitative or qualitative assigned value.
SECOND
SCHEDULE
[See Rules
13(5), 13(7), 20(2), 21(2), 25(3), 29(1), 31(1), 34(2), 34(4), 35(2), 37,
40(2), 42(1), 51(2), 59(2), 63(1), 64(1), 81(1), 84, [37][87A(3),
87C(1) and] 91]
Fee
payable for licence, permission and registration certificate
| 
   Sl. No.  | 
  
   Rule  | 
  
   Subject  | 
  
   In rupees (INR) except where specified in dollars
  ($)  | 
 
| 
   (1)  | 
  
   (2)  | 
  
   (3)  | 
  
   (4)  | 
 
| 
   1.  | 
  
   13(5)  | 
  
   Registration of Notified Body.  | 
  
   25000  | 
 
| 
   2.  | 
  
   13(7)  | 
  
   Registration retention fee of Notified Body.  | 
  
   25000  | 
 
| 
   3.  | 
  
   20(2)  | 
  
   Manufacturing licence or loan licence to
  manufacture [38][Class
  A (other than non-sterile and non-measuring)] or Class B medical device for,—  | 
  
   —  | 
 
| 
   4.  | 
  
      | 
  
   (a) one site; and  | 
  
   5000  | 
 
| 
   5.  | 
  
      | 
  
   (b) each distinct medical device.  | 
  
   500  | 
 
| 
   6.  | 
  
   21(2)  | 
  
   Manufacturing licence or loan licence to
  manufacture Class C or Class D medical device for,—  | 
  
   —  | 
 
| 
   7.  | 
  
      | 
  
   (a) one site; and  | 
  
   50000  | 
 
| 
   8.  | 
  
      | 
  
   (b) each distinct medical device.  | 
  
   1000  | 
 
| 
   9.  | 
  
   29(1)  | 
  
   Manufacturing licence or loan licence retention
  fee for,—  | 
  
   —  | 
 
| 
   10.  | 
  
      | 
  
   (a) one site manufacturing [39][Class
  A (other than non-sterile and non-measuring)] or Class B medical device; or  | 
  
   5000  | 
 
| 
   11.  | 
  
      | 
  
   (b) one site of manufacturing Class C or Class D
  medical device; or  | 
  
   50000  | 
 
| 
   12.  | 
  
      | 
  
   (c) each distinct medical device of [40][Class
  A (other than non-sterile and non-measuring)] or Class B; or  | 
  
   500  | 
 
| 
   13.  | 
  
      | 
  
   (d) each distinct medical device of Class C or
  Class D.  | 
  
   1000  | 
 
| 
   14.  | 
  
   31(1)  | 
  
   Test licence to manufacture for clinical
  investigations, test, evaluation, examination, demonstration or training for
  each distinct medical device.  | 
  
   500  | 
 
| 
   15.  | 
  
   34(2)  | 
  
   Import licence for [41][Class
  A (other than non-sterile and non-measuring)] medical device other than in
  vitro diagnostic medical device for,—  | 
  
      | 
 
| 
   16.  | 
  
      | 
  
   (a) one site; and  | 
  
   $1000  | 
 
| 
   17.  | 
  
      | 
  
   (b) each distinct medical device.  | 
  
   $50  | 
 
| 
   18.  | 
  
   34(2)  | 
  
   Import licence for Class B medical device other
  than in vitro diagnostic medical device for,—  | 
  
      | 
 
| 
   19.  | 
  
      | 
  
   (a) one site; and  | 
  
   $2000  | 
 
| 
   20.  | 
  
      | 
  
   (b) each distinct medical device.  | 
  
   $1000  | 
 
| 
   21.  | 
  
   34(2)  | 
  
   Import licence for [42][Class
  A (other than non-sterile and non-measuring)] or Class B in vitro diagnostic
  medical device for,—  | 
  
      | 
 
| 
   22.  | 
  
      | 
  
   (a) one site; and  | 
  
   $1000  | 
 
| 
   23.  | 
  
      | 
  
   (b) each distinct in vitro diagnostic medical
  device.  | 
  
   $10  | 
 
| 
   24.  | 
  
   34(2)  | 
  
   Import licence for Class C or Class D medical
  device other than in vitro diagnostic medical device for,—  | 
  
   —  | 
 
| 
   25.  | 
  
      | 
  
   (a) one site; and  | 
  
   $3000  | 
 
| 
   26.  | 
  
      | 
  
   (b) each distinct medical device.  | 
  
   $1500  | 
 
| 
   27.  | 
  
   34(2)  | 
  
   Import licence for Class C or Class D in vitro
  diagnostic medical device for,—  | 
  
      | 
 
| 
   28.  | 
  
      | 
  
   (a) one site; and  | 
  
   $3000  | 
 
| 
   29.  | 
  
      | 
  
   (b) each distinct in vitro diagnostic medical
  device.  | 
  
   $500  | 
 
| 
   30.  | 
  
   35(2)  | 
  
   Inspection of the overseas manufacturing site.  | 
  
   $6000  | 
 
| 
   31.  | 
  
   37  | 
  
   Import licence retention fee for,—  | 
  
   —  | 
 
| 
   32.  | 
  
      | 
  
   (a) one overseas site manufacturing [43][Class
  A (other than non-sterile and non-measuring)] medical device other than in
  vitro diagnostic medical device; or  | 
  
   $1000  | 
 
| 
   33.  | 
  
      | 
  
   (b) one overseas site manufacturing Class B
  medical device other than in vitro diagnostic medical device; or  | 
  
   $2000  | 
 
| 
   34.  | 
  
      | 
  
   (c) one overseas site manufacturing Class C or
  Class D medical device other than in vitro diagnostic medical device; or  | 
  
   $3000  | 
 
| 
   35.  | 
  
      | 
  
   (d) each distinct medical device of [44][Class
  A (other than non-sterile and non-measuring)] other than in vitro diagnostic
  medical device; or  | 
  
   $50  | 
 
| 
   36.  | 
  
      | 
  
   (e) each distinct medical device of Class B other
  than in vitro diagnostic medical device; or  | 
  
   $1000  | 
 
| 
   37.  | 
  
      | 
  
   (f) each distinct medical device of Class C or
  Class D other than in vitro diagnostic medical device.  | 
  
   $1500  | 
 
| 
   38.  | 
  
      | 
  
   (g) one overseas site manufacturing [45][Class
  A (other than non-sterile and non-measuring)] or Class B in vitro diagnostic
  medical device;  | 
  
   $1000  | 
 
| 
   39.  | 
  
      | 
  
   (h) one overseas site manufacturing Class C or
  Class D medical device other than in vitro diagnostic medical device;  | 
  
   $3000  | 
 
| 
   40.  | 
  
      | 
  
   (i) each distinct in vitro diagnostic medical
  device of [46][Class
  A (other than non-sterile and non-measuring)] or Class B in vitro diagnostic
  medical device;  | 
  
   $10  | 
 
| 
   41.  | 
  
      | 
  
   (j) each distinct in vitro diagnostic medical device
  of Class C or Class D in vitro diagnostic medical device;  | 
  
   $500  | 
 
| 
   42.  | 
  
   40(2)  | 
  
   Fee for Import licence for test, evaluation or
  demonstration or training for each distinct medical device.  | 
  
   $100  | 
 
| 
   43.  | 
  
   42(1)  | 
  
   Fee for Import of investigational medical device
  by Government hospital or statutory medical institution for treatment of
  patient of each distinct medical device.  | 
  
   500  | 
 
| 
   44.  | 
  
   51(2)(a)  | 
  
   Permission to conduct pilot clinical
  investigation.  | 
  
   100000  | 
 
| 
   45.  | 
  
   51(2)(b)  | 
  
   Permission to conduct pivotal clinical
  investigation.  | 
  
   100000  | 
 
| 
   46.  | 
  
   59(2)  | 
  
   Permission to conduct clinical performance
  evaluation.  | 
  
   25000  | 
 
| 
   47.  | 
  
   63(1)  | 
  
   Permission to import or manufacture a medical
  device which does not have its predicate device.  | 
  
   50000  | 
 
| 
   48.  | 
  
   64(1)  | 
  
   Permission to import or manufacture new in vitro diagnostic
  medical device.  | 
  
   25000  | 
 
| 
   49.  | 
  
   81(1)  | 
  
   Registration of medical device testing laboratory
  to carry out testing or evaluation of a medical device on behalf of
  manufacturer.  | 
  
   20000  | 
 
| 
   50.  | 
  
   84  | 
  
   Registration retention fee for medical device
  testing laboratory.  | 
  
   20000  | 
 
| 
   51.  | 
  
   91  | 
  
   Certificate to export of each [47][category
  of] medical device.  | 
  
   1000  | 
 
| 
   [48][52.  | 
  
   87A (3)  | 
  
   Registration certificate for sale of medical
  devices  | 
  
   3000  | 
 
| 
   53.  | 
  
   87C (1)  | 
  
   Retention fee for registration certificate for
  sale of medical devices  | 
  
   3000]  | 
 
THIRD
SCHEDULE
[See Rules
13(5), 13(9), 14, 15, 20(4), 20(6)]
Documents
required for registration of Notified Body, its duties and functions
Part I Documents to be
furnished along with application in Form MD-1 for grant of certificate of
registration
(1)     A Notified Body shall
furnish duly signed copy of the following documents to the Central Licensing
Authority.
(i)       Constitution details of the
Notified Body;
(ii)      Brief profile of the
organisation and business profile related to audit of medical device
manufacturing sites;
(iii)     Accreditation Certificate
issued by the National Accreditation Body referred to in the Rule 11.
(iv)    Quality manual of the
organisation;
(v)      List of all Standard
Operating Procedures;
(vi)    List of all technical
personnel including any outside experts along with their qualification,
experience and responsibilities.
(2)     Undertaking to be submitted
stating that the,—
(i)       Notified body including its
directors, executives and personnel responsible for carrying out evaluation and
verification activities shall not be the designer, manufacturer, supplier or
installer of devices within the product category for which the body has been
designated, nor the authorised representative of any of those parties.
(ii)      Directors, executives and
personnel responsible for carrying out evaluation and verification activities
shall be independent of both the manufacturers for whom the notified body
conducts assessments and the commercial competitors of those manufacturers,
during their employment by the notified body for the product range it is
notified for.
(iii)     Notified body personnel
shall not be involved in consultancy activities relating to devices in
question, their manufacturing control or test procedures, or their
manufacturer.
Part II Duties and
functions of Notified Body
1. Duties:
(1)     Notified body shall perform
the audit of manufacturer who applied under sub-rule (1) of Rule 13. The
specific application shall be allotted to the notified body by the State
Licensing Authority through the portal of the Central Government. The audit
shall relevant to domestic manufacturing site of Class A [49][(other
than non-sterile and non-measuring)] or Class B medical devices.
(2)     The notified body shall
have standard operating procedure for identification, review and resolution of
all cases where conflict of interest is suspected or proven. Record of such
review and decision shall be maintained.
2. Functions:
A notified body shall,—
(i)       impart training to its
staff covering all the evaluation and verification operations for which the
notified body has been designated;
(ii)      ensure that staff has
adequate knowledge and experience of the requirement of the control;
(iii)     carry out the evaluation
and verification operations with the highest degree of professional integrity
independently with technical competence;
(iv)    ensure that manufacturing site
and products comply with prescribed standards referred in Rule 7;
(v)      not provide training or
consultancy to the manufacturers whose site is being audited;
(vi)    ensure that their auditors
possess required qualification and expertise in the relevant field for carrying
out assessments of manufacturing site and medical device that they are
undertaking;
(vii)   establish and maintain
procedure and record which demonstrate its compliance with quality management
system.
3. Procedure for
audit:
The notified body shall
carry out the audit in the following manner,—
(i)       technical review of
respective documents as prescribed in the Fourth Schedule;
(ii)      on-site audit of the
manufacturer's quality management system to establish conformity by examination
of objective evidence, and that of sub-contractor wherever applicable, the
requirements of the Fifth Schedule;
(iii)     establish conformity by
examination and provision of objective evidences to the essential principles
laid down by the Central Government from time to time;
(iv)    establish design conformity
by review of the design documents during assessment of medical device to ensure
its quality, safety, and performance;
(v)      record post approval
changes, if any;
(vi)    assess conformity to the
product and process standards as per provisions of these rules;
(vii)   inform the manufacturer
about the observed non-compliances during audit, if any, and provide a copy of
the audit report to the manufacturer;
(viii)  when any major
non-compliance is observed during audit by the notified body which may affect
quality of the device, it may provide reasonable time to rectify the
non-compliance followed by compliance verification of the manufacturing site;
(ix)    The Notified Body, after
assessment and verification, shall submit detailed report giving its findings
on each aspect of audit along with its recommendations after completion of
audit to the State Licensing Authority and a copy of the same to the
manufacturer.
FOURTH
SCHEDULE
[See Rules
20(2), 21(2), 34(2), 34(4), 63(1) and 64(1)]
Documents
required for grant of licence to manufacture for sale or for distribution or
import
Part I POWER OF ATTORNEY
(To
be authenticated in India either by a Magistrate of First Class or by Indian
Embassy in the country of origin or by an equivalent authority through
apostille)
Power of Attorney to
accompany an application for issuance of import licence
I ………………… working as
……………………… authorised to sign this Power of Attorney, on behalf of M/s
………………………………… (full address/telephone no., e-mail) having manufacturing site at
……………………… (full address, telephone no., e-mail), hereby delegate Power of
Attorney to M/s………………….……, (full address, as per wholesale licence or
manufacturing licence [50][or
registration certificate], with telephone, fax and e-mail address), hereinafter
to be known as authorised agent, intends to apply for import licence under the
provisions of these rules, to import into India for the following medical
devices manufactured at below manufacturing site.
| 
   Sl. No.  | 
  
   Name and address of foreign manufacturer (full
  address with telephone, fax and e-mail address)  | 
  
   Name and address of manufacturing site (full
  address with telephone, fax and e-mail address of the manufacturing site)  | 
 
| 
      | 
  
      | 
  
      | 
 
| 
      | 
  
      | 
  
      | 
 
Following are the details
of medical device proposed to be imported (A separate list may be annexed, if
required in below given format).
| 
   Sl. No.  | 
  
   Generic Name  | 
  
   Brand Name (if any)  | 
  
   Model No. (if any)  | 
  
   Dimension  | 
  
   Intended Use  | 
  
   Shelf life  | 
  
   Sterile or Non-sterile  | 
  
   Class of medical device  | 
 
| 
   (2)  | 
  
   Our Authorised agent shall,—  | 
 
| 
   (a)  | 
  
   act as the official representative
  for obtaining import licence in India.  | 
 
| 
   (b)  | 
  
   submit all necessary documents, as
  defined in the Fourth Schedule, for the import licence of medical device.  | 
 
| 
   (3)  | 
  
   I shall comply with all the
  conditions imposed on the import licence and with provisions of the Medical
  Devices Rules, 2017.  | 
 
| 
   (4)  | 
  
   I declare that M/s ………… is carrying
  on the manufacture of the listed medical device at the manufacturing site
  specified above.  | 
 
| 
   (5)  | 
  
   I shall allow the Central Licensing
  Authority or any person authorised by it in that behalf to enter and inspect
  or audit the manufacturing premise and to examine the process, procedure and
  documents in respect of any manufacturing site or to take sample of listed
  medical device for which the application for import licence has been made.  | 
 
| 
   (6)  | 
  
   (6) In case of any violation of Drugs
  and Cosmetics Act, 1940 (23 of 1940) and the Medical Devices Rules, 2017, the
  authorised agent shall continue to be responsible even after withdraw of this
  Power of Attorney for the devices imported in India.  | 
 
| 
   (7)  | 
  
   I do hereby state and declare that
  all the photocopies or scanned copies in the application are true copies of
  the original documents.  | 
 
| 
   (8)  | 
  
   I do hereby state and declare that
  all the documents submitted by the undersigned are true and correct.  | 
 
| 
   Place:  | 
  
      | 
 
| 
   Date:  | 
  
   Signature of the manufacturer  | 
 
| 
      | 
  
   (Name and Designation)  | 
 
| 
      | 
  
   Seal/Stamp  | 
 
Undertaking
from the authorised agent
I …………………, age………., working
as ………………………… at M/s ……………………………… (Full address/telephone no., e-mail) agrees
to act upon the Power of Attorney as the authorised agent on behalf of M/s
………………………………… (Full address/telephone no., e-mail) having manufacturing site at
……………………… (Full address, telephone no., e-mail).
| 
   Place:  | 
  
      | 
 
| 
   Date:  | 
  
   Signature of the authorised agent  | 
 
| 
      | 
  
   (Name and Designation)  | 
 
| 
      | 
  
   Seal/Stamp  | 
 
Part II
| 
   (i)  | 
  
   Documents to be submitted with the
  application for grant of Import Licence or licence to manufacture for sale or
  for distribution of a Class A [51][(other
  than non-sterile and non-measuring)] medical device,—  | 
 
| 
   (a)  | 
  
   The applicant shall submit documents
  as specified in the table below—  | 
 
| 
   Sl. No.  | 
  
   For medical devices other than in vitro
  diagnostic medical device  | 
  
   For in vitro diagnostic medical device  | 
 
| 
   1.  | 
  
   device description, intended use of the device,
  specification including variants and accessories;  | 
  
   device description, intended use of the device,
  specification including variants and accessories;  | 
 
| 
   2.  | 
  
   material of construction;  | 
  
   a summary of analytical technology, relevant
  analytes and test procedure;  | 
 
| 
   3.  | 
  
   working principle and use of a novel technology
  (if any);  | 
  
   working principle and use of a novel technology
  (if any);  | 
 
| 
   4.  | 
  
   labels, package inserts ([52][instructions
  for use or electronic instructions for use], etc.,), user manual, wherever
  applicable,  | 
  
   labels and package inserts ([53][instructions
  for use or electronic instructions for use], etc.,), user manual, wherever
  applicable;  | 
 
| 
   5.  | 
  
   summary of any reported Serious Adverse Event in
  India or in any of the countries where device is marketed and action taken by
  the manufacturer and National Regulatory Authority concerned;  | 
  
   analytical performance summary including
  sensitivity and specificity;  | 
 
| 
   6.  | 
  
   site or plant master file as specified in
  Appendix I of this Schedule;  | 
  
   site or plant master file as specified in
  Appendix I of this Schedule;  | 
 
| 
   7.  | 
  
   constitution details of the firm (of domestic
  manufacturer or authorised agent);  | 
  
   constitution details of the firm (of domestic
  manufacturer or authorised agent);  | 
 
| 
   8.  | 
  
   essential principles checklist for demonstrating
  conformity to the essential principles of safety and performance of the
  medical device;  | 
  
   essential principles checklist for demonstrating
  conformity to the essential principles of safety and performance of the in
  vitro medical device;  | 
 
| 
   9.  | 
  
   undertaking signed by the manufacturer stating
  that the manufacturing site is in compliance with the provisions of the Fifth
  Schedule;  | 
  
   undertaking signed by the manufacturer stating
  that the manufacturing site is in compliance with the provisions of the Fifth
  Schedule;  | 
 
| 
   (b)  | 
  
   In case of application for import
  licence, the authorised agent shall submit—  | 
 |
| 
      | 
  
   A.  | 
  
   notarised copy of overseas
  manufacturing site or establishment or plant registration, by whatever name
  called, in the country of origin issued by the competent authority and Free
  Sale Certificate issued by the National Regulatory Authority or equivalent
  competent authority of the country concerned as referred under Rule 36.  | 
 
| 
      | 
  
   B.  | 
  
   notarised copy of Quality Management
  System certificate or Full Quality Assurance certificate or Production
  Quality Assurance certificate issued by the competent authority, in respect
  of the manufacturing site.  | 
 
| 
      | 
  
   C.  | 
  
   self-attested copy of valid whole
  sale licence or manufacturing licence issued under these rules.  | 
 
| 
      | 
  
   D.  | 
  
   copy of latest inspection or audit
  report carried out by Notified bodies or National Regulatory Authority or
  Competent Authority within last 3 years, if any.  | 
 
| 
   (ii)  | 
  
   Documents to be submitted with the application
  for grant of licence to manufacture or import Class B, Class C or Class D
  medical device,—  | 
 
| 
      | 
  
   The domestic manufacturer or
  authorised agent shall submit the duly signed following information
  pertaining to Manufacturing site.  | 
 
| 
      | 
  
   (a)  | 
  
   Constitution details of domestic
  manufacturer or authorised agent;  | 
 
| 
      | 
  
   (b)  | 
  
   Site or plant master file as
  specified in Appendix I of this Schedule;  | 
 
| 
      | 
  
   (c)  | 
  
   Device master file as specified in
  Appendix II for medical devices other than in vitro diagnostic medical
  devices, or Appendix III for in vitro diagnostic medical devices of this
  Schedule;  | 
 
| 
      | 
  
   (d)  | 
  
   Essential Principles checklist for
  demonstrating conformity to the Essential Principles of Safety and
  Performance of the Medical Device including in vitro diagnostic medical
  device;  | 
 
| 
      | 
  
   (e)  | 
  
   Test licence obtained for testing and
  generation of quality control data (for domestic manufacturers only), if any;  | 
 
| 
      | 
  
   (f)  | 
  
   Undertaking signed stating that the
  manufacturing site is in compliance with the provisions of the Fifth
  Schedule.  | 
 
| 
      | 
  
   (g)  | 
  
   Documents as specified in the clause
  (b) of paragraph (i) of this part.  | 
 
| 
      | 
  
   [54][(h)  | 
  
   In case of in-vitro diagnostic
  medical devices, performance evaluation report by the manufacturer shall be
  submitted by the applicant: Provided that when the State
  Licensing Authority specifically requires for Class B or the Central Licence
  Authority for Class B, Class C and Class D in-vitro diagnostic medical
  devices, as the case may be, applicant shall submit the report issued by the central
  medical devices testing laboratory or a medical device testing laboratory
  registered under Rule 83 or by any laboratory accredited by the National
  Accreditation Board for Testing and Calibration Laboratories or by any
  hospital accredited by National Accreditation Board for Hospitals and
  Healthcare Providers or by any Central Government or State Government
  Laboratory of any hospital or of any institute, specified by the concerned
  State Licensing Authority or the Central Licensing Authority.]  | 
 
Part III
APPENDIX
I
Contents
of a site or plant master file
The manufacturer shall
prepare a succinct document in the form of site master file containing specific
information about the production and/or control of device manufacturing carried
out at the premises. It shall contain the following information,—
1. General
Information:
(i)          
brief
information on the site (including name and address), relation to other sites;
(ii)        
manufacturing
activities;
(iii)      
any
other operations carried out on the site;
(iv)      
name
and exact address of the site, including telephone, fax numbers, web site URL
and e-mail address;
(v)       
type
of medical devices handled on the site and information about specifically toxic
or hazardous substances handled, mentioning the way they are handled and
precautions taken;
(vi)      
short
description of the site (size, location and immediate environment and other
activities on the site);
(vii)    
number
of employees engaged in production, quality control, warehousing, and
distribution;
(viii)  
use
of outside scientific, analytical or other technical assistance in relation to
the design, manufacture and testing;
(ix)      
short
description of the quality management system of the company;
(x)       
devices
details registered with foreign countries;
(xi)      
brief
description of testing facility.
2. Personnel:
(i)          
organisation
chart showing the arrangements for key personnel;
(ii)        
qualifications,
experience and responsibilities of key personnel;
(iii)      
outline
of arrangements for basic and in-service training and how records are
maintained;
(iv)      
health
requirements for personnel engaged in production;
(v)       
personnel
hygiene requirements, including clothing.
3. Premises and
Facilities:
(i)       layout of premises with
indication of scale;
(ii)      nature of construction,
finishes/fixtures and fittings;
(iii)     brief description of
ventilation systems. More details should be given for critical areas with
potential risks of airborne contamination (including schematic drawings of the
systems). Classification of the rooms used for the manufacture of sterile
products should be mentioned;
(iv)    special areas for the
handling of highly toxic, hazardous and sensitising materials;
(v)      brief description of water
systems (schematic drawings of the systems are desirable) including sanitation;
(vi)    maintenance (description of
planned preventive maintenance programmes for premises and recording system);
4. Equipment:
(i)          
Brief
description of major production and quality control laboratories equipment (a
list of the equipment is required);
(ii)        
maintenance
(description of planned preventive maintenance programmes and recording system);
(iii)      
qualification
and calibration, including the recording system. Arrangements for computerised
systems validation.
5. Sanitation:
Availability of written
specifications and procedures for cleaning the manufacturing areas and
equipment.
6. Production:
(i)       Brief description of
production operations using, wherever possible, flow sheets and charts
specifying important parameters;
(ii)      arrangements for the
handling of starting materials, packaging materials, bulk and finished
products, including sampling, quarantine, release and storage;
(iii)     arrangements for
reprocessing or rework;
(iv)    arrangements for the
handling of rejected materials and products;
(v)      brief description of
general policy for process validation;
(vi)    Brief description of
sterilisation facility.
7. Quality Assurance:
Description of the quality
assurance system and of the activities of the quality assurance department.
Procedures for the release of finished products.
8. Storage:
Policy on the storage of
medical device.
9. Documentation:
Arrangements for the
preparation, revision and distribution of necessary documentation, including
storage of master documents.
10. Medical Device
Complaints and Field Safety Corrective Action:
(i)       Arrangements for the
handling of complaints;
(ii)      Arrangements for the handling
of field safety corrective action.
11. Internal Audit:
Short Description of the
internal audit system.
12. Contract
Activities:
Description of the way in
which the compliance of the contract acceptor is assessed.
APPENDIX
II
Device
Master File for Medical Devices other than in Vitro Diagnostic Medical Devices
| 
   EXECUTIVE SUMMARY:  | 
 |
| 
   1.  | 
  
   An executive summary shall be
  provided by the manufacturer and shall contain:  | 
 
| 
   1.1  | 
  
   Introductory descriptive information
  on the medical device, the intended use and indication for use, class of
  device, novel features of the device (if any), shelf-life of the device and a
  synopsis on the content of the dossier.  | 
 
| 
   1.2  | 
  
   Information regarding sterilisation
  of the device (whether it is sterile or non-sterile; if sterile, mode of
  sterilisation).  | 
 
| 
   1.3  | 
  
   Risk Management Plan, Risk Analysis,
  evaluation and control documents.  | 
 
| 
   1.4  | 
  
   Clinical Evidence and evaluation (if
  applicable).  | 
 
| 
   1.5  | 
  
   Regulatory status of the similar
  device in India (approved or not approved in India).  | 
 
| 
   1.6  | 
  
   Design Examination Certificate,
  Declaration of Conformity, Mark of Conformity Certificate, Design Certificate
  (if applicable). Copy of such certificate(s) shall be enclosed.  | 
 
| 
   1.7  | 
  
   Marketing history of the device from
  the date of introducing the device in the market.  | 
 
| 
   1.8  | 
  
   Domestic price of the device in the
  currency followed in the country of origin.  | 
 
| 
   1.9  | 
  
   List of regulatory approvals or
  marketing clearance obtained (submit respective copies of approval
  Certificates):  | 
 
| 
   Country  | 
  
   Approved Indication  | 
  
   Approved Shelf life  | 
  
   Class of Device  | 
  
   Date of First Approval  | 
 
| 
   USA  | 
  
      | 
  
      | 
  
      | 
  
      | 
 
| 
   Australia  | 
  
      | 
  
      | 
  
      | 
  
      | 
 
| 
   Japan  | 
  
      | 
  
      | 
  
      | 
  
      | 
 
| 
   Canada  | 
  
      | 
  
      | 
  
      | 
  
      | 
 
| 
   European Union  | 
  
      | 
  
      | 
  
      | 
  
      | 
 
| 
   Others[55]  | 
  
      | 
  
      | 
  
      | 
  
      | 
 
Status
of market clearance pending, rejected or withdrawn
| 
   Regulatory Agency of the country  | 
  
   Indication for use  | 
  
   Registration status and date  | 
  
   Reason for rejection/withdrawal, if any  | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
 
| 
   1.10  | 
  
   Safety and performance related
  information on the device:  | 
 |
| 
      | 
  
   (a)  | 
  
   Summary of reportable event and field
  safety corrective action from the date of introduction—  | 
 
For Serious Adverse Event:
| 
   Sl. No.  | 
  
   Serious Adverse Event (SAE)  | 
  
   Duration  | 
  
   Number of the SAE reported  | 
  
   Total Units sold  | 
  
   Lot/Batch No.  | 
 |
| 
      | 
  
      | 
  
   From  | 
  
   To  | 
  
      | 
  
      | 
  
      | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
 
For Field Safety Corrective
Action (FSCA):
| 
   Date of FSCA  | 
  
   Reason for FSCA  | 
  
   Countries where FSCA was conducted  | 
  
   Description of the action taken  | 
 
| 
      | 
  
   (b)  | 
  
   If the device contains any of the
  followings, then descriptive information on the following need to be provided.  | 
 
| 
      | 
  
   1.  | 
  
   Animal or human cells tissues or
  derivatives thereof, rendered non-viable (e.g. Porcine Heart Valves).  | 
 
| 
      | 
  
   2.  | 
  
   Cells, tissues or derivatives of
  microbial recombinant origin (e.g. Dermal fillers based on Hyaluronic acid
  derived from bacterial fermentation process).  | 
 
| 
      | 
  
   3.  | 
  
   Irradiating components, ionising or
  non-ionising.  | 
 
| 
   2.  | 
  
   Device description and product
  specification, including variants and accessories  | 
 |
| 
   2.1  | 
  
   The dossier should contain the
  following descriptive information for the device—  | 
 |
| 
      | 
  
   (a)  | 
  
   A general description including its
  generic name, model name, model no., materials of construction, intended use,
  indications, [56][instructions
  for use or electronic Instructions for use], contraindications, warnings,
  precautions and potential adverse effects;  | 
 
| 
      | 
  
   (b)  | 
  
   the intended patient population and
  medical condition to be diagnosed or treated and other considerations such as
  patient selection criteria;  | 
 
| 
      | 
  
   (c)  | 
  
   principle of operation or mode of
  action, accompanies by animation or videos (if available);  | 
 
| 
      | 
  
   (d)  | 
  
   an explanation of any novel features;  | 
 
| 
      | 
  
   (e)  | 
  
   a description of the accessories,
  other medical device and other product that are not medical device, which are
  intended to be used in combination with it and it should also be clarified
  whether these accessories or device are supplied as a system or separate
  components;  | 
 
| 
      | 
  
   (f)  | 
  
   a description or complete list of the
  various configurations or variants of the device that will be made available;  | 
 
| 
      | 
  
   (g)  | 
  
   a general description of the key
  functional elements e.g. its parts or components (including software if
  appropriate), its formulation, its composition, its functionality and where
  appropriate, this will include: labeled pictorial representations (e.g.
  diagrams, photographs, and drawings), clearly indicating key parts or
  components, including sufficient explanation to understand the drawings and
  diagrams;  | 
 
| 
      | 
  
   (h)  | 
  
   a description of the materials
  incorporated into key functional elements and those making either direct
  contact with a human body or indirect contact with the body e.g. during
  extracorporeal circulation of body fluids. Complete chemical, biological and
  physical characterisation of the material(s) of the medical device;  | 
 
| 
      | 
  
   (i)  | 
  
   for medical devices intended to emit
  ionizing radiation, information on radiation source (e.g. radioisotopes) and
  the material used for shielding of unintended, stray or scattered radiation
  from patients, users and other persons shall be provided.  | 
 
| 
   2.2  | 
  
   Product Specification:  | 
 |
| 
      | 
  
   The dossier should contain a list of
  the features, dimensions and performance attributes of the medical device,
  its variants and accessories, that would typically appear in the product
  specification made available to the end user e.g. in brochures, catalogues,
  etc.  | 
 |
| 
   2.3  | 
  
   Reference to predicate or previous
  generations of the device:  | 
 |
| 
      | 
  
   Where relevant to demonstrating
  conformity to the essential principles, and to the provision of general
  background information, the dossier should contain an overview of:  | 
 |
| 
      | 
  
   (a)  | 
  
   the manufacturer's previous
  generation of the device, if such exist;  | 
 
| 
      | 
  
   (b)  | 
  
   predicate devices available on the
  local and international markets; and  | 
 
| 
      | 
  
   (c)  | 
  
   comparative analysis to prove
  substantial equivalence to the predicate device(s) as claimed.  | 
 
| 
   3.  | 
  
   LABELLING:  | 
 |
| 
      | 
  
   The dossier should typically contain
  a complete set of labeling associated with the device as per the requirements
  of Chapter VI of these rules. Information on labelling should include the
  following—  | 
 |
| 
      | 
  
   (a)  | 
  
   Copy of original label of the device,
  including accessories if any, and its packaging configuration;  | 
 
| 
      | 
  
   [57][(b)  | 
  
   Instructions for use or electronic
  instructions for use (Prescriber's manual);]  | 
 
| 
      | 
  
   (c)  | 
  
   Product brochure; and  | 
 
| 
      | 
  
   (d)  | 
  
   Promotional material.  | 
 
| 
   4.  | 
  
   DESIGN AND MANUFACTURING INFORMATION:  | 
 |
| 
      | 
  
   4.1  | 
  
   Device Design:  | 
 
| 
      | 
  
      | 
  
   The dossier should contain
  information to allow the reviewer to obtain a general understanding of the
  design stages applied to the device. The information may take in form of flow
  chart. Device design validation data should be submitted.  | 
 
| 
      | 
  
   4.2  | 
  
   Manufacturing Processes:  | 
 
| 
      | 
  
      | 
  
   The dossier should contain
  information to allow the reviewer to obtain a general understanding of the
  manufacturing processes. The information may take the form of flow chart
  showing an overview of production, manufacturing environment, facilities and
  controls used for manufacturing, assembly, any final product testing,
  labelling and packaging and storage of the finished medical device. If the
  manufacturing process is carried out at multiple sites, the manufacturing
  activities at each site should be clearly specified.  | 
 
| 
   5.  | 
  
   ESSENTIAL PRINCIPLES CHECKLIST:  | 
 ||
| 
      | 
  
   (i)  | 
  
   The dossier should contain the
  following—  | 
 |
| 
      | 
  
      | 
  
   (a)  | 
  
   the essential principles;  | 
 
| 
      | 
  
      | 
  
   (b)  | 
  
   whether each essential principle
  applies to the device and if not, why not;  | 
 
| 
      | 
  
      | 
  
   (c)  | 
  
   the method used to demonstrate
  conformity with each essential principle that applies;  | 
 
| 
      | 
  
      | 
  
   (d)  | 
  
   a reference for the method employed
  (e.g. standard); and  | 
 
| 
      | 
  
      | 
  
   (e)  | 
  
   the precise identity of the
  controlled document that offers evidence of conformity with each method used.  | 
 
| 
      | 
  
   (ii)  | 
  
   Methods used to demonstrate
  conformity may include one or more of the following:  | 
 |
| 
      | 
  
      | 
  
   (a)  | 
  
   conformity with standards as referred
  to in Rule 7;  | 
 
| 
      | 
  
      | 
  
   (b)  | 
  
   conformity with an in-house test method;  | 
 
| 
      | 
  
      | 
  
   (c)  | 
  
   the evaluation of pre-clinical and
  clinical evidence;  | 
 
| 
      | 
  
      | 
  
   (d)  | 
  
   comparison to a similar device
  already available on the market.  | 
 
| 
      | 
  
   (iii)  | 
  
   The essential principles checklist
  should incorporate a cross-reference to the location of such evidence both
  within the full technical documentation held by the manufacturer and within
  the dossier. A template for a checklist is shown in as under:  | 
 |
| 
   Essential Principle  | 
  
   Relevant Yes/No  | 
  
   Specification/standard Sub-clause/reference  | 
  
   Complies Yes/No  | 
  
   Document Reference Justification and/or comments  | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
  
      | 
 
| 
   6.  | 
  
   Risk analysis and control summary:  | 
 
| 
      | 
  
   The dossier should contain a summary
  of the risks identified during the risk analysis process and how these risks
  have been controlled to an acceptable level. This risk analysis should be
  based on prescribed standards and be part of the manufacturer's risk
  management plan based on complexity and risk class of the device. The
  technique used to analyse the risk must be specified, to ensure that it is
  appropriate for the medical device and risk involved. The risks and benefits
  associated with the use of the medical device should be described. The risk
  analysis submitted shall have periodic updation of the risks identified as
  per risk management plan.  | 
 
| 
   7.  | 
  
   Verification and validation of the
  medical device  | 
 
| 
   7.1  | 
  
   General:  | 
 ||
| 
      | 
  
   (A)  | 
  
   The dossier should contain product
  verification and validation documentation. The dossier should summarise the
  results of verification and validation studies undertaken to demonstrate
  conformity of the device with the essential principles that apply to it. Such
  information would typically cover wherever applicable:  | 
 |
| 
      | 
  
      | 
  
   (a)  | 
  
   engineering tests;  | 
 
| 
      | 
  
      | 
  
   (b)  | 
  
   laboratory tests;  | 
 
| 
      | 
  
      | 
  
   (c)  | 
  
   simulated use testing;  | 
 
| 
      | 
  
      | 
  
   (d)  | 
  
   any animal tests for demonstrating
  feasibility or proof of concept of the finished device;  | 
 
| 
      | 
  
      | 
  
   (e)  | 
  
   any published literature regarding
  the device or substantially similar devices.  | 
 
| 
      | 
  
   (B)  | 
  
   Such summary information may include:  | 
 |
| 
      | 
  
      | 
  
   (i)  | 
  
   declaration or certificate of
  conformity to a recognised standard and summary of the data if no acceptance
  criteria are specified in the standard;  | 
 
| 
      | 
  
      | 
  
   (ii)  | 
  
   declaration or certificate of
  conformity to a published standard that has not been recognised, supported by
  a rationale for its use, and summary of the data if no acceptance criteria is
  specified in the standard;  | 
 
| 
      | 
  
      | 
  
   (iii)  | 
  
   declaration or certificate of
  conformity to a professional guideline, industry method, or in-house test
  method, supported by a rationale for its use, a description of the method
  used, and summary of the data in sufficient detail to allow assessment of its
  adequacy;  | 
 
| 
      | 
  
      | 
  
   (iv)  | 
  
   a review of published literature
  regarding the device or substantially similar devices.  | 
 
| 
      | 
  
   (C)  | 
  
   In addition, where applicable to the
  device, the dossier should contain detailed information on:  | 
 |
| 
      | 
  
      | 
  
   (a)  | 
  
   bio-compatibility studies data as per
  prescribed standards;  | 
 
| 
      | 
  
      | 
  
   (b)  | 
  
   medicinal substances incorporated
  into the device, including compatibility of the device with the medicinal
  substance;  | 
 
| 
      | 
  
      | 
  
   (c)  | 
  
   biological safety of devices
  incorporating animal or human cells, tissues or their derivatives;  | 
 
| 
      | 
  
      | 
  
   (d)  | 
  
   sterilisation;  | 
 
| 
      | 
  
      | 
  
   (e)  | 
  
   software verification and validation;  | 
 
| 
      | 
  
      | 
  
   (f)  | 
  
   animal studies that provide direct
  evidence of safety and performance of the device, especially when no clinical
  investigation of the device was conducted;  | 
 
| 
      | 
  
      | 
  
   (g)  | 
  
   clinical evidence.  | 
 
| 
      | 
  
   (D)  | 
  
   Detailed information will describe
  test design, complete test or study protocols, methods of data analysis, in
  addition to data summaries and test conclusions. Where no new testing has
  been undertaken, the dossier should incorporate a rationale for that decision,
  e.g. bio-compatibility testing on the identical materials was conducted when
  these were incorporated in a previous, legally marketed version of the
  device. The rationale may be incorporated into the Essential Principle
  checklist.  | 
 |
| 
   7.2  | 
  
   Bio-compatibility:  | 
 |
| 
      | 
  
   (i)  | 
  
   The dossier should contain a list of
  all materials in direct or indirect contact with the patient or user.  | 
 
| 
      | 
  
   (ii)  | 
  
   Where bio-compatibility testing has
  been undertaken (as per prescribed standards) to characterize the physical,
  chemical, toxicological and biological response of a material, detailed
  information should be included on the tests conducted, standards applied,
  test protocols, the analysis of data and the summary of results. At a
  minimum, tests should be conducted on samples from the finished, sterilised
  (when supplied sterile) device.  | 
 
| 
      | 
  
   (iii)  | 
  
   Depending on nature and intended use
  of the investigational medical device, device performance for its actions
  (including mechanical, electrical, thermal, radiation and any other of this
  type) and safety should be assessed in healthy or diseased animal model
  (intended to be treated by such medical device), as appropriate,
  demonstrating reaction to active and basic parts of the devices on absolute
  tissue, local tissue as well as whole organ, clearly recording local, general
  and systemic adverse reactions, risks or potential risks and performance of
  device in line with intended use. Wherever possible, histopathology,
  pathophysiology and path anatomy should be carried out.  | 
 
| 
      | 
  
   (iv)  | 
  
   ISO-10993, Biological Evaluation of
  Medical Devices, should be followed for conducting bio-compatibility study
  for invasive medical devices should be carried out. A report of
  bio-compatibility study along with rationale for selecting specific tests
  carried out should be prepared including conclusion of the study.  | 
 
| 
   7.3  | 
  
   Medicinal substances:  | 
 
| 
      | 
  
   Where the medical device incorporates
  a medicinal substance, the dossier should provide detailed information
  concerning that medicinal substance, its identity and source, the intended
  reason for its presence, and its safety and performance in the intended
  application.  | 
 
| 
   7.4  | 
  
   Biological safety:  | 
 |
| 
      | 
  
   (i)  | 
  
   The dossier should contain a list of
  all materials of animal or human origin used in the device. For these
  materials, detailed information should be provided concerning the selection
  of sources or donors; the harvesting, processing, preservation, testing and handling
  of tissues, cells and substances of such origin should also be provided.
  Process validation results should be included to substantiate that
  manufacturing procedures are in place to minimise biological risks, in
  particular, with regard to viruses and other transmissible agents.
  Transmissible Spongiform Encephalopathies (TSE) or Bovine Spongiform
  Encephalopathy (BSE) Certificates should also be submitted: [58][Provided that the
  requirement of Transmissible Spongiform Encephalopathies (TSEs) or Bovine
  Spongiform Encephalopathy (BSE) Certificates is not necessary, if the source
  is from an animal species from a country of origin recognised as having
  negligible Bovine Spongiform Encephalopathy risk in accordance with the
  recommendations of the World Organisation for Animal Health.]  | 
 
| 
      | 
  
   (ii)  | 
  
   The system for record keeping to
  allow traceability from sources to the finished device should be fully
  described.  | 
 
| 
   7.5  | 
  
   Sterilisation:  | 
 |
| 
      | 
  
   (i)  | 
  
   Where the device is supplied sterile,
  the dossier should contain the detailed information of the initial
  sterilisation validation including steriliser qualification, bio-burden
  testing, pyrogen testing, testing for sterilant residues (if applicable) and
  packaging validation as per prescribed standards. Typically, the detailed
  validation information should include the method used, sterility assurance level
  attained, standards applied, the sterilisation protocol developed in
  accordance with prescribed standards, and a summary of results.  | 
 
| 
      | 
  
   (ii)  | 
  
   Evidence of the ongoing revalidation
  of the process should also be provided. Typically this would consist of
  arrangements for, or evidence of, revalidation of the packaging and
  sterilisation processes.  | 
 
| 
   7.6  | 
  
   Software verification and validation:  | 
 
| 
      | 
  
   The dossier should contain
  information on the software design and development process and evidence of
  the validation of the software, as used in the finished device. This
  information should typically include the summary results of all verification,
  validation and testing performed both in-house and in a simulated or actual
  user environment prior to final release. It should also address all of the
  different hardware configurations and, where applicable, operating systems
  identified in the labelling.  | 
 
| 
   7.7  | 
  
   Animal studies:  | 
 |
| 
      | 
  
   (i)  | 
  
   Where studies in an animal model have
  been undertaken to provide evidence of conformity with the Essential
  Principles related to functional safety and performance, detailed information
  should be contained in the dossier.  | 
 
| 
      | 
  
   (ii)  | 
  
   The dossier should describe the study
  objectives, methodology, results, analysis and conclusions and document
  conformity with Good Laboratory Practices. The rationale (and limitations) of
  selecting the particular animal model should be discussed.  | 
 
| 
   7.8  | 
  
   Stability data:  | 
 
| 
      | 
  
   If available, real-time aging data
  shall be submitted to support the claimed shelf-life. However, if real-time
  data is not available, accelerated stability data shall be submitted to
  support the claimed shelf-life. Such a provisional claimed shelf-life may be
  approved provided that the manufacturer immediately initiates real-time
  stability testing to validate the proposed shelf-life. After completion of
  the real time stability analysis, real-time stability data shall be submitted
  in support of the claimed shelf-life.  | 
 
| 
   7.9  | 
  
   Clinical evidence:  | 
 
| 
      | 
  
   The dossier should contain the
  clinical evidence that demonstrates conformity of the device with the
  Essential Principles that apply to it. It needs to address the elements
  contained in the Clinical Investigation, as specified under the Seventh
  Schedule. If a predicate device is available, the manufacturer needs to
  submit the substantial equivalence evaluation along with relevant published
  literature in accordance with these rules.  | 
 
| 
   7.10  | 
  
   Post Marketing Surveillance data
  (Vigilance reporting):  | 
 
| 
      | 
  
   The dossier should contain the Post
  Marketing Surveillance or Vigilance Reporting procedures and data collected
  by the manufacturer encompassing the details of the complaints received and
  corrective and Preventive actions taken for the same.  | 
 
| 
      | 
  
   Note:  | 
 |
| 
      | 
  
   1.  | 
  
   All reports submitted as a part of
  the dossier should be signed and dated by the responsible person.  | 
 
| 
      | 
  
   2.  | 
  
   Batch Release Certificates and
  Certificate of Analysis of finished product for minimum 3 consecutive batches
  should be submitted.  | 
 
| 
      | 
  
   3.  | 
  
   All certificates submitted must be
  within the validity period.  | 
 
| 
      | 
  
   4.  | 
  
   Any information which is not relevant
  for the subject device may be stated as ‘Not Applicable’ in the relevant
  Sections/Columns of the above format, and reasons for non-applicability
  should be provided.  | 
 
APPENDIX
III
Device
Master File for in Vitro Diagnostic Medical Devices
| 
   1.  | 
  
   EXECUTIVE SUMMARY:  | 
 
| 
      | 
  
   An executive summary shall be
  provided by the manufacturer and shall contain:  | 
 
| 
   1.1  | 
  
   Introductory descriptive information
  on the in vitro diagnostic medical device, the intended use and risk Class of
  in vitro diagnostic medical device, novel features (if any), claimed
  shelf-life and a synopsis on the content of the dossier.  | 
 
| 
   1.2  | 
  
   Regulatory status of the similar
  device in India (approved or new in vitro diagnostic medical device).  | 
 
| 
   1.3  | 
  
   Domestic price of the in vitro
  diagnostic medical device in the currency followed in the country of origin.  | 
 
| 
   1.4  | 
  
   Marketing history of the in vitro
  diagnostic medical device from the date of introducing the in vitro
  diagnostic medical device in the market.  | 
 
| 
   1.5  | 
  
   List of regulatory approvals or
  marketing clearance obtained in below format (submit respective copy of
  approval certificate).  | 
 
| 
   Sl. No.  | 
  
   Name of the country  | 
  
   Approved indication  | 
  
   Approved shelf-life  | 
  
   Composition  | 
  
   Risk Class  | 
  
   Date of first approval  | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
 
| 
   1.6  | 
  
   Status of pending request for market
  clearance  | 
 
| 
   Regulatory Agency of the country  | 
  
   Intended use  | 
  
   Indication for use  | 
  
   Registration status and date  | 
  
   Reason for rejection/withdrawal, if any  | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
  
      | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
  
      | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
  
      | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
  
      | 
 
| 
   1.7  | 
  
   Safety and performance related
  information on the in vitro diagnostic medical device:  | 
 
| 
   (a)  | 
  
   Summary of reportable events and
  field safety corrective action from the date of introduction.  | 
 
For adverse event (false
diagnosis or any other hazard during its use)
| 
   Adverse event (false diagnosis)  | 
  
   Frequency of occurrence during the period (number
  of report/total units sold)  | 
 
| 
      | 
  
      | 
 
| 
      | 
  
      | 
 
| 
      | 
  
      | 
 
For Field Safety Corrective
Action (FSCA)
| 
   Date of FSCA  | 
  
   Reason for FSCA  | 
  
   Countries where FSCA was conducted  | 
  
   Description of the action taken  | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
 
| 
   (b)  | 
  
   If the in vitro diagnostic medical
  device contains any of the following then descriptive information on the
  following need to be provided.  | 
 |
| 
      | 
  
   (1)  | 
  
   Animal or human fluids or derivatives
  thereof, rendered non-viable.  | 
 
| 
      | 
  
   (2)  | 
  
   Cells, tissues and/or derivatives of
  microbial recombinant origin.  | 
 
| 
   2.  | 
  
   Description and specification,
  including variants and accessories of the in vitro diagnostic medical device  | 
 
| 
   2.1  | 
  
   Description  | 
 
| 
      | 
  
   The device master file should include
  the following device descriptive information:  | 
 
| 
      | 
  
   (a)  | 
  
   it may include—  | 
 
| 
      | 
  
      | 
  
   (1)  | 
  
   what is detected;  | 
 
| 
      | 
  
      | 
  
   (2)  | 
  
   its function (for example screening,
  monitoring, diagnostic or aid to diagnosis, staging or aid to staging of
  disease);  | 
 
| 
      | 
  
      | 
  
   (3)  | 
  
   the specific disorder, condition or
  risk factor of interest that it is intended to detect, define or
  differentiate;  | 
 
| 
      | 
  
      | 
  
   (4)  | 
  
   whether it is automated or not;  | 
 
| 
      | 
  
      | 
  
   (5)  | 
  
   whether it is qualitative or
  quantitative;  | 
 
| 
      | 
  
      | 
  
   (6)  | 
  
   the type of specimen required (e.g.
  serum, plasma, whole blood, tissue biopsy, urine);  | 
 
| 
      | 
  
      | 
  
   (7)  | 
  
   testing population;  | 
 
| 
      | 
  
   (b)  | 
  
   the intended user (lay person or
  professional);  | 
 
| 
      | 
  
   (c)  | 
  
   a general description of the
  principle of the assay method;  | 
 
| 
      | 
  
   (d)  | 
  
   the risk based Class of the device;  | 
 
| 
      | 
  
   (e)  | 
  
   a description of the components (e.g.
  reagents, assay controls and calibrators) and where appropriate, a
  description of the reactive ingredients of relevant components (such as
  antibodies, antigens, nucleic acid primers) where applicable;  | 
 
| 
      | 
  
   (f)  | 
  
   a description of the specimen
  collection and transport materials provided with the in vitro diagnostic
  medical device or descriptions of specifications recommended for use;  | 
 
| 
      | 
  
   (g)  | 
  
   for instruments of automated assays;
  a description of the appropriate assay characteristics or dedicated assays;  | 
 
| 
      | 
  
   (h)  | 
  
   for automated assays; a description
  of the appropriate instrumentation characteristics or dedicated
  instrumentation;  | 
 
| 
      | 
  
   (i)  | 
  
   a description of any software to be
  used with the in vitro diagnostic medical device;  | 
 
| 
      | 
  
   (j)  | 
  
   a description or complete list of the
  various configurations/variants of the in vitro diagnostic medical device
  that will be made available;  | 
 
| 
      | 
  
   (k)  | 
  
   a description of the accessories,
  other in vitro diagnostic medical device and other products that are not in
  vitro diagnostic medical device, which are intended to be used in combination
  with the in vitro diagnostic medical device.  | 
 
| 
      | 
  
   Reference to the manufacturer's
  previous device generation(s) or similar devices or device history.  | 
 |
| 
   2.2  | 
  
   For a new in vitro diagnostic medical
  device:  | 
 
| 
      | 
  
   Where relevant to demonstrating
  conformity to the essential principles, and to provide general background
  information, the device master file may provide a summary of Clinical
  Performance Evaluation reports.  | 
 
| 
   2.3  | 
  
   For an in vitro diagnostic medical
  device already available on the market in India:  | 
 |
| 
      | 
  
   (i)  | 
  
   This information may include a
  summary of the number of adverse event reports related to the safety and
  performance of this in vitro diagnostic medical device in relation to the
  number of in vitro diagnostic medical devices placed on the market.  | 
 
| 
      | 
  
   (ii)  | 
  
   External certificates and documents
  which give written evidence of conformity with the essential principles may
  be annexed to the device master file.  | 
 
| 
      | 
  
   (iii)  | 
  
   comparative analysis to prove
  substantial equivalence to the predicate device(s), if claimed in the
  application.  | 
 
| 
   3.  | 
  
   Essential principles checklist:  | 
 
| 
      | 
  
   (i)  | 
  
   The device master file should include
  an essential principles checklist that identifies:  | 
 |
| 
      | 
  
      | 
  
   (a)  | 
  
   the essential principles of safety
  and performance;  | 
 
| 
      | 
  
      | 
  
   (b)  | 
  
   whether each essential principle
  applies to the in vitro diagnostic medical device and if not, why not;  | 
 
| 
      | 
  
      | 
  
   (c)  | 
  
   the method used to demonstrate
  conformity with each essential principle that applies; and  | 
 
| 
      | 
  
      | 
  
   (d)  | 
  
   the reference to the actual technical
  documentation that offers evidence of conformity with each method used.  | 
 
| 
      | 
  
   (ii)  | 
  
   The method used to demonstrate
  conformity may include one or more of the following—  | 
 |
| 
      | 
  
      | 
  
   (a)  | 
  
   conformity with recognised or other
  standards;  | 
 
| 
      | 
  
      | 
  
   (b)  | 
  
   conformity with a commonly accepted
  industry test method (reference method);  | 
 
| 
      | 
  
      | 
  
   (c)  | 
  
   conformity with appropriate in house
  test methods that have been validated and verified;  | 
 
| 
      | 
  
      | 
  
   (d)  | 
  
   comparison to an in vitro diagnostic
  medical device already available on the market.  | 
 
| 
      | 
  
   (iii)  | 
  
   The essential principles checklist
  should include a cross-reference to the location of such evidence both within
  the full technical documentation held by the manufacturer and within the
  Device master file (when such documentation is specifically required for inclusion
  in the Summary Technical Documentation as outlined in this guidance).  | 
 |
| 
   4.  | 
  
   Risk analysis and control summary:  | 
 
| 
      | 
  
   The device master file should contain
  a summary of the risks identified during the risk analysis process and a
  description of how these risks have been controlled to an acceptable level.
  Preferably, this risk analysis should be based on recognised standards and be
  part of the manufacturer's risk management plan.  | 
 
| 
      | 
  
   The summary should address possible
  hazards for the in vitro diagnostic medical device such as the risk from
  false positive or false negative results, indirect risks which may result
  from in vitro diagnostic medical device associated hazards, such as
  instability, which could lead to erroneous results, or from user related
  hazards, such as reagents containing infectious agents. The results of the
  risk analysis should provide a conclusion with evidence that remaining risks
  are acceptable when compared to the benefits.  | 
 
| 
   5.  | 
  
   Design and manufacturing information:  | 
 
| 
   5.1  | 
  
   Device design:  | 
 
| 
      | 
  
   The Device master file should contain
  information to allow a reviewer to obtain a general understanding of the
  design applied to the in vitro diagnostic medical device. It should include a description of
  the critical ingredients of an assay such as antibodies, antigens, enzymes
  and nucleic acid primers provided or recommended for use with the in vitro
  diagnostic medical device. This section is not intended to take
  the place of the more detailed information required for a QMS audit or other
  conformity assessment activity. If design takes place at multiple sites, a
  controlling site must be identified.  | 
 
| 
   5.2  | 
  
   Manufacturing processes:  | 
 
| 
      | 
  
   The device master file should contain
  information to allow a reviewer to obtain a general understanding of the
  manufacturing processes. It is not intended to take the place of the more
  detailed information required for a QMS audit or other conformity assessment
  activity. The information may take the form of a process flow chart showing,
  for example, an overview of production including the technologies used,
  assembly, any in-process and final product testing, and packaging of the
  finished in vitro medical device.  | 
 
| 
   5.3  | 
  
   Manufacturing sites:  | 
 
| 
      | 
  
   The device master file should
  identify the sites where these activities are performed (this does not
  include the sites of all suppliers of raw materials but only the sites that
  are involved in critical manufacturing activities). If Quality Management
  System certificates, or the equivalent, exist for these sites, they may be
  annexed to the device master file.  | 
 
| 
   6.  | 
  
   Product validation and verification:  | 
 
| 
      | 
  
   The information provided in the
  product validation and verification section of the device master file will
  vary in the level of detail as determined by the class of the device. The
  device master file should summarise the results of validation and verification
  studies undertaken to demonstrate conformity of the in vitro diagnostic
  medical device with the essential principles that apply to it. Where
  appropriate, such information might come from literature.  | 
 
| 
      | 
  
   For the purpose of the device master
  file document, summary and detailed information are defined as follows:  | 
 
| 
      | 
  
   (i)  | 
  
   Summary information:  | 
 |
| 
      | 
  
      | 
  
   A summary should provide enough to
  assess the validity of that information by the regulatory authorities. This
  summary should contain a brief description of:  | 
 |
| 
      | 
  
      | 
  
   (a)  | 
  
   the study protocol;  | 
 
| 
      | 
  
      | 
  
   (b)  | 
  
   the study results;  | 
 
| 
      | 
  
      | 
  
   (c)  | 
  
   the study conclusion.  | 
 
| 
      | 
  
      | 
  
   This summary may include:  | 
 |
| 
      | 
  
      | 
  
   (a)  | 
  
   Where a recognised standard exists, a
  declaration/certificate of conformity to a recognised standard can be
  provided with a summary of the data if no acceptance criteria are specified
  in the standard;  | 
 
| 
      | 
  
      | 
  
   (b)  | 
  
   In the absence of a recognised
  standard, a declaration/certificate of conformity to a published standard
  that has not been recognised might be provided if it is supported by a
  rationale for its use, and summary of the data, and a conclusion, if no
  acceptance criteria are specified in the standard;  | 
 
| 
      | 
  
      | 
  
   (c)  | 
  
   In the absence of a recognised
  standard and non-recognised published standards, a professional guideline,
  industry method, or in-house standard may be referred to in the summarised
  information. However, it should be supported by a rationale for its use, a
  description of the method used, a summary of the data in sufficient detail
  and a conclusion to allow assessment of its adequacy;  | 
 
| 
      | 
  
      | 
  
   (d)  | 
  
   A review of relevant published
  literature regarding the device/analyte (measurand) or substantially similar
  in vitro diagnostic medical devices.  | 
 
| 
      | 
  
   (ii)  | 
  
   Detailed information:  | 
 |
| 
      | 
  
      | 
  
   Detailed information should include:  | 
 |
| 
      | 
  
      | 
  
   (a)  | 
  
   complete study protocol;  | 
 
| 
      | 
  
      | 
  
   (b)  | 
  
   method of data analysis;  | 
 
| 
      | 
  
      | 
  
   (c)  | 
  
   complete study report;  | 
 
| 
      | 
  
      | 
  
   (d)  | 
  
   study conclusion.  | 
 
| 
      | 
  
   For detailed information, when a
  recognised standard exists that contains the protocol and the method of data
  analysis, this information can be substituted by a declaration or certificate
  of conformity to the recognised standard along with a summary of the data and
  conclusions, where appropriate, actual test result summaries with their
  acceptance criteria should be provided and not just pass/fail statements.  | 
 ||
| 
   7.  | 
  
   Analytical Studies:  | 
 
| 
      | 
  
   The statements and descriptions in
  the following sections refer to all in vitro diagnostic medical devices. It
  must be noted however that there are applicability differences between
  instrumentation and reagent-based assays, and that the assays themselves may
  be quantitative, semi-quantitative or qualitative in nature. There may be
  limited applicability of some of the following sub-sections for qualitative
  or semi-quantitative assays. Where possible, comments regarding
  instrumentation or qualitative assays appear in the sub-sections.  | 
 
| 
   8.  | 
  
   Specimen type:  | 
 
| 
      | 
  
   (a)  | 
  
   This section should describe the
  different specimen types that can be used. This should include their
  stability and storage conditions. Stability includes storage and where
  applicable transport conditions. Storage includes elements such as duration,
  temperature limits and freeze/thaw cycles.  | 
 
| 
      | 
  
   (b)  | 
  
   This section should include summary
  information for each matrix and anticoagulant when applicable, including a
  description of the measurement procedure for comparison or determination of
  measurement accuracy. This includes information such as specimen type tested,
  number of samples, sample range (using spiked samples as appropriate) or
  target concentrations tested, calculations and statistical methods, results
  and conclusions.  | 
 
| 
   9.  | 
  
   Analytical performance
  characteristics.  | 
 
| 
   9.1  | 
  
   Accuracy of measurement:  | 
 
| 
      | 
  
   This section should describe both
  trueness and precision studies.  | 
 
| 
      | 
  
   Explanation.—The general term
  measurement accuracy is currently used to cover both trueness and precision,
  whereas this term was used in the past to cover only the one component now
  named trueness.  | 
 
| 
      | 
  
   While measurement trueness, affected
  by systematic error, is normally expressed in terms of bias, measurement
  precision, affected by random error, is naturally expressed in terms of
  standard deviation. Accuracy is affected by a combination
  of systematic and random effects that contribute as individual components of
  the total error of measurement.  | 
 
| 
   9.2  | 
  
   Reproducibility:  | 
 
| 
      | 
  
   This section should include
  reproducibility estimates and information about the studies used to estimate,
  as appropriate, variability between days, runs, sites, lots, operators and
  instruments. Such variability is also known as “Intermediate Precision”.
  Reproducibility data is obtained for instrumentation in conjunction with an
  appropriate assay.  | 
 
| 
      | 
  
   Note 1.—Such studies should include
  the use of samples that represent the full range of expected analyte
  (measurand) that can be measured by the test as claimed by the manufacturer.  | 
 
| 
      | 
  
   Note 2.—If a recognised standard is
  used, a declaration/certificate of conformity to the recognised standard along
  with a summary of the data and conclusions.  | 
 
| 
   10.  | 
  
   Analytical sensitivity:  | 
 
| 
      | 
  
   This section should include
  information about the study design and results. It should provide a
  description of specimen type and preparation including matrix, analyte
  (measurand) levels, and how levels were established. The number of replicates
  tested at each concentration should also be provided as well as a description
  of the calculation used to determine assay sensitivity. For example:  | 
 
| 
      | 
  
   (a)  | 
  
   Number of standard deviations above
  the mean value of the sample without analyte (measurand), commonly referred
  to as limit of blank (LoB).  | 
 
| 
      | 
  
   (b)  | 
  
   Lowest concentration distinguishable
  from zero, based on measurements of samples containing analyte (measurand),
  commonly referred to as limit of detection (LoD).  | 
 
| 
      | 
  
   (c)  | 
  
   Lowest concentration at which
  precision and/or trueness are within specified criteria, commonly referred to
  as limit of quantitation (LoQ).  | 
 
| 
      | 
  
   For Class C and Class D in vitro
  diagnostic medical devices, detailed information would be provided.  | 
 |
| 
   11.  | 
  
   Analytical specificity:  | 
 |
| 
      | 
  
   (i)  | 
  
   This section should describe
  interference and cross reactivity studies to determine the analytical
  specificity, defined as the ability of a measurement procedure to detect or
  measure only the analyte (measurand) to be detected, in the presence of other
  substances/agents in the sample.  | 
 |
| 
      | 
  
   (ii)  | 
  
   Provide information on the evaluation
  of potentially interfering and cross reacting substances/agents on the assay.
  Information should be provided on the substance/agent type and concentration
  tested, sample type, analyte (measurand) test concentration, and results.  | 
 |
| 
      | 
  
   (iii)  | 
  
   Interferents and cross reacting
  substances/agents, which vary greatly depending on the assay type and design,
  could derive from exogenous or endogenous sources such as:  | 
 |
| 
      | 
  
      | 
  
   (a)  | 
  
   substances used for patient treatment
  (e.g. therapeutic drugs, anticoagulants, etc.);  | 
 
| 
      | 
  
      | 
  
   (b)  | 
  
   substances ingested by the patient
  (e.g. over the counter medications, alcohol, vitamins, foods, etc.);  | 
 
| 
      | 
  
      | 
  
   (c)  | 
  
   substances added during sample
  preparation (e.g. preservatives, stabilizers);  | 
 
| 
      | 
  
      | 
  
   (d)  | 
  
   substances encountered in specific
  specimens types (e.g. haemoglobin, lipids, bilirubin, proteins);  | 
 
| 
      | 
  
      | 
  
   (e)  | 
  
   analytes of similar structure (e.g.
  precursors, metabolites) or medical conditions unrelated to the test
  condition including specimens negative for the assay but positive for a
  condition that may mimic the test condition (e.g. for a hepatitis A assay:
  test specimens negative for hepatitis A virus, but positive for hepatitis B
  virus).  | 
 
| 
      | 
  
      | 
  
   Explanation.—Interference studies
  involve adding the potential interferent to the sample and determining any
  bias of the test parameter relative to the control sample to which no
  interferent has been added.  | 
 |
| 
   12.  | 
  
   Metrological traceability of
  calibrator and control material values:  | 
 
| 
      | 
  
   Where applicable, summarise the
  information about metrological traceability of values assigned to calibrators
  and trueness control materials. Include, for example, methods and acceptance
  criteria for the metrological traceability to reference materials and/or
  reference measurement procedures and a description of value assignment and
  validation. Precision control materials, used
  when establishing the reproducibility of a measurement procedure do not
  require the assessment of metrological traceability to a reference material
  or a reference method.  | 
 
| 
   13.  | 
  
   Measuring range of the assay:  | 
 
| 
      | 
  
   This section should include a summary
  of studies which define the measuring range (linear and non-linear measuring
  systems) including the limit of detection and describe information on how
  these were established. This summary should include a description of specimen
  type, number of samples, number of replicates, and preparation including
  information on matrix, analyte (measurand) levels and how levels were
  established. If applicable, add a description of high dose hook effect and
  the data supporting the mitigation (e.g. dilution) steps.  | 
 
| 
   14.  | 
  
   Definition of Assay Cut-off:  | 
 
| 
      | 
  
   This section should provide a summary
  of analytical data with a description of the study design including methods
  for determining the assay cut-off, including:  | 
 
| 
      | 
  
   (a)  | 
  
   the population(s) studied
  (demographics/selection/inclusion and exclusion criteria/number of
  individuals included);  | 
 
| 
      | 
  
   (b)  | 
  
   method or mode of characterisation of
  specimens; and  | 
 
| 
      | 
  
   (c)  | 
  
   Statistical methods e.g. Receiver
  Operator Characteristic (ROC) to generate results and if applicable, define
  gray-zone/equivocal zone.  | 
 
| 
   15.  | 
  
   Stability (excluding specimen
  stability):  | 
 |
| 
      | 
  
   This section should describe claimed
  shelf-life, in use stability and shipping studies.  | 
 |
| 
   16.  | 
  
   Claimed Shelf life:  | 
 |
| 
      | 
  
   This section should provide
  information on stability testing studies to support the claimed shelf-life.
  Testing should be performed on at least three different lots manufactured
  under conditions that are essentially equivalent to routine production
  conditions (these lots do not need to be consecutive lots). Accelerated
  studies or extrapolated data from real time data are acceptable for initial
  shelf-life claim but need to be followed up with real time stability studies.
  Such detailed information should describe:  | 
 |
| 
      | 
  
   (a)  | 
  
   the study report (including the
  protocol, number of lots, acceptance criteria and testing intervals);  | 
 
| 
      | 
  
   (b)  | 
  
   when accelerated studies have been
  performed in anticipation of the real time studies, the method used for accelerated
  studies;  | 
 
| 
      | 
  
   (c)  | 
  
   conclusions and claimed shelf-life.  | 
 
| 
      | 
  
   Explanation.—Shelf life can be
  derived from the lot with the longest real time stability data as long as
  accelerated or extrapolated data from all three lots are comparable.  | 
 |
| 
   17.  | 
  
   In use stability:  | 
 |
| 
      | 
  
   This section should provide
  information on in use stability studies for one lot reflecting actual routine
  use of the device (real or simulated). This may include open vial stability
  and/or, for automated instruments, on board stability. In the case of
  automated instrumentation if calibration stability is claimed, supporting
  data should be included. Such detailed information should describe:  | 
 |
| 
      | 
  
   (a)  | 
  
   the study report (including the
  protocol, acceptance criteria and testing intervals);  | 
 
| 
      | 
  
   (b)  | 
  
   conclusions and claimed in use
  stability.  | 
 
| 
   18.  | 
  
   Shipping stability:  | 
 |
| 
      | 
  
   This section should provide
  information on shipping stability studies for one lot to evaluate the
  tolerance of products to the anticipated shipping conditions. Shipping
  studies can be done under real and/or simulated conditions and should include
  variable shipping conditions such as extreme heat or cold. Such information
  should describe:  | 
 |
| 
      | 
  
   (a)  | 
  
   the study report (including the
  protocol, acceptance criteria);  | 
 
| 
      | 
  
   (b)  | 
  
   method used for simulated conditions;  | 
 
| 
      | 
  
   (c)  | 
  
   conclusion and recommended shipping
  conditions.  | 
 
| 
   19.  | 
  
   Clinical Evidence:  | 
 
| 
      | 
  
   The device master file should contain
  the Clinical Evidence, Evaluation report that demonstrates conformity of the
  in vitro diagnostic medical device to the Essential Principles that apply to
  it.  | 
 
| 
   20.  | 
  
   Labelling:  | 
 
| 
      | 
  
   The device master file should
  typically contain a complete set of labeling associated with the in vitro
  medical device as described in Chapter VI.  | 
 
| 
   21.  | 
  
   Post-marketing surveillance data
  (vigilance reporting):  | 
 
| 
      | 
  
   The dossier should contain the
  post-marketing surveillance or vigilance reporting procedures and data
  collected by the manufacturer encompassing the details of the complaints
  received and corrective and Preventive actions taken for the same.  | 
 
| 
   22.  | 
  
   Information required to be submitted
  for the in vitro diagnostic medical device:  | 
 |
| 
      | 
  
   (1)  | 
  
   The details of source antigen or
  antibody as the case may be and characterisation of the same. Process control
  of coating of antigen or antibody on the base material like Nitrocellulose
  paper, strips or cards or ELISA wells, etc. Detailed composition of the in
  vitro diagnostic medical device and manufacturing flow chart process of the in
  vitro diagnostic medical device showing the specific flow diagram of
  individual components or source of the individual components.  | 
 
| 
      | 
  
   (2)  | 
  
   Test protocol of the in vitro
  diagnostic medical device showing the specifications and method of testing.
  In house evaluation report of sensitivity, specificity and stability studies
  carried out by the manufacturer.  | 
 
| 
      | 
  
   (3)  | 
  
   In case of imported diagnostic in
  vitro diagnostic medical devices, the report of evaluation in details
  conducted by the National Control Authority of country of origin.  | 
 
| 
      | 
  
   (4)  | 
  
   Specimen batch test report for at
  least consecutive 3 batches showing specification of each testing parameter.  | 
 
| 
      | 
  
   (5)  | 
  
   The detailed test report of all the
  components used/packed in the finished in vitro diagnostic medical device.  | 
 
| 
      | 
  
   (6)  | 
  
   Pack size and labeling.  | 
 
| 
      | 
  
   (7)  | 
  
   Product inserts.  | 
 
| 
      | 
  
   (8)  | 
  
   Specific evaluation report, if done
  by any laboratory in India, showing the sensitivity and specificity of the in
  vitro diagnostic medical device.  | 
 
| 
      | 
  
   (9)  | 
  
   Specific processing like safe handling,
  material control, area control, process control, and stability studies,
  storage at quarantine stage and finished stage, packaging should be
  highlighted in the product dossier.  | 
 
| 
      | 
  
   Note:  | 
 |
| 
      | 
  
   1.  | 
  
   All the test reports submitted as a
  part of the dossier should be signed and dated by the responsible person.  | 
 
| 
      | 
  
   2.  | 
  
   Batch Release Certificates and
  Certificate of Analysis of finished product for minimum 3 consecutive batches
  should be submitted.  | 
 
| 
      | 
  
   3.  | 
  
   All certificates submitted must be
  within the validity period.  | 
 
| 
      | 
  
   4.  | 
  
   Any information which is not relevant
  for the subject in vitro diagnostic medical device may be stated as ‘Not
  Applicable’ in the relevant sections/columns of the above format, and reasons
  for non-applicability should be provided.  | 
 
Part IV Information
required to be submitted with the Application Form for import or manufacture of
medical devices which does not have predicate device
| 
   (a)  | 
  
   Data to be submitted along with the
  application (for medical devices other than new in vitro diagnostic)—  | 
 |
| 
   1.  | 
  
   Design Analysis data including,
  (whichever applicable)—  | 
 |
| 
      | 
  
   (a)  | 
  
   design input and design output
  documents;  | 
 
| 
      | 
  
   (b)  | 
  
   mechanical and electrical tests;  | 
 
| 
      | 
  
   (c)  | 
  
   reliability tests;  | 
 
| 
      | 
  
   (d)  | 
  
   validation of software relating to
  the function of the device;  | 
 
| 
      | 
  
   (e)  | 
  
   any performance tests;  | 
 
| 
      | 
  
   (f)  | 
  
   in vitro tests.  | 
 
| 
   2.  | 
  
   Bio-compatibility tests data, Report
  of bio-compatibility tests along with rationale for selecting these tests.
  Summary report of the bio-compatibility study including the conclusion of the
  study.  | 
 |
| 
   3.  | 
  
   Risk Management data;  | 
 |
| 
   4.  | 
  
   Animal Performance study data;  | 
 |
| 
   5.  | 
  
   Pilot or Pivotal Clinical
  Investigation data, including that carried out in other countries if any;  | 
 |
| 
   6.  | 
  
   In case, if waiver from clinical
  investigation is claimed in accordance with the provisions of Medical Device
  Rules, 2017, the information or supporting data shall be submitted.  | 
 |
| 
   7.  | 
  
   Regulatory status and Restriction on
  use in other countries (if any) where marketed or approved;  | 
 |
| 
   [59][8.  | 
  
   Proposed instructions for use or
  electronic instructions for use and labels.]  | 
 |
| 
   (b)  | 
  
   Data to be submitted along with the
  application (for new in vitro diagnostic medical devices)—  | 
 |
| 
   1.  | 
  
   Device data including, (whichever
  applicable)—  | 
 |
| 
      | 
  
   (i)  | 
  
   design input, design output
  documents, stability data;  | 
 
| 
      | 
  
   (ii)  | 
  
   device specification including
  specificity, sensitivity, reproducibility and reputability;  | 
 
| 
      | 
  
   (iii)  | 
  
   product validation and software
  validation relating to the function of the device (if any);  | 
 
| 
      | 
  
   (iv)  | 
  
   performance evaluation report from a
  laboratory designated under sub-rule (1) of Rule 19.  | 
 
| 
   2.  | 
  
   Risk Management data.  | 
 |
| 
   3.  | 
  
   Clinical Performance Evaluation data
  carried out in India and in other countries (if any).  | 
 |
| 
   4.  | 
  
   Regulatory status and Restriction on
  use in other countries (if any) where marketed or approved.  | 
 |
| 
   [60][5.  | 
  
   Proposed instructions for use or
  electronic instructions for use and labels.]  | 
 |
FIFTH
SCHEDULE
[See Rules
20(3), 20(5), 20(8), 22(i)]
Quality
Management System for medical devices and in vitro diagnostic medical devices
1. General Requirements:
1.1. This schedule specifies
requirements for a quality management system that shall be used by the
manufacturer for the design and development, manufacture, packaging, labelling,
testing, installation and servicing of medical devices and in vitro diagnostic
medical devices. If the manufacturer does not carry out design and development
activity, the same shall be recorded in the quality management system. The
manufacturer shall maintain conformity with this Schedule to reflect the
exclusions.
1.2. If any requirement in
Paragraph 7 (product realisation) of this Schedule is not applicable due to the
nature of the medical device and in vitro diagnostic medical devices for which
the quality management system is applied, the manufacturer does not need to
include such a requirement in its quality management system.
1.3. The processes required
by this Schedule, which are applicable to the medical device and in-vitro
diagnostic medical device, but which are not performed by the manufacturer are
the responsibility of the manufacturer and are accounted for in the
manufacturer's quality management system.
1.4. If a manufacturer
engages in only some operations subject to the requirements of this part, and
not in others, that manufacturer need only to comply with those requirements
which are applicable to the operations in which it is engaged.
1.5. It is emphasised that
the quality management system requirements specified in this Schedule are in
addition to complementary to technical requirements for products.
1.6. Manufacturers of
components or parts of finished devices and in vitro diagnostic medical devices
are encouraged to use appropriate provisions of this schedule as guidance.
2. Applicability:
The provisions of this
Schedule shall be applicable to manufacturers of finished devices, in vitro
diagnostic medical devices, mechanical contraceptives (condoms, intrauterine
devices, tubal rings), surgical dressings, surgical bandages, surgical
staplers, surgical sutures and ligatures, blood and blood components collection
bags with or without anticoagulants.
3. Terms and definitions:
3.1 Active implantable
medical device.—Active medical device which is intended to be totally or
partially introduced, surgically or medically, into the human or animal body or
by medical intervention into a natural orifice and which is intended to remain
after the procedure.
3.2 Active medical
device.—Medical device relying for its functioning on a source of electrical
energy or any source of power other than that directly generated by the human
or animal body or gravity.
3.3 Advisory
notice.—Notice issued by the manufacturer, subsequent to delivery of the
medical device and in vitro diagnostic medical devices, to provide
supplementary information or to advise what action should be taken in or both
in—
(a)      the use of a medical device
and in vitro diagnostic medical devices;
(b)      the modification of a
medical device and in vitro diagnostic medical devices;
(c)      the return of the medical
device and in vitro diagnostic medical devices to the organisation that
supplied it; or
(d)      the destruction of a
medical device and in vitro diagnostic medical devices.
3.4 Customer
complaint.—Written, electronic or oral communication that alleges deficiencies
related to the identity, quality, durability, reliability, safety,
effectiveness or performance of a medical device and in vitro diagnostic
medical devices that has been placed on the market.
3.5 Implantable
medical device.—Medical device intended—
(a)      to be totally or partially
introduced into the human or animal body or a natural orifice; or
(b)      to replace an epithelial
surface or the surface of the eye;
by surgical intervention,
and which is intended to remain after the procedure for at least thirty days,
and which can only be removed by medical or surgical intervention.
3.6 Component means any raw
material, substance, piece, part, software, firmware, labeling, or assembly
which is intended to be included as part of the finished, packaged, and labeled
device.
3.7 Design input means the
physical and performance requirements of a device that are used as a basis for
device design.
3.8 Design output means the
results of a design effort at each design phase and at the end of the total
design effort. The finished design output is the basis for the device master
record. The total finished design output consists of the device, its packaging
and labeling, and the device master record.
3.9 Design review means a
documented, comprehensive, systematic examination of a design to evaluate the
adequacy of the design requirements, to evaluate the capability of the design
to meet these requirements, and to identify problems.
3.10 Finished device means
any device or accessory to any device that is suitable for use or capable of
functioning, whether or not it is packaged, labeled or sterilised.
3.11 Management with
executive responsibility means those senior employees of a manufacturer who
have the authority to establish or make changes to the manufacturer's quality
policy and quality system.
3.12 Medical device
including substances used for in vitro diagnosis referred to in Rule 3 of these
rules.
3.13 Quality audit means a
systematic, independent examination of a manufacturer's quality system that is
performed at defined intervals and at sufficient frequency to determine whether
both quality system activities and the results of such activities comply with
quality system procedures, that these procedures are implemented effectively,
and that these procedures are suitable to achieve quality system objectives.
3.14 Quality policy means
the overall intention and direction of an organisation with respect to quality,
as established by management with executive responsibility.
3.15 Quality system means
the organisational structure, responsibilities, procedures, processes, and
resources for implementing quality management.
3.16 Rework means action
taken on a non-conforming product that will fulfil the specified Device Master
File requirements before it is released for distribution.
3.17 Specification means
any requirement with which a product, process, service, or other activity must conform.
3.18 Validation means
confirmation by examination and provision of objective evidence that the
particular requirement for a specific intended use can be consistently
fulfilled.
3.18.1 Process validation
means establishing by objective evidence that a process consistently produces a
result or product meeting its predetermined specifications.
3.18.2 Design validation
means establishing by objective evidence that device specifications conform
with user needs and intended use(s).
3.19 Verification means
confirmation by examination and provision of objective evidence that specified
requirements have been fulfilled.
4. Quality management
system:
4.1 General: The
manufacturer shall establish, document, implement and maintain a quality
management system and maintain its effectiveness in accordance with the
requirements of this schedule.
The manufacturer shall—
(a)      identify the processes
needed for the quality management system and their application throughout the
organisation;
(b)      determine the sequence and interaction
of these processes;
(c)      determine criteria and
methods needed to ensure that both the operation and control of these processes
are effective;
(d)      ensure the availability of
resources and information necessary to support the operation and monitoring of
these processes;
(e)      monitor, measure and
analyse these processes; and
(f)       implement actions necessary
to achieve planned results and maintain the effectiveness of these processes.
These processes shall be
managed by the manufacturer in accordance with the requirements of this
Schedule. Where a manufacturer chooses to outsource any process that affects
product conformity with requirements, the manufacturer shall ensure control
over such processes. Control of such outsourced processes shall be identified
within the quality management system.
Note: Processes needed for
the quality management system referred to above shall include processes for
management activities, provision of resources, product realisation and
measurement.
4.2 Documentation
requirements:
4.2.1 General: The
quality management system documentation shall include—
(a)      documented statements of a
quality policy and quality objectives;
(b)      a quality manual;
(c)      documented procedures
required by this schedule;
(d)      documents needed by the
manufacturer to ensure the effective planning, operation and control of its
processes;
(e)      records required by this
Schedule, and
where this Schedule
specifies that a requirement, procedure, activity or special arrangement be
“documented”, it shall, in addition, be implemented and maintained.
For each type of medical
device or in vitro diagnostic medical devices, the manufacturer shall establish
and maintain a file either containing or identifying documents defining product
specifications and quality management system requirements. These documents
shall define the complete manufacturing process and, if applicable,
installation.
The manufacture shall
prepare documentation for device or in vitro diagnostic medical devices in a
form of a Device Master File containing specific information as referred to in
Fourth Schedule.
Data may be recorded by
electronic data processing systems or other reliable means, but documents and
record relating to the system in use shall also be available in a hard copy to
facilitate checking of the accuracy of the records. Wherever documentation is
handled by electronic data processing methods, authorised persons shall enter
or modify data in the computer. There shall be record of changes and deletions.
Access shall be restricted by ‘passwords’ or other means and the result of
entry of critical data shall be independently checked. Batch records
electronically stored shall be protected by a suitable back-up. During the
period of retention, all relevant data shall be readily available.
4.2.2 Quality manual:
The manufacturer shall establish and maintain a quality manual that includes—
(a)      the scope of the quality
management system, including details of and justification for any exclusion or
non-application or both;
(b)      the documented procedures
established for the quality management system, or reference to them; and
(c)      a description of the
interaction between the processes of the quality management system.
The quality manual shall
outline the structure of the documentation used in the quality management
system.
The manufacturer shall
prepare documentation in a form of a Plant Master File containing specific
information about the facilities, personnel and other details as prescribed in
Fourth.
4.2.3 Control of
documents: Documents required by the quality management system shall be controlled.
Records are a special type of document and shall be controlled according to the
requirements given in the control of records. Documents shall be approved,
signed and dated by the appropriate and the authorised person.
A documented procedure shall
be established to define the controls needed—
(a)      to review and approve
documents for adequacy prior to issue;
(b)      to review and update as
necessary and re-approve documents;
(c)      to ensure that changes and
the current revision status of documents are identified;
(d)      to ensure that relevant
versions of applicable documents are available at points of use;
(e)      to ensure that documents
remain legible and readily identifiable;
(f)       to ensure that documents of
external origin are identified and their distribution controlled; and
(g)      to prevent the unintended
use of obsolete documents, and to apply suitable identification to them if they
are retained for any purpose.
Changes to document shall
be reviewed and approved. Change records shall be maintained which will include
a description of the change, identification of the affected documents, the
signature of the approving individual, the approval date, and when the change
becomes effective.
The manufacturer shall
ensure that changes to documents are reviewed and approved either by the
original approving functionary or another designated functionary which has
access to pertinent background information upon which to base its decisions.
The manufacturer shall
define the period for which at least one copy of obsolete controlled documents
shall be retained.
This period shall ensure
that documents to which medical devices or in vitro diagnostic medical devices
have been manufactured and tested are retained for at least one year after the
date of expiry of the medical device or in vitro diagnostic medical devices as
defined by the manufacturer.
4.2.4 Control of
records: Records shall be established and maintained to provide evidence of
conformity to the requirements and of the effective operation of the quality
management system. Records shall remain legible, readily identifiable and
retrievable. A documented procedure shall be established to define the controls
needed for the identification, storage, protection, retrieval, retention time
and disposition of records.
The manufacturer shall
retain the records for a period of time at least one year after the date of
expiry of the medical device or in vitro diagnostic medical devices as defined
by the manufacturer, but not less than two years from the date of product
release by the manufacturer.
5. Management
responsibility:
5.1 Management
commitment: Top management of the manufacturer shall provide evidence of its
commitment to the development and implementation of the quality management
system and maintaining its effectiveness by—
(a)      communicating to the
employees the importance of meeting customer as well as statutory and
regulatory requirements;
(b)      establishing the quality
policy;
(c)      ensuring that quality
objectives are established;
(d)      conducting management
reviews; and
(e)      ensuring the availability
of resources.
5.2 Customer focus:
Top management of the manufacturer shall ensure that customer requirements are
determined and are met.
5.3 Quality policy:
Top management of the manufacturer shall ensure that the quality policy—
(a)      is appropriate to the
purpose of the manufacturing facility;
(b)      includes a commitment to
comply with requirements and to maintain the effectiveness of the quality
management system;
(c)      provides a framework for
establishing and reviewing quality objectives;
(d)      is communicated and understood
within the manufacturer's organisation; and
(e)      is reviewed for continuing
suitability.
5.4 Planning:
5.4.1 Quality
objectives: Top management of the manufacturer shall ensure that quality
objectives, including those needed to meet requirements for product, are
established at relevant functions and levels within the manufacturing
organisation. The quality objectives shall be measurable and consistent with
the quality policy.
5.4.2 Quality
management system planning: Top management of the manufacturer shall ensure
that—
(a)      the planning of the quality
management system is carried out in order to meet the specified requirements,
as well as the quality objectives; and
(b)      the integrity of the
quality management system is maintained when changes to the quality management
system are planned and implemented.
5.5 Responsibility,
authority and communication—
5.5.1 Responsibility
and authority: Top management of the manufacturer shall ensure that
responsibilities and authorities are defined, documented and communicated
within the manufacturing organisation.
Top management of the
manufacturer shall establish the interrelation of all personnel who manage,
perform and verify work affecting quality, and shall ensure the independence
and authority necessary to perform these tasks.
5.5.2 Management
representative: Top management shall appoint a member of management who,
irrespective of other responsibilities, shall have responsibility and authority
that includes—
(a)      ensuring that processes
needed for the quality management system are established, implemented and
maintained;
(b)      reporting to top management
on the performance of the quality management system and any need for
improvement; and
(c)      ensuring the promotion of
awareness of regulatory and customer requirements throughout the manufacturing
organisation.
5.5.3 Internal
communication: Top management shall ensure that appropriate communication
processes are established within the Manufacturing organisation and that
communication takes place regarding the effectiveness of the quality management
system.
5.6 Management review:
5.6.1 General: Top
management shall review the organisation's quality management system, at
planned intervals, to ensure its continuing suitability, adequacy and
effectiveness. This review shall include assessing opportunities for
improvement and the need for changes to the quality management system,
including the quality policy and quality objectives. Records from management
reviews shall be maintained.
5.6.2 Review input:
The input to management review shall include information on—
(a)      results of audits,
(b)      customer feedback,
(c)      process performance and
product conformity,
(d)      status of preventive and
corrective actions,
(e)      follow-up actions from
previous management reviews,
(f)       changes that could affect
the quality management system,
(g)      recommendations for
improvement, and
(h)     new or revised regulatory
requirements as and when issued.
5.6.3 Review output:
The output from the management review shall include any decisions and actions
related to—
(a)      improvements needed to
maintain the effectiveness of the quality management system and its processes,
(b)      improvement of product
related to customer requirements, and
(c)      resource needs.
6. Resource management:
6.1 Provision of
resources: The manufacturing organisation shall determine and provide the
resources needed—
(a)      to implement the quality
management system and to maintain its effectiveness, and
(b)      to meet regulatory and
customer requirements.
6.2 Human resources:
6.2.1 General:
Personnel performing work affecting product quality shall be competent on the
basis of appropriate education, training, skills and experience. Number of
personnel employed shall be adequate and in direct proportion to the workload.
Prior to employment, all personnel, shall undergo medical examination including
eye examination, and shall be free from communicable or contagious diseases.
Thereafter, they should be medically examined periodically, at least once a
year.
Records shall be maintained
thereof.
6.2.2 Competence, awareness
and training: The manufacturer shall—
(a)      determine the necessary
competence for personnel performing work affecting product quality,
(b)      provide training or take
other actions to satisfy these needs,
(c)      evaluate the effectiveness
of the actions taken,
(d)      ensure that its personnel
are aware of the relevance and importance of their activities and how they
contribute to the achievement of the quality objectives,
(e)      maintain appropriate
records of education, training, skills and experience, and
(f)       establish documented
procedures for identifying training needs and ensure that all personnel are
trained to adequately perform their assigned responsibilities.
6.3 Infrastructure:
The organisation shall
determine, provide and maintain the infrastructure needed to achieve conformity
to product requirements. Infrastructure includes, as applicable—
(a)      buildings, workspace and
associated utilities,
(b)      process equipment (both
hardware and software), and
(c)      supporting services (such
as transport or communication).
The manufacturer shall
establish documented requirements for maintenance activities, including their
frequency, when such activities or lack thereof can affect product quality.
Records of such maintenance shall be maintained.
6.4 Work environment:
The organisation shall
determine and manage the work environment needed to achieve conformity to
product requirements. Following requirements shall apply, namely—
(a)      the manufacturer shall
establish documented requirements for health, cleanliness and clothing of
personnel if contact between such personnel and the product or work environment
could adversely affect the quality of the product;
(b)      if work environment
conditions can have an adverse effect on product quality, the manufacturer
shall establish documented requirements as per Annexure ‘A’ of this Schedule
for the work environment conditions and documented procedures or work
instructions to monitor and control these work environment condition;
(c)      the manufacturer shall
ensure that all personnel who are required to work temporarily under special
environmental conditions within the work environment are appropriately trained
and supervised by a trained person;
(d)      if appropriate, special
arrangements shall be established and documented for the control of contaminated
or potentially contaminated product in order to prevent contamination of other
product, the work environment or personnel;
(e)      all personnel shall bear
clean body covering appropriate to their duties. Smoking, eating, drinking,
chewing or keeping food and drink shall not be permitted in production,
laboratory and storage areas.
7. Product realisation:
7.1 Planning of
product realisation: The manufacturer shall plan and develop the processes
needed for product realisation. Planning of product realisation shall be
consistent with the requirements of the other processes of the quality
management system.
In planning product
realisation, the manufacturer shall determine the following, as appropriate—
(a)      quality objectives and
requirements for the product;
(b)      the need to establish
processes, documents, and provide resources specific to the product;
(c)      required verification,
validation, monitoring, inspection and test activities specific to the product
and the criteria for product acceptance;
(d)      records needed to provide
evidence that the realisation processes and resulting product meet
requirements.
The output of this planning
shall be in a form suitable for the manufacturer's method of operations.
The manufacturer
organisation shall establish documented requirements for risk management (as
per the IS or ISO 14971) throughout product realisation. Records arising from
risk management shall be maintained.
7.2 Customer-related
processes:
7.2.1 Determination of
requirements related to the product: The manufacturer shall determine—
(a)      requirements specified by
the customer, including the requirements for delivery and post-delivery
activities;
(b)      requirements not stated by
the customer but necessary for specified or intended use, where known;
(c)      statutory requirements related
to the product, and
(d)      any additional requirements
determined by the manufacturer.
7.2.2 Review of
requirements related to the product:
The manufacturer shall
review the requirements related to the product. This review shall be conducted
prior to the manufacturer's commitment to supply a product to the customer and
shall ensure that—
(a)      product requirements are
defined and documented;
(b)      contract or order
requirements differing from those previously expressed are resolved; and
(c)      the manufacturer has the ability
to meet the defined requirements.
Records of the results of
the review and actions arising from the review shall be maintained.
Where the customer provides
no documented statement of requirement, the customer requirements shall be
confirmed by the manufacturer before acceptance.
Where product requirements
are changed, the manufacturer shall ensure that relevant documents are amended
and that relevant personnel are made aware of the changed requirements.
7.2.3 Customer
communication:
The manufacturer shall
determine and implement effective arrangements for communicating with customers
in relation to—
(a)      product information;
(b)      enquiries, contracts or
order handling, including amendments;
(c)      customer feedback,
including customer complaints; and
(d)      advisory notices.
7.3 Design and development:
7.3.1 Design and
development planning: The manufacturer shall establish documented procedures
for design and development. The manufacturer shall plan and control the design
and development of product. During the design and development planning, the
manufacturer shall determine—
(a)      the design and development
stages;
(b)      the review, verification,
validation and design transfer activities that are appropriate at each design
and development stage; and
(c)      the responsibilities and authorities
for design and development.
The manufacturer shall
manage the interfaces between different groups involved in design and
development to ensure effective communication and clear assignment of
responsibility.
Planning output shall be
documented, and updated as appropriate, as the design and development
progresses.
Note: Design transfer
activities during the design and development process ensure that design and
development outputs are verified as suitable for manufacturing before becoming
final production specifications.
7.3.2 Design and
development inputs:
Inputs relating to product
requirements shall be determined and records maintained. The design
requirements relating to a device are appropriate and address the intended use
of the device, including the needs of the user and patients.
These inputs shall include—
(a)      functional, performance and
safety requirements, according to the intended use;
(b)      applicable statutory and
regulatory requirements;
(c)      where applicable,
information derived from previous similar designs;
(d)      other requirements
essential for design and development; and
(e)      output(s) of risk
management.
These inputs shall be
reviewed for adequacy and approved by designated individual.
Requirements shall be
complete, unambiguous and not in conflict with each other.
7.3.3 Design and
development outputs:
The outputs of design and
development shall be provided in a form that enables verification against the
design and development input and shall be documented, reviewed, and approved
prior to release.
Design and development
outputs shall—
(a)      meet the input requirements
for design and development;
(b)      provide appropriate
information for purchasing, production and for service provision;
(c)      contain or reference
product acceptance criteria; and
(d)      specify the characteristics
of the product that are essential for its safe and proper use.
Records of the design and
development outputs shall be maintained.
Records of design and
development outputs can include specifications, manufacturing procedures,
engineering drawings, and engineering or research logbooks.
7.3.4 Design and
development review:
At suitable stages,
systematic reviews of design and development shall be performed in accordance
with planned arrangements—
(a)      to evaluate the ability of
the results of design and development to meet requirements; and
(b)      to identify any problems
and propose necessary actions.
Participants in such
reviews shall include representatives of functions concerned with the design
and development stage being reviewed, as well as other specialist personnel.
Records of the results of the reviews and any necessary actions shall be
maintained.
7.3.5 Design and
development verification:
Verification shall be
performed in accordance with planned arrangements to ensure that the design and
development outputs have met the design and development input requirements.
Records of the results of the verification and any necessary actions shall be
maintained.
7.3.6 Design and
development validation:
Design and development
validation shall be performed in accordance with planned arrangements to ensure
that the resulting product is capable of meeting the requirements for the
specified application or intended use.
Design validation shall be
performed under defined operating conditions on initial production units, lots,
or batches or their equivalence. Design validation shall include software
validation and risk analysis, where appropriate validation shall be completed
prior to the delivery or implementation of the product.
Records of the results of
validation and any necessary actions shall be maintained.
As part of design and
development validation, the manufacturer shall perform clinical evaluations
and/or evaluation of performance of the medical device or in vitro diagnostic
medical devices.
Note 1.—If a medical device
or in vitro diagnostic medical devices can only be validated following assembly
and installation at point of use, delivery is not considered to be complete
until the product has been formally transferred to the customer.
Note 2.—Provision of the
medical device for purposes of clinical evaluations and/or evaluation of
performance is not considered to be delivery.
7.3.7 Control of
design and development changes:
Design and development
changes shall be identified and records maintained. The changes shall be
reviewed, verified and validated, as appropriate, and approved before
implementation. The review of design and development changes shall include
evaluation of the effect of the changes on constituent parts and product
already delivered. Records of the results of the review of changes and any
necessary actions shall be maintained.
Note.—Each manufacturer
shall establish and maintain a Design History File for each type of device. The
Design History File shall contain or reference the records necessary to
demonstrate that the design was developed in accordance with the approved
design plan and the requirements of design and development.
7.4 Purchasing:
7.4.1 Purchasing
process: The manufacturer organisation shall establish documented procedures to
ensure that purchased product conforms to specified purchase requirements. The
type and extent of control applied to the supplier and the purchased product
shall be dependent upon the effect of the purchased product on subsequent
product realisation or the final product.
The manufacturer shall
evaluate and select suppliers based on their ability to supply product in
accordance with the manufacturer's requirements. Criteria for selection,
evaluation and re-evaluation shall be established.
Records of the results of
evaluations and any necessary actions arising from the evaluation shall be
maintained.
7.4.2 Purchasing
information: Purchasing information shall describe the product to be purchased,
including where appropriate—
(a)      requirements for approval
of product, procedures, processes and equipment;
(b)      requirements for
qualification of personnel; and
(c)      quality management system
requirements.
The manufacturer shall
ensure the adequacy of specified purchase requirements prior to their
communication to the supplier.
To the extent required for
traceability, the manufacturer shall maintain documents and records of relevant
purchasing information.
7.4.3 Verification of
purchased product: The manufacturer shall establish and implement the
inspection or other activities necessary for ensuring that purchased product
meets specified purchase requirements. Where the manufacturer intends to
perform verification at the supplier's premises, the manufacturer shall state
the intended verification arrangements and method of product release in the
purchasing information. Records of the verification shall be maintained.
7.5 Production and service
provision:
7.5.1 Control of production
and service provision:
7.5.1.1 General
requirements: The manufacturer shall plan and carry out production and service
provision under controlled conditions. Controlled conditions shall include, as
applicable—
(a)      the availability of
information that describes the characteristics of the product;
(b)      the availability of
documented procedures, documented requirements, work instructions; and
reference materials and reference measurement procedures as necessary;
(c)      the use of suitable
equipment;
(d)      the availability and use of
monitoring and measuring devices;
(e)      the implementation of
monitoring and measurement;
(f)       the implementation of
release, delivery and post-delivery activities; and
(g)      the implementation of
defined operations for labeling and packaging.
The manufacturer shall
establish and maintain a record for each batch of medical device or in vitro
diagnostic medical devices that provides traceability and identifies the amount
manufactured and amount approved for distribution. The batch record shall be
verified and approved.
7.5.1.2 Control of
production and service provision — Specific requirements
7.5.1.2.1 Cleanliness of product
and contamination control: The manufacturer shall establish documented
requirements for cleanliness of product if—
(a)      product is cleaned by the
manufacturer prior to sterilisation or its use; or
(b)      product is supplied
non-sterile to be subjected to a cleaning process prior to sterilisation or its
use; or
(c)      product is supplied to be
used non-sterile and its cleanliness is of significance in use; or
(d)      process agents are to be
removed from product during manufacture.
If the product is cleaned
in accordance with clause (a) or clause (b) above, the requirements content in
clause (a) and (b) of sub-paragraph (6.4) do not apply prior to the cleaning
process.
7.5.1.2.2 Installation
activities: If appropriate, the manufacturer shall establish documented
requirements which contain acceptance criteria for installing and verifying the
installation of the medical device or in vitro diagnostic medical devices.
If the agreed customer
requirements allow installation to be performed other than by manufacturer or
its authorised agent, the manufacturer shall provide documented requirements
for installation and verification. Records of installation and verification
performed by the manufacturer or its authorised agent shall be maintained.
7.5.1.3 Particular
requirements for sterile medical devices: The manufacturer shall maintain
records of the process parameters for the sterilisation process which was used
for each sterilisation batch. Sterilization records shall be traceable to each
production batch of medical device.
7.5.2 Validation of
processes for production and service provision:
7.5.2.1 General: The
manufacturer shall validate any processes for production and service provision
where the resulting output cannot be verified by subsequent monitoring or
measurement. This includes any processes where deficiencies become apparent
only after the product is in use. Validation shall demonstrate the ability of
these processes to achieve planned results.
The manufacturer shall
establish arrangements for these processes including, as applicable—
(a)      defined criteria for review
and approval of the processes;
(b)      approval of equipment and
qualification of personnel;
(c)      use of specific methods and
procedures;
(d)      requirements for records;
and
(e)      revalidation.
The manufacturer shall
establish documented procedures for the validation of the application of
computer software (and its changes to such software or its application) for
production and service provision that affect the ability of the product conform
to specified requirements. Such software applications shall be validated prior
to initial use.
Records of validation shall
be maintained.
7.5.2.2 Particular
requirements for sterile medical devices: The manufacturer shall establish
documented procedures for the validation of sterilisation processes. Sterilization
processes shall be validated prior to initial use. The records of validation of
each sterilisation process shall be maintained.
7.5.3 Identification and
traceability:
7.5.3.1 Identification:
The manufacturer shall identify the product by suitable means throughout
product realisation, and shall establish documented procedures for such product
identification. The manufacturer shall establish documented procedures to
ensure that medical devices and in vitro diagnostic medical devices returned to
the manufacturer are identified and distinguished from conforming product.
7.5.3.2 Traceability:
7.5.3.2.1 General: The
manufacturer shall establish documented procedures for traceability. Such
procedures shall define the extent of product traceability and the records
required.
Where traceability is a
requirement, the manufacturer shall control and record the unique
identification of the product.
Note.—Configuration
management is a means by which identification and traceability can be
maintained.
7.5.3.2.2 Particular
requirements for active implantable medical devices and implantable medical
devices: In defining the records required for traceability, the manufacturer
shall include records of all components, materials and work environment conditions,
if these could cause the medical device not to satisfy its specified
requirements.
The manufacturer shall
require that its agents or distributors maintain records of the distribution of
active implantable medical devices and implantable medical devices to allow
traceability and that such records are available for inspection.
Records of the name and
address of the shipping package consignee shall be maintained.
7.5.3.3 Status
identification: The manufacturer shall identify the product status with respect
to monitoring and measurement requirements. The identification of product
status shall be maintained throughout production, storage, implant, usage and
installation of the product to ensure that only product that has passed the
required inspections and tests (or released under an authorised concession) is
despatched, used or installed.
7.5.4 Customer
property: The manufacturer shall exercise care with customer property while it
is under the manufacturer's control or being used by the manufacturer. The
manufacturer shall identify, verify, protect and safeguard customer property
provided for use or incorporation into the product. If any customer property is
lost, damaged or otherwise found to be unsuitable for use, this shall be
reported to the customer and records maintained.
Note.—Customer property can
include intellectual property or confidential health information.
7.5.5 Preservation of
product: The manufacturer shall establish documented procedures or documented
work instructions for preserving the conformity of product during internal
processing and delivery to the intended destination. This preservation shall
include identification, handling, packaging, storage and protection.
Preservation shall also apply to the constituent parts of a product.
The manufacturer shall
establish documented procedures or documented work instructions for the control
of product with a limited shelf-life or requiring special storage conditions.
Such special storage conditions shall be controlled and recorded.
7.6 Control of monitoring
and measuring devices: The manufacturer shall determine the monitoring and
measurement to be undertaken and the monitoring and measuring devices needed to
provide evidence of conformity of product to determined requirements.
The manufacturer shall establish
documented procedures to ensure that monitoring and measurement can be carried
out and are carried out in a manner that is consistent with the monitoring and
measurement requirements.
Where necessary to ensure
valid results, measuring equipment shall be—
(a)      calibrated or verified at
specified intervals, or prior to use, against measurement standards traceable
to Bureau of Indian Standards wherever available; where no such standards
exist, the basis used for calibration or verification shall be recorded;
(b)      adjusted or re-adjusted as
necessary;
(c)      identified to enable the
calibration status to be determined;
(d)      safeguarded from
adjustments that would invalidate the measurement result;
(e)      protected from damage and
deterioration during handling, maintenance and storage.
In addition, the
manufacturer shall assess and record the validity of the previous measuring
results when the equipment is found not to conform to requirements. The
manufacturer shall take appropriate action on the equipment and any product affected.
Records of the results of calibration and verification shall be maintained.
When used in the monitoring
and measurement of specified requirements, the ability of computer software to
satisfy the intended application shall be confirmed. This shall be undertaken
prior to initial use and reconfirmed as necessary.
8. Measurement, analysis
and improvement:
8.1 General: The
manufacturer shall plan and implement the monitoring, measurement, analysis and
improvement processes needed—
(a)      to demonstrate conformity
of the product;
(b)      to ensure conformity of the
quality management system; and
(c)      to maintain the
effectiveness of the quality management system.
This shall include
determination of applicable methods, including statistical techniques, and the
extent of their use.
Note.—If relevant Indian
standards are not available, International standards are applicable. In case no
Indian or International standards are available, validated testing process of
the manufacturer is applicable.
8.2 Monitoring and
measurement:
8.2.1 Feedback: As one
of the measurements of the performance of the quality management system, the
manufacturer shall monitor information relating to whether the manufacturer has
met customer or regulatory requirements. The methods for obtaining and using
this information shall be determined.
The manufacturer shall
establish a documented procedure for a feedback system to provide early warning
of quality problems and for input into the corrective and preventive action
processes.
8.2.2 Internal audit:
The manufacturer shall conduct internal audits at planned intervals to
determine whether the quality management system—
(a)      conforms to the planned
arrangements, to the requirements of this Schedule and to the quality
management system requirements established by the manufacturer; and
(b)      is effectively implemented
and maintained.
An audit programme shall be
planned, taking into consideration the status and importance of the processes
and areas to be audited, as well as the results of previous audits. The audit
criteria, scope, frequency and methods shall be defined.
Selection of auditors and
conduct of audits shall ensure objectivity and impartiality of the audit
process. Auditors shall not audit their own work.
The responsibilities and
requirements for planning and conducting audits, and for reporting results and
maintaining records shall be defined in a documented procedure. The management
responsible for the area being audited shall ensure that actions are taken
without undue delay to eliminate detected non-conformities and their causes.
Follow-up activities shall include the verification of the actions taken and
the reporting of verification results.
8.2.3 Monitoring and
measurement of processes: The manufacturer shall apply suitable methods for
monitoring and, where applicable, measurement of the quality management system
processes. These methods shall demonstrate the ability of the processes to
achieve planned results.
When planned results are
not achieved, correction and corrective action shall be taken, as appropriate,
to ensure conformity of the product.
8.2.4 Monitoring and
measurement of product—
8.2.4.1 General
requirements: The manufacturer shall monitor and measure the characteristics of
the product to verify that product requirements have been met. This shall be
carried out at appropriate stages of the product realisation process in
accordance with the planned arrangements and documented procedures.
Evidence of conformity with
the acceptance criteria shall be maintained. Records shall indicate the
person(s) authorising release of product. Product release shall not proceed
until the planned arrangements have been satisfactorily completed.
8.2.4.2 Particular
requirement for active implantable medical devices and implantable medical
Devices wherever applicable: The manufacturer shall record the identity of
personnel performing any inspection or testing.
8.3 Control of
non-conforming product: The manufacturer shall ensure that product which does
not conform to product requirements is identified and controlled to prevent its
unintended use or delivery. The controls and related responsibilities and
authorities for dealing with non-conforming product shall be defined in a
documented procedure.
The manufacturer shall deal
with non-conforming product by one or more of the following ways:
(a)      by taking action to
eliminate the detected non-conformity;
(b)      by authorising its use,
release or acceptance under concession;
(c)      by taking action to
preclude its original intended use or application.
The manufacturer shall
ensure that non-conforming product is accepted by concession only if regulatory
requirements are met. Records of the identity of the person authorising the
concession shall be maintained.
Records of the nature of
non-conformities and any subsequent actions taken, including concessions
obtained, shall be maintained.
When non-conforming product
is corrected it shall be subject to re-verification to demonstrate conformity
to the requirements. When non-conforming product is detected after delivery or
use has started, the manufacturer shall take action appropriate to the effects,
or potential effects, of the non-conformity.
If product needs to be
reworked (one or more times), the manufacturer shall document the rework
process in a work instruction that has undergone the same authorisation and
approval procedure as the original work instruction. Prior to authorisation and
approval of the work instruction, a determination of any adverse effect of the
rework upon product shall be made and documented.
8.4 Analysis of data:
The manufacturer shall establish documented procedures to determine, collect
and analyse appropriate data to demonstrate the suitability and effectiveness
of the quality management system and to evaluate whether improvement of the
effectiveness of the quality management system can be made.
This shall include data
generated as a result of monitoring and measurement and from other relevant
sources.
The analysis of data shall
provide information relating to—
(a)      feedback;
(b)      conformity to product
requirements;
(c)      characteristics and trends
of processes and products including opportunities for preventive action; and
(d)      suppliers.
Records of the results of
the analysis of data shall be maintained.
8.5 Improvement:
8.5.1 General: The
manufacturer shall identify and implement any changes necessary to ensure and
maintain the continued suitability and effectiveness of the quality management
system through the use of the quality policy, quality objectives, audit
results, analysis of data, corrective and preventive actions and management
review.
The manufacturer shall
establish documented procedures for the issue and implementation of advisory
notices. These procedures shall be capable of being implemented at any time.
Records of all customer complaint investigations shall be maintained. If
investigation determines that the activities outside the manufacturer's
organisation contributed to the customer complaint, relevant information shall
be exchanged between the organisations involved.
If any complaint is not
investigated, justification shall be documented. Any correction or corrective
action resulting from the compliant handling process shall be documented.
Manufacturer shall notify the adverse event to the regulatory authority and
establish documented procedures for the same.
8.5.2 Corrective
action: The manufacturer shall take action to eliminate the cause of
non-conformities in order to prevent recurrence. Corrective actions shall be
appropriate to the effects of the non-conformities encountered. A documented
procedure shall be established to define requirements for—
(a)      reviewing non-conformities
(including customer complaints);
(b)      determining the causes of
non-conformities;
(c)      evaluating the need for
action to ensure that non-conformities do not recur;
(d)      determining and
implementing action needed, including, if appropriate, updating documentation;
(e)      recording of the results of
any investigation and of action taken; and
(f)       reviewing the corrective
action taken and its effectiveness.
8.5.3 Preventive
action: The manufacturer shall determine action to eliminate the causes of
potential non-conformities in order to prevent their occurrence. Preventive
actions shall be appropriate to the effects of the potential problems. A
documented procedure shall be established to define requirements for—
(a)      determining potential
non-conformities and their causes,
(b)      evaluating the need for
action to prevent occurrence of non-conformities,
(c)      determining and
implementing action needed,
(d)      recording of the results of
any investigations and of action taken, and
(e)      reviewing preventive action
taken and its effectiveness.
ANNEXURE
‘A’
[refer
sub-paragraph 6.4(b)]
Environmental
requirement for medical devices and in vitro diagnostic medical devices
| 
   Name of Device  | 
  
   Type of Operation  | 
  
   ISO Class (At rest)  | 
 
| 
   Cardiac stent/Drug Eluting Stent  | 
  
   Primary Packing and Crimping  | 
  
   5  | 
 
| 
      | 
  
   Washing, Ultrasonic cleaning and Drug coating  | 
  
   7  | 
 
| 
      | 
  
   Assembly, Wrapping and Packaging  | 
  
   8  | 
 
| 
      | 
  
   Laser cutting, Descaling, Annealing and Electro
  polishing  | 
  
   9  | 
 
| 
   Heart Valves  | 
  
   Valve Packing  | 
  
   5  | 
 
| 
      | 
  
   Ultrasonic Cleaning and Visual Inspection  | 
  
   7  | 
 
| 
      | 
  
   Frame and Disc Assembly  | 
  
   7  | 
 
| 
   Intra Ocular Lenses  | 
  
   Primary Packing and Sealing  | 
  
   5  | 
 
| 
      | 
  
   Final Inspection  | 
  
   7  | 
 
| 
      | 
  
   Power Checking and Final Cleaning  | 
  
   8  | 
 
| 
      | 
  
   Tumble Polishing and Lathe Cutting  | 
  
   9  | 
 
| 
   Bone Cements  | 
  
   Final Product Filling  | 
  
   5  | 
 
| 
      | 
  
   Sieving and Calcinations  | 
  
   7  | 
 
| 
      | 
  
   Powder Preparation, Granulation and Drying  | 
  
   8  | 
 
| 
   Internal Prosthetic Replacement  | 
  
   Primary Packing  | 
  
   5  | 
 
| 
      | 
  
   Product Preparation  | 
  
   7  | 
 
| 
      | 
  
   Component Preparation  | 
  
   8  | 
 
| 
   Orthopaedic Implants  | 
  
   Cleaning and packaging (to be sterilised in
  factory premises)  | 
  
   7  | 
 
| 
      | 
  
   Cleaning and packaging (Non-Sterile to be
  sterilised in Hospital)  | 
  
   8  | 
 
| 
      | 
  
   Cutting, lathing, and Polishing  | 
  
   9  | 
 
| 
   Catheters/Ablation Device/IV Cannulae/Scalp Vein
  Set/Hypodermic Syringes/Hypodermic Needles/Perfusion Sets  | 
  
   Assembly, Coating, Wrapping and Packing  | 
  
   7  | 
 
| 
   Component Preparation and Cleaning  | 
  
   8  | 
 |
| 
   Moulding  | 
  
   9  | 
 |
| 
   [61][Condoms  | 
  
   Compounding  | 
  
   Well ventilated area with neat and clean
  environment, free from dust and other particulate matter  | 
 
| 
      | 
  
   Moulding  | 
  
   Well ventilated area with neat and clean
  environment, free from dust and other particulate matter  | 
 
| 
      | 
  
   Vulcanising  | 
  
   Normal Air  | 
 
| 
      | 
  
   Primary Packing  | 
  
   Air conditioned]  | 
 
| 
   Intra Uterine Devices  | 
  
   Moulding  | 
  
   Well-ventilated Area with minimum 5 micron filter  | 
 
| 
      | 
  
   Assembling  | 
  
   7  | 
 
| 
      | 
  
   Primary Packaging  | 
  
   7  | 
 
| 
   Tubal ring  | 
  
   Extrusion  | 
  
   7  | 
 
| 
      | 
  
   Cutting and Assembly  | 
  
   7  | 
 
| 
      | 
  
   Primary Packaging  | 
  
   7  | 
 
| 
   Blood bags  | 
  
   Moulding/Extrusion of components  | 
  
   8  | 
 
| 
      | 
  
   Assembly  | 
  
   7  | 
 
| 
      | 
  
   Filing  | 
  
   5  | 
 
| 
   Suture  | 
  
   Extrusion  | 
  
   9  | 
 
| 
      | 
  
   Assembly  | 
  
   8  | 
 
| 
      | 
  
   Primary Packing  | 
  
   8  | 
 
| 
   Staplers  | 
  
   Staple formation  | 
  
   9  | 
 
| 
      | 
  
   Staple assembly  | 
  
   8  | 
 
| 
      | 
  
   Staple Primary pack  | 
  
   8  | 
 
| 
   Ligatures  | 
  
   Extrusion  | 
  
   9  | 
 
| 
      | 
  
   Cutting and assembly  | 
  
   8  | 
 
| 
      | 
  
   Final Primary Packing  | 
  
   8  | 
 
| 
   [62][Sterile surgical dressings  | 
  
   Final Primary Packing  | 
  
   9]  | 
 
| 
   In vitro diagnostic medical devices
  (Kit/Reagents)  | 
  
   Dry, Liquid Reagent Preparation  | 
  
   Well Lighted and Ventilated controlled
  temperature & humidity as per process or product requirement  | 
 
| 
      | 
  
   Coating of sheets, etc.  | 
  
      | 
 
| 
      | 
  
   Assembly and primary packing  | 
  
      | 
 
| 
      | 
  
   Filling  | 
  
   Well Lighted and Ventilated controlled
  temperature and humidity as per process or product requirement. Provision of
  Laminar hood if required, Clean Room Class 8 or Class 9 as per
  product/process requirement  | 
 
| 
      | 
  
   Secondary Packing  | 
  
   Well Lighted and Ventilated controlled
  temperature if required  | 
 
| 
      | 
  
   Storage  | 
  
   As per recommended storage condition of the product  | 
 
SIXTH
SCHEDULE
[See Rules
26(iii), 26(iv), 38(v) and 38(vii)]
Post
approval change
(A)     Changes in respect of
following shall be considered as major change in—
(1)     material of construction;
(2)     design which shall affect
quality in respect of its specifications, indication for use; performance and
stability of the medical device;
(3)     the intended use or
indication for use;
(4)     the method of
sterilisation;
(5)     the approved Shelf life;
(6)     the name or address of—
(i)       the domestic manufacturer
or its manufacturing site;
(ii)      overseas manufacturer or
its manufacturing site (for import only);
(iii)     authorised agent (for
import only);
(7)     label excluding change in
font size, font type, color, label design;
(8)     manufacturing process,
equipment or testing which shall affect quality of the device;
(9)     primary packaging material.
(B)     Changes in respect of
following shall be considered as minor change in—
(1)     design which shall not
affect quality in respect of its specifications, indication for use,
performance and stability of the medical device;
(2)     in the manufacturing
process, equipment, or testing which shall not affect quality of the device;
(3)     packaging specifications
excluding primary packaging material.
SEVENTH
SCHEDULE
[See Rules
51(1), 51(2), 53(ii), 53(v), 59(3)]
Requirements
for permission to import or manufacture investigational medical device for
conducting clinical investigation
1. Application for
permission—
(1)     an application in Form
MD-22 shall be made to the Central Licensing Authority along with following
data in accordance with tables, namely—
(i)       Design analysis data as per
Table 1.
(ii)      Biocompatibility and Animal
Performance Study as per Table 2.
(iii)     Information specified in
Table 3 shall be submitted along with Investigator's Brochure as prescribed in
Table 4, Clinical Investigational Plan as prescribed in Table 5, Case Report
Form as prescribed in Table 6, Serious adverse event reported, if any, as
prescribed in Table 7, Informed Consent Form as prescribed in Table 8,
investigator's undertaking as prescribed in Table 9, of this schedule and Ethics
Committee approval, if available, as prescribed in Appendix VIII of Schedule Y
of the Drugs and Cosmetics Rules, 1945.
(iv)    Regulatory status in other
countries, including information in respect of restrictions imposed, if any, on
use of investigational medical device in other countries, prescription based
device, exclusion of certain age groups, warning about adverse device effect.
Likewise, if the investigational medical device has been withdrawn in any
country by the manufacturer or by regulatory authority, such information shall
also be furnished along with reasons and its relevance, if any. This
information must continue to be submitted by the sponsor to the Central
Licensing Authority during the entire duration of marketing of the said medical
device in the Country.
(v)      Proposed [63][instructions
for use or electronic instructions for use] or direction for use and labels
shall be submitted as part of the application. The drafts of label shall comply
with provisions of labeling rules specified in Medical Devices Rules, 2017:
Provided that after
submission and approval by the Central Licensing Authority, no change in
the [64][instructions
for use or electronic instructions for use] shall be effected without such
changes having been approved by the Central Licensing Authority;
(vi)    Report of clinical
investigation should be in consonance with the format as prescribed in Table
10, such reports shall be certified by Principal Investigator.
(2)     For investigational medical
device developed in India, clinical investigation is required to be carried out
in India right from Pilot clinical investigation or first in human study and
data generated should be submitted.
(3)     For investigational medical
devices developed and studied in country other than India, Pilot Clinical
Investigation or relevant clinical study data should be submitted along with
the application. After submission of such data generated outside India to the
Central Licensing Authority, permission may be granted to repeat pilot study or
to conduct Pivotal Clinical Investigation. Pivotal Clinical Investigation is
required to be conducted in India before permission to market the medical
device in India except investigational medical device classified under Class A,
in exceptional cases, the Central Licensing Authority, may, for reasons to be
recorded in writing, if consider it necessary, mandate conduct of clinical
investigation, depending on the nature of the medical device.
(4)     The number of study
subjects and sites to be involved in the conduct of clinical investigation
shall depend on the nature and objective of the clinical investigation.
2. Clinical Investigation:
(1)     Approval for clinical
investigation—
(i)       Clinical investigation on
an investigational medical device shall be initiated only after approval has
been obtained from the Ethics Committee(s), registered under Rule 122-DD of
Drugs and Cosmetics Rules, 1945, and permission granted by Central Licensing
Authority. The investigation shall be initiated at the respective sites only
after obtaining such approval from the Ethics Committee for that site.
(ii)      All investigators should
possess appropriate qualification, training and experience and should have
access to such investigational and treatment facilities as are relevant to the
proposed clinical investigation. A qualified physician (or dentist, when
appropriate), who is an investigator or a sub-investigator for the
investigation, shall be responsible for all investigation related decisions
concerning medical or dental issues. Laboratories used for generating data for
clinical investigation should be compliant with Good Laboratory Practices or
should have accreditation certificate issued by National Accreditation Board
for Testing and Calibration Laboratories. In all cases, information about
laboratory or facility to be used for the investigation, if other than those at
the investigation site, should be furnished to the Central Licensing Authority
prior to initiation of investigation at such site.
(iii)     Clinical investigational
plan amendments, if it becomes necessary, to so amend it, before initiation or
during the course of a clinical investigation, shall be notified to the Central
Licensing Authority in writing along with approval of the Ethics Committee, if
available, which has granted the approval for the study. No deviations from or
changes to clinical investigational plan shall be implemented without prior
written approval of the Ethics Committee and the Central Licensing Authority
except when it is necessary to eliminate immediate hazards to the study subject
or when changes involve only logistic or administrative aspects of
investigation. All such exceptions shall be immediately notified to the Ethics
Committee as well as to the Central Licensing Authority within 30 days.
(2)     Responsibilities of
Sponsor:
(i)       The sponsor is responsible
for implementing and maintaining quality assurance system to ensure that the
clinical investigation is designed, conducted, monitored, and that data is
generated, documented, recorded and reported in compliance with clinical
investigational plan and Good Clinical Practices (GCP) Guidelines issued by the
Central Drugs Standards Control Organisation, Directorate General of Health
Services, Government of India and applicable rules.
(ii)      The Sponsor is required to
submit a status report on Clinical Investigation to the Central Licensing
Authority, at the prescribed periodicity including safety summary and
deviations.
(iii)     Report of any serious
adverse event occurring during clinical investigation, after due analysis,
shall be forwarded by the sponsor to the Chairman of the Ethics Committee,
Central Licensing Authority, and the Head of institution where the clinical
investigation has been conducted within 14 calendar days from knowledge of
occurrence of the serious adverse event as prescribed in Table 7 of this
schedule.
(iv)    In case of injury or death
occurring to the clinical investigation subject, the sponsor or his
representative whosoever, had obtained permission from the Central Licensing
Authority for conduct of clinical investigation, shall make payment for medical
management of the subject and also provide financial compensation for clinical
investigation related injury or death in the manner as specified in the Drugs
and Cosmetics Rules, 1945.
(v)      The sponsor or his
representative, whosoever, had obtained permission from the Central Licensing
Authority for conduct of clinical investigation shall submit details of
compensation paid for clinical investigation related injury or death to the
Central Licensing Authority within thirty days of the receipt of the order from
Central Licensing Authority.
(vi)    The sponsor shall ensure
that the clinical investigation report, whether for a completed or prematurely
terminated clinical investigation, is provided to the Ethics Committee,
participating investigators and to the Central Licensing Authority.
(vii)   In case, an investigation
need to be discontinued prematurely for any reason including lack of commercial
interest, the sponsor shall need to inform to the Central Licensing Authority
and also submit summary report within a period of ninety days having a
description of the investigation, the number of patients exposed to the
investigational medical device, details of adverse device affect or serious
adverse event, compensation paid, if any, and the reason for discontinuation of
the investigation or non-pursuit of the investigational medical device
application.
(3)     Responsibilities of the
Investigator:
(i)       The investigator shall be
responsible for the conduct of the investigation according to clinical
investigation plan, GCP guidelines and also for compliance as per the
undertaking by the investigator as given in Table 9 of this schedule. Standard
operating procedures are required to be documented by the investigators for the
tasks performed by them. During and following a subject's participation in an
investigation, the investigator should ensure that adequate medical care is
provided to the participant for any adverse events. Investigator shall report
all serious adverse events to the Central Licensing Authority, sponsor or his
representative, whosoever had obtained permission from the Central Licensing
Authority for conduct of the clinical investigation, and the Ethics Committee
that accorded approval to the clinical investigation plan, within forty eight
hours of their occurrence. In case the Investigator fails to report any serious
adverse event within the stipulated period, he shall have to furnish the reason
for the delay to the Central Licensing Authority along with the report of the
serious adverse event. The detailed report of the serious adverse event, after
due analysis, shall be forwarded by the Investigator to Chairman of the Ethics
Committee, Central Licensing Authority and the head of the Institution where
investigation has been conducted within fourteen calendar days of occurrence of
the serious adverse event.
(ii)      The Investigator shall
provide information to the clinical investigation subject through informed
consent process as provided in Table 8 about the essential elements of the
clinical investigation and the subject's right to claim compensation in case of
investigation related injury or death. He shall also inform the subject or
his/her nominee(s) of their rights to contact the Sponsor or his representative
whosoever had obtained permission from the Central Licensing Authority for
conduct of the clinical investigation for making claims in case of
investigation related injury or death.
(4)     Responsibilities of the
Ethics Committee:
(i)       It is the responsibility of
the Ethics Committee that reviews and accords its approval to a Clinical
Investigation Plan to safeguard the rights, safety and well-being of all study
subjects. The Ethics Committee should exercise particular care to protect the
rights, safety and well-being of all vulnerable subjects participating in the
study.
Explanation.—The vulnerable
subject means the members of a group with hierarchical structure (e.g.
prisoners, armed forces personnel, staff and students of medical, nursing and
pharmacy institutions), patients with incurable diseases, unemployed or
impoverished persons, patients in emergency situation, ethnic minority groups,
homeless persons, nomads, refugees, minors or others incapable of personally
giving consent. Ethics committee(s) get documented ‘standard operating
procedures’ and should maintain a record of its proceedings.
(ii)      Ethics Committee(s) shall,
at appropriate intervals, undertake an ongoing review of the investigation of
the Clinical Investigation Plan. Such review may be based on periodic study
progress reports furnished by investigators or monitoring and internal audit
reports furnished by the Sponsor.
(iii)     In case, an Ethics
Committee revokes sites approval accorded to a Clinical Investigation Plan, it
shall record the reasons for doing so and at once, communicate such a decision
to the Investigator as well as to the Central Licensing Authority.
(iv)    Any report of serious
adverse event occurring during clinical investigation, after due analysis,
shall be forwarded by the Chairman of Ethics Committee to the Central Licensing
Authority and to the Head of institution where the clinical investigation has
been conducted within 14 calendar days of the knowledge of occurrence of the
serious adverse event.
(5)     Informed consent:
(i)       In all investigations, a
freely given, informed, written consent is required to be obtained from each
study subject. The investigator shall provide information about the study
verbally and through the patient information sheet, in a language that is
non-technical and is understandable by the study subject. The Subject's consent
must be obtained in writing using an ‘Informed Consent Form’. The patient
information sheet as well as the Informed Consent Form shall be approved by the
Ethics Committee and furnished to the Central Licensing Authority. Any change
in the informed consent documents should be approved by the Ethics Committee
and submitted to the Central Licensing Authority before such changes are
implemented.
(ii)      Where a subject is not able
to give informed consent (e.g. an unconscious person or a minor or those
suffering from severe mental illness or disability), the same may be obtained
from a legally acceptable representative. If the subject or his legally
acceptable representative is unable to read or write, an impartial witness
should be present during the entire informed consent process who must append
his signatures to the consent form.
Explanation.—a legally
acceptable representative means a person who is able to give consent or
authorise an intervention in the patient as provided by the law in India.
(iii)     A checklist of essential
elements to be included in the study subject's informed consent document as
well as a format for the Informed Consent Form for study Subjects is given in
Table 8 of this schedule.
(v)      The informed consent
process, in case of vulnerable subjects in clinical investigations of an
innovative medical device which is not approved anywhere in the world, shall be
audio-video recorded.
(6)     Pilot Clinical Investigation—
(i)       Pilot clinical
investigation is defined as those clinical investigations which are used to
acquire specific essential information about a device before beginning the
pivotal clinical investigation. Pilot clinical investigation is exploratory study
which may be conducted in a few numbers of patients with the disease or
condition being studied before moving to large population and scope that give
insight into the performance and safety of a device but cannot provide
definitive support for specific mechanistic or therapeutic claims.
(ii)      The objectives of a pilot
clinical investigation typically include assessing feasibility (e.g.
preliminary device performance), exploring eligibility criteria and their
practical application for pivotal controlled investigation, ascertaining
potential harm (preliminary safety evaluations), studying device mechanism,
validating a method for determining an outcome measure, using a defined device
mechanism to validate a surrogate outcome measure, and evaluating the logistics
of pivotal investigation for performance.
(iii)     If the application is for
conduct of clinical investigation as a part of multi-national clinical
development of medical device, the number of sites and the patients as well as
justification to conduct such clinical investigation in India shall be provided
to the Central Licensing Authority.
(7)     Pivotal Clinical
Investigation:
(i)       The pivotal clinical
investigation is a definitive study in which evidence is gathered to support
the safety and effectiveness evaluation of the medical device for its intended
use. Pivotal clinical investigation is confirmatory study that may be conducted
in large number of patients with disease or condition being studied and scope
to provide the effectiveness and adverse effects.
(ii)      For investigational medical
device which does not have a predicate medical device but has been approved for
sale or distribution in any country other than India, pivotal studies need to
be carried out primarily to generate evidence of safety and effectiveness of the
medical device in Indian patients when used as recommended in the prescribing
information except in cases of investigational medical device classified under
Class A which shall be governed as per permission of Para 6 above. Prior to
conduct of pivotal clinical investigation in Indian subjects, the Central
Licensing Authority may require making the pilot study data available to assess
whether the pilot data is in conformity to the data already generated outside
the country.
(iii)     If the application is for
conduct of clinical investigation as part of a Global Clinical Investigation of
medical device, the number of sites and patients as well as justification for
undertaking such clinical investigation in India shall be provided to the
Central Licensing Authority.
(8)     Post Marketing Clinical
Investigation: Post marketing clinical investigation is the study other than
surveillance performed after marketing approval has been given to the medical
device in relation to the approved indication. This investigation may not be considered
necessary at the time of medical device approval but may be required by the
Central Licensing Authority for optimising the intended use of the medical
device. Post Marketing Clinical investigation includes additional drug device
interaction, safety studies, investigation designed to support use under the
approved indication e.g. mortality or morbidity studies, etc.
(9)     Studies in special
populations: The clinical investigation data of the medical device is required
to be submitted to support the claim sought to be made for use of medical
device in children, pregnant women, nursing women, elderly patients with renal
or other organ system failure as given below:
(i)       Geriatrics: Geriatrics
patients can be included in pivotal study (and in pilot study at the sponsor's
option) in meaningful numbers, if—
(a)      the disease intended to be
treated is characteristically a disease of aging; or
(b)      the population to be
treated is known to be included in substantial numbers of geriatric patients;
or
(c)      there is specific reason to
expect that conditions common in the elderly are likely to be encountered; or
(d)      the investigational medical
device is likely to alter the geriatric patient's response in regard to safety
or performance compared with that of non-geriatric patient.
(ii)      Paediatrics:
(a)      The timing of pediatric
studies in the medical device development program shall depend on the device,
the type of disease being treated, safety consideration, and the safety and
effectiveness of available treatment.
(b)      The medical device expected
to be used in children; the performance and safety shall be made in the
appropriate age group. When clinical investigation is required to be conducted
in children, it is usually appropriate to begin with older children before
extending the investigation to younger children and then infants.
(c)      If the medical device is
predominantly or exclusively used in paediatric patients, clinical
investigation data should be generated in paediatric population except for
initial safety and performance data, which will usually be obtained in adults
unless such initial safety studies in adults would yield little useful
information or expose them to inappropriate risk.
(d)      If the medical device is
intended to treat serious or life-threatening diseases, occurring in both adults
and paediatric patients, for which there are currently no or limited
therapeutic options, paediatric population may be included in the clinical
investigation early, following assessment of initial safety data and reasonable
evidence of potential benefit. In circumstances where this is not possible,
lack of data has to be justified.
(e)      If the medical device has a
potential for use in paediatric patient, paediatric studies may be conducted.
These studies may be initiated at various stages of clinical development or
after post-marketing surveillance in adults, if a safety concern exists. In
cases where there is limited paediatric data at the time of submission of
application, more data in paediatric patients would be expected after marketing
authorisation for use in children is granted.
(f)       Paediatric subjects are
legally unable to provide written informed consent, and are dependent on their
parents or legal guardian to assume responsibility for their participation in
clinical investigation. Written informed consent shall be obtained from the
parent or legal guardian. However, all paediatric participants shall be
informed to the fullest extent possible about the study in a language and in
terms that they are able to understand. Where appropriate, paediatric participants
should additionally assent to enroll in the study. Mature minors and
adolescents should personally sign and date a separately designed written
consent form. Although a participant's wish to withdraw from a study shall be
respected, there may be circumstances in therapeutic studies for serious or
life-threatening diseases in which, in the opinion of the investigator and
parent or legal guardian, the welfare of a pediatric patient would be
jeopardised by his or her failing to participate in the study. In this
situation, continued parental or legal guardian consent will be sufficient to
allow participation in the study.
(g)      For clinical investigations
conducted in paediatric population, the reviewing Ethics Committee shall
include members who are knowledgeable about pediatric, ethical, clinical and
psychosocial issues.
(iii)     Pregnant or nursing women:
(a)      Pregnant or nursing women
shall be included in clinical investigation only when the medical device is
intended for use by pregnant or nursing women or fetuses or nursing infants and
where the data generated from women who are not pregnant or nursing, would not
be suitable.
(b)      For medical device intended
for use during pregnancy, follow-up data pertaining to a period appropriate for
that medical device on the pregnancy, foetus and child will be required.
3. Post Marketing
Surveillance:
(i)       Subsequent to approval of
an Investigational medical device, it shall be closely monitored for their
clinical safety once they are marketed. The applicants shall furnish Periodic
Safety Update Reports (PSURs) in order to—
(a)      report all the relevant new
information from appropriate sources;
(b)      relate these data to
patient exposure;
(c)      summarise the market
authorisation status in different countries and any significant variations
related to safety; and
(d)      indicate whether changes
will be made to product information in order to optimise the use of the
product.
(i)       One medical device should
be covered in one PSUR. Within the single PSUR separate presentation of data
for different indications or separate population need to be given.
(ii)      All relevant clinical and
non-clinical safety data will cover only the period of the report (interval
data). The PSURs shall be submitted every six months for the first two years
after marketing approval of the medical device. For subsequent two years, the
PSURs need to be submitted annually. The Central Licensing Authority may extend
the total duration for submission of PSURs if it is considered necessary in the
interest of public health. PSURs due for a period must be submitted within
thirty calendar days of the last day of the reporting period. However, all
cases involving suspected unexpected serious adverse event shall be reported to
the licensing authority within fifteen days of initial receipt of information
by the applicant. If marketing of the medical device is delayed by the
applicant after obtaining approval to market, such data will have to be
provided on the deferred basis beginning from the time the medical device is
marketed.
(iii)     New studies specifically
planned or conducted to examine a safety issue should be described in the
PSURs.
(iv)    A PSUR should be structured
as follows:
(a)      Title Page: The title page
of PSUR shall capture the name of the Medical device; reporting interval;
approved indication of Medical devices; date of approval of the medical device;
date of marketing of medical device; licence name and address.
(b)      Introduction: This section
of PSUR shall capture the reporting interval; medical device's intended use,
mode of action or principle of operation, risk class and a brief description of
the approved indication and population.
(c)      Current worldwide marketing
authorization status: This section of PSUR shall capture the brief narrative
overview including details of countries where the device is currently approved
along with date of first approval, date of marketing and if the product was
withdrawn in any of the countries with reasons thereof.
(d)      Actions taken in reporting
interval for safety reasons: This section of PSUR shall include a description
of significant actions related to safety that have been taken during the
reporting interval, related to either investigational uses or marketing
experience by the licence holder, sponsor of a clinical investigation,
regulatory authorities, data monitoring committee, or Ethics Committee.
(e)      Changes to reference safety
information: This section of PSUR shall capture any significant changes to the
reference safety information within the reporting interval. Such changes will
include information relating to contraindications, warnings, precautions,
adverse events, and important findings from ongoing and completed clinical
investigations and significant non-clinical findings.
(f)       Estimated patient exposure:
This section of PSUR shall provide the estimates of the size and nature of the
population exposed to the medical device. Brief descriptions of the method used
to estimate the subject exposure shall be provided in terms of—
(i)       Cumulative and interval
subject exposure in Clinical investigation;
(ii)      Cumulative and interval
patient exposure from Marketing Experience in India;
(iii)     Cumulative and interval
patient exposure from Marketing Experience from the rest of the world.
(g)      Presentation of individual
case histories: This section of PSUR shall include the individual case
information available to a licence holder and provide brief case narrative,
medical history indication treated with suspect medical device, causality
assessment and provide following information:
(i)       Reference prescribing
information
(ii)      Individual cases received
from India
(iii)     Individual cases received
from rest of the world
(iv)    Cumulative and interval
summary tabulations of serious adverse events from clinical investigations
(v)      Cumulative and interval
summary tabulations from post-marketing data sources.
(h)     Studies: This section of
PSUR shall capture the brief summary of clinically important emerging efficacy
or effectiveness and safety findings obtained from the licence holder sponsored
clinical investigation and published safety studies that became available
during the reporting interval which has the potential impact the product safety
information.
(i)       Summaries of Significant
Safety Findings from Clinical investigation during the reporting period
(ii)      Findings from Non-interventional
Studies
(iii)     Findings from Non-Clinical
Studies
(iv)    Findings from Literature
(i)       Other information: This
section of PSUR shall include details about signals and Risk Management Plan,
if any, put in place by the licence holder.
(a)      Signal and risk evaluation:
In this section the licence holder shall provide details of signal and risk
identified during the reporting period and evaluation of signals identified
during the same period.
(b)      Risk Management Plan: In
this section the licence holder shall provide brief details of safety concerns
and necessary action taken by him to mitigate such safety concerns.
(j)       Overall Safety Evaluation:
This section of PSUR shall capture the overall safety evaluation of medical
device based on its risk benefit evaluation for approved indication.
(i)       Summary of Safety Concerns
(ii)      Benefit Evaluation
(iii)     Benefit Risk Analysis
Evaluation
(k)      Conclusion: This section of
PSUR shall provide details on the safety profile of medical device and
necessary action taken by the licence holder in this regards.
(l)       Appendix: The appendix
includes the copy of marketing authorisation in India, copy of prescribing
information, line listings with narrative of Individual Case Safety Reports
(ICSR).
Note: Table means “Table”
given below this Schedule.
Table 1 DESIGN ANALYSIS
DATA
Design Analysis Data for a
medical device shall include the following:
(i)       Physical and Metrological
Standardisation.
(ii)      Design control documents
and a predefined procedure of the medical device at the time of manufacturing.
The Design Analysis should
be carried out in accordance with the Standards as detailed in the Medical
Devices Rules, 2017.
A comprehensive report of
design analysis including the basic design features of the device, drawings,
and tests adapted for design analysis (with specifications) and rationale for
selecting those tests and design control procedures shall be prepared.
Table 2 BIO-COMPATIBILITY
AND ANIMAL PERFORMANCE STUDY FOR INVESTIGATIONAL MEDICAL DEVICE
(1)     Recent version of
ISO-10993, Biological Evaluation of Medical Devices shall be followed for
conducting bio-compatibility study for invasive medical devices. A report of
bio-compatibility study along with rationale for selecting specific tests
carried out should be prepared including conclusion of the study.
(2)     Depending on the nature and
intended use of investigational medical device, device performance for its
actions (including mechanical, electrical, thermal, radiation and any other of
this type) and safety shall be assessed in healthy or diseased animal model
(intended to be treated by such medical device), as appropriate, demonstrating
reaction to active and basic parts of the devices on absolute tissue, local
tissue as well as whole organ, clearly recording local, general and systemic
adverse reactions, risks or potential risks and performance of device in line
with intended use. Wherever possible, histopathology, pathophysiology and path
anatomy shall be carried out.
(3)     If the active component of
device is a drug, data for its animal studies as per Schedule Y of the Drugs
and Cosmetics Rules, 1945 should be submitted.
Table 3 INFORMATION TO BE
SUBMITTED ALONG WITH THE APPLICATION
(1)     Design Analysis data
including, (whichever applicable)—
(a)      design input and design
output documents;
(b)      mechanical and electrical
tests;
(c)      reliability tests;
(d)      validation of software
relating to the function of the device;
(e)      any performance tests;
(f)       ex vivo tests.
(2)     The agreement between the
Sponsor and Principal and coordinating investigator(s).
(3)     Appropriate insurance
certificate, if any.
(4)     Forms for reporting any
adverse event and serious adverse event.
(5)     Report of bio-compatibility
tests along with rationale for selecting these tests including a summary report
and conclusion of the study.
(6)     Results of the risk
analysis.
(7)     Animal Performance study
data.
(8)     Clinical Investigational
Plan, Investigator's Brochure as per Table 4, Case Report Form as per Table 6,
Informed Consent Form as per Table 8, investigator's undertaking and Ethics
Committee clearance.
(9)     Pilot and Pivotal Clinical
Investigation data including that, if any, carried out in other countries.
(10)   Regulatory status and
Restriction on use in other countries, if any, where marketed or approved.
[65][(11) Proposed instructions
for use or electronic instructions for use and labels.]
Table 4 INVESTIGATOR'S
BROCHURE (IB)
(1)     General
1.1 Introduction:
Information Brochure shall contain, as a minimum, information on all topics
listed in this Table.
1.2 Identification of the
IB—
(a)      Name of the investigational
medical device.
(b)      Document reference number,
if any.
(c)      Version or date of the IB.
(d)      Confidentiality statement,
if appropriate.
(e)      Summary of the revision
history in case of amendments, if appropriate.
(f)       A version or issue number
and reference number, if any, with page number and the total number of pages on
each page of the IB.
1.3 Sponsor or
manufacturer: Name and address of the sponsor or manufacturer of the
investigational medical device.
(2)     Investigational medical
device information:
(a)      Summary of literature and
evaluation supporting the rationale for the design and intended use of
investigational medical device.
(b)      Statement concerning
regulatory classification of investigational medical device, if relevant.
(c)      General description of the
investigational medical device and its components including materials used.
(d)      Summary of relevant
manufacturing processes and related validation processes.
(e)      Description of the
mechanism of action of investigational medical device, along with supporting
scientific literature.
(f)       Manufacturer's instructions
for installation and use of investigational medical device, including any
necessary storage and handling requirements, preparation for use and any
intended reuse (e.g. sterilisation), any pre-use safety or performance checks
and any precautions to be taken after use (e.g. disposal), if relevant.
(g)      Description of the intended
clinical performance.
(3)     Pre-clinical testing:
Summary of preclinical testing that has been performed on the investigational
medical device, together with an evaluation of results of such testing
justifying its use in human subjects.
The summary shall include
or, where applicable, refer to the results of:
(a)      design input and design
output documents,
(b)      in vitro tests,
(c)      mechanical and electrical
tests,
(d)      reliability tests,
(e)      validation of software
relating to the function of the device,
(f)       any performance tests,
(g)      ex vivo tests, and
(h)     biological safety
evaluation.
(4)     Existing clinical data:
(a)      Summary of relevant
previous clinical experience with the investigational medical device and with
medical devices that have similar characteristics, including such
characteristics that relate to other indications for use of the investigational
medical device.
(b)      Analysis of adverse device
effects and any history of modification or recall.
(5)     Risk management:
(a)      Summary of the risk
analysis, including identification of residual risks.
(b)      Result of the risk
assessment.
(c)      Anticipated risks,
contra-indications, warnings, etc. for the investigational medical device.
(6)     Regulatory and other
references:
(a)      List of International
Standards, if any, complied with in full or in part.
(b)      Statement of conformity
with national regulations, where appropriate.
(c)      List of references, if
relevant.
Table 5 CLINICAL
INVESTIGATION PLAN
1.1 General:
1.1.1 Introduction:
This document specifies the content of a clinical investigation plan
(hereinafter to be referred as CIP). If the required information is written in
other documentation, for example the IB, such documentation shall be referenced
in the CIP. The content of a CIP and any subsequent amendments shall include all
the topics listed in this document, together with a justification for each
topic if this is not self-explanatory.
1.1.2 Identification of the
clinical investigation plan—
(a)      Title of the clinical
investigation.
(b)      Reference number
identifying the specific clinical investigation, if any.
(c)      Version or date of the CIP.
(d)      Summary of the revision
history in the case of amendments.
(e)      Version or issue number and
reference number, if any, with the page number and the total number of pages on
each page of the CIP.
1.1.3 Sponsor: Name,
address and contact details (e-mail id, phone number, etc.) of the sponsor of
the clinical investigation.
1.1.4 Principal
investigator, coordinating investigator and investigation site—
(a)      Name, address, and
professional position of—
(i)       Principal Investigator,
(ii)      Coordinating investigator,
if appointed
(b)      Name and address of the
investigation site in which the clinical investigation will be conducted.
(c)      Name and address of other
institutions involved in the clinical investigation.
The sponsor shall maintain
an updated list of principal investigators, investigation sites, and
institutions.
1.1.5 Overall synopsis
of the clinical investigation: A summary or overview of the clinical
investigation shall include all the relevant information regarding clinical
investigation design such as inclusion or exclusion criteria, number of
subjects, duration of clinical investigation, follow-up, objective and
endpoint.
1.2 Identification and
description of the investigational medical device:
(a)      Summary description of the
investigational medical device and its intended purpose.
(b)      Details concerning the
manufacturer of the investigational medical device.
(c)      Name or number of the model
or type, including software version and accessories, if any, to permit full,
identification.
(d)      Description as to how
traceability shall be achieved during and after clinical investigation, for
example by assignment of lot numbers, batch numbers or serial numbers.
(e)      Intended purpose of the
investigational medical device in the proposed clinical investigation.
(f)       The populations and
indications for which the investigational medical device is intended.
(g)      Description of
investigational medical device including any materials that will be in contact
with tissues or body fluids. (This shall include details of any medicinal
products, human or animal tissues or their derivatives, or other biologically
active substances.)
(h)     Summary of the necessary
training and experience needed to use the investigational medical device.
(i)       Description of the specific
medical or surgical procedures involved in the use of investigational medical
device.
1.3 Justification for
the design of the clinical investigation: Justification for the design of
clinical investigation, which shall be based on conclusions of the evaluation,
and shall comprise a section on justification for the design of the clinical
investigation and include:
(a)      an evaluation of the
results of the relevant pre-clinical testing or assessment carried out to
justify the use of investigational medical device in human subjects; and
(b)      an evaluation of clinical
data that are relevant to the proposed clinical investigation.
1.4 Risks and benefits
of the investigational medical device and clinical investigation:
(a)      Anticipated clinical
benefits.
(b)      Anticipated adverse device effects.
(c)      Residual risks associated
with investigational medical device, as identified in the risk analysis report.
(d)      Risks associated with
participation in the clinical investigation.
(e)      Possible interactions with
concomitant medical treatments.
(f)       Steps that will be taken to
control or mitigate the risks.
(g)      Risk-to-benefit rationale.
1.5 Objectives and
hypotheses of the clinical investigation:
(a)      Objectives, primary and
secondary.
(b)      Hypotheses, primary and
secondary, to be accepted or rejected by statistical data from the clinical
investigation.
(c)      Claims and intended
performance of investigational medical device that are to be verified.
(d)      Risks and anticipated
adverse device effects that are to be assessed.
1.6 Design of the
clinical investigation:
1.6.1 General:
(a)      Description of the type of
clinical investigation to be performed (e.g. comparative double-blind, parallel
design, with or without a comparator group) with rationale for the choice.
(b)      Description of the measures
to be taken to minimise or avoid bias, including randomisation and blinding or
masking.
(c)      Primary and secondary
endpoints, with rationale for their selection and measurement.
(d)      Methods and timing for
assessing, recording, and analysing variables.
(e)      Equipment to be used for
assessing the clinical investigation variables and arrangements for monitoring
maintenance and calibration.
(f)       Any procedures for
replacement of subjects.
1.6.2 Investigational
medical device and comparator:
(a)      Description of the exposure
to the investigational medical device or comparator, if used.
(b)      Justification of the choice
of comparator.
(c)      List of any other medical
device or medication to be used during clinical investigation.
(d)      Number of investigational
medical devices to be used, together with a justification.
1.6.3 Subjects:
(a)      Inclusion criteria for
subject selection.
(b)      Exclusion criteria for
subject selection.
(c)      Criteria and procedures for
subject withdrawal or discontinuation.
(d)      Point of enrolment.
(e)      Total expected duration of
the clinical investigation.
(f)       Expected duration of each
subject's participation.
(g)      Number of subjects required
to be included in clinical investigation.
(h)     Estimated time needed to
select this number (i.e. enrolment period).
1.6.4 Procedures:
(a)      Description of all the
clinical investigation related procedures that subjects undergo during clinical
investigation.
(b)      Description of those
activities performed by sponsor representatives (excluding monitoring).
(c)      Any known or foreseeable
factors that may compromise the outcome of clinical investigation or
interpretation of results. The follow-up period during clinical investigation
shall permit demonstration of performance over a period of time sufficient to
represent a realistic test of the performance of the investigational medical
device and allow any risks associated with adverse device effects over that
period to be identified and assessed.
The Clinical investigation
plan shall specifically address what medical care, if any, will be provided to
the subjects after the clinical investigation has been completed.
1.6.5 Monitoring plan:
General outline of the monitoring plan to be followed, including access to
source data and the extent of source data verification planned.
1.7 Statistical
considerations: With reference to 1.5 and 1.6, the description of and
justification for—
(a)      statistical design, method
and analytical procedures;
(b)      sample size;
(c)      the level of significance
and the power of the clinical investigation;
(d)      expected drop-out rates;
(e)      pass or fail criteria to be
applied to the results of the clinical investigation;
(f)       the provision for an
interim analysis, where applicable;
(g)      criteria for the
termination of the clinical investigation on statistical grounds;
(h)     procedures for reporting
any deviation from the original statistical plan;
(i)       the specification of
sub-groups for analysis;
(j)       procedures that take into
account all the data;
(k)      the treatment of missing,
unused or spurious data, including drop-outs and withdrawals;
(l)       the exclusion of particular
information from the testing of the hypothesis, if relevant; and
(m)    in multicenter clinical investigations,
the minimum and maximum number of subjects to be included for each center.
Special reasoning and
sample size(s) may apply for the early clinical investigation(s), e.g.
feasibility clinical investigation(s).
1.8 Data management:
(a)      Procedures used for data
review, database cleaning, and issuing and resolving data queries.
(b)      Procedures for
verification, validation and securing of electronic clinical data systems, if
applicable.
(c)      Procedures for data
retention.
(d)      Specified retention period.
(e)      Other aspects of clinical
quality assurance, as appropriate.
1.9 Amendments to the
Clinical investigation plan: Description of the procedures to amend the
Clinical investigation plan.
1.10 Deviations from
clinical investigation plan:
(a)      Statement specifying that
the investigator is not allowed to deviate from the Clinical investigation
plan, except without appropriate notifications or approvals from Ethics
Committee and Central Licensing authority, as the case may be.
(b)      Procedures for recording,
reporting and analysing Clinical investigation plan deviations.
(c)      Notification requirements
and time frames.
(d)      Corrective and preventive
actions and principal investigator disqualification criteria.
1.11 Device
accountability: Description of the procedures for the accountability of
investigational medical devices should be maintained.
1.12 Statements of
compliance:
(a)      Statement specifying that
the clinical investigation shall be conducted in accordance with the ethical
principles as prescribed in Good Clinical Practices.
(b)      Statement specifying that
the clinical investigation shall not begin until the required approval from the
Ethics Committee.
(c)      Statement specifying that
any additional requirements imposed by the Ethics Committee or Central
Licensing Authority shall be followed, if appropriate.
(d)      Statement specifying the
type of insurance that shall be provided for subjects, if appropriate.
1.13 Informed consent
process:
(a)      Description of the general
process for obtaining informed consent, including the process for providing
subjects with new information, as needed.
(b)      Description of the informed
consent process in circumstances where the subject is unable to give it; in the
case of emergency treatment, process should be clearly specified.
1.14 Adverse events,
adverse device effects and device deficiencies:
(a)      Definitions of adverse
events and adverse device effects.
(b)      Definition of device
deficiencies.
(c)      Definitions of serious
adverse events and serious adverse device effects and, where appropriate,
unanticipated serious adverse device effects.
(d)      Time period in which the
principal investigator shall report all adverse events and device deficiencies
to the sponsor and, where appropriate, to Ethics Committee and the regulatory
authority.
(e)      Details of the process for
reporting adverse events (date of the adverse event, treatment, resolution,
assessment of both the seriousness and the relationship to the investigational
medical device).
(f)       Details of the process for
reporting device deficiencies.
(g)      List of foreseeable adverse
events and anticipated adverse device effects, together with their likely
incidence, mitigation or treatment.
(h)     Emergency contact details
for reporting serious adverse events and serious adverse device effects.
1.15 Vulnerable
population:
(a)      Description of the
vulnerable population.
(b)      Description of the specific
informed consent process.
(c)      Description of the Ethics
Committee specific responsibility.
(d)      Description of what medical
care, if any, will be provided for subjects after the clinical investigation
has been completed.
1.16 Suspension or premature
termination of the clinical investigation:
(a)      Criteria and arrangements
for suspension or premature termination of the whole clinical investigation or
of the clinical investigation in one or more investigation sites.
(b)      Criteria for access to and breaking
the blinding or masking code in case of suspension or premature termination of
the clinical investigation, if the clinical investigation involves a blinding
or masking technique.
(c)      Requirements for subject
follow-up.
1.17 Publication
policy:
(a)      Statement indicating
whether the results of the clinical investigation will be submitted for
publication.
(b)      Statement indicating the
conditions under which the results of the clinical investigation will be
offered for publication.
Table 6 CASE REPORT FORM
(CRF)
1. General:
(i)       Case Report Forms are
established to implement the clinical investigation plan, to facilitate subject
observation and to record subject and investigational medical device data
during the clinical investigation according to the clinical investigation plan.
They can exist as printed, optical, or electronic documents and can be
organised into a separate section for each subject.
(ii)      The Case Report Forms
should reflect the clinical investigation plan and take account of the nature
of the investigational medical device.
2. Content and format:
2.1 Overall considerations:
(i)       The Case Report Forms can
be organised such that they reflect all the data from a single procedure or a
single visit or other grouping that makes clinical or chronological sense.
(ii)      The format of Case Report
Forms shall be such as to minimise errors that can be made by those who enter
data and those who transcribe the data into other systems.
(iii)     The data categories and
format listed in this Table can be considered when designing a Case Report
Form.
2.2 Cover page or login
screen:
(1)     Name of sponsor or sponsor
logo.
(2)     Clinical investigation plan
version and date (if required).
(3)     Version number of Case
Report Forms.
(4)     Name of clinical
investigation or reference number (if applicable).
2.3 Header or footer or
Case Report Form identifier:
(a)      Name of the clinical
investigation or reference number.
(b)      Version number of Case
Report Forms.
(c)      Investigation
site/principal investigator identification number.
(d)      Subject identification
number and additional identification such as date of birth or initials, if
allowed by national regulations.
(e)      Case Report Form number or
date of visit or visit number.
(f)       Page/screen number of CRF
and total number of pages/screens (e.g. “page x of xx”).
2.4 Types of Case
Report Forms: The following is a suggested list of CRFs that may be developed
to support a clinical investigation. This is not an exhaustive list and is
intended to be used as a guideline:
(a)      Screening.
(b)      Documentation of subject's
informed consent.
(c)      Inclusion/exclusion.
(d)      Baseline visit:
(1)     demographics;
(2)     medical diagnosis;
(3)     relevant previous
medications or procedures;
(4)     date of enrolment;
(5)     other characteristics.
(e)      Intervention(s) or
treatment(s).
(f)       Follow-up visit(s).
(g)      Clinical investigation
procedure(s).
(h)     Adverse event(s).
(i)       Device deficiencies.
(j)       Concomitant
illness(es)/medication(s).
(k)      Unscheduled visit(s).
(l)       Subject diary.
(m)    Subject withdrawal or lost
to follow-up.
(n)     Form signifying the end of
the clinical investigation, signed by the principal investigator or his/her authorised
designee.
(o)      CIP deviation(s).
3. Procedural issues:
A system shall be
established to enable cross-referencing of CRFs and CIP versions.
Supplemental CRFs may be
developed for collecting additional data at individual investigation sites in
multicenter investigations.
Table 7 DATA ELEMENTS FOR
REPORTING SERIOUS ADVERSE EVENTS OCCURRING IN A CLINICAL INVESTIGATION
1. Patient details:
(a)      Initials and other relevant
identifier (hospital/Out Patient Department's record number, etc.);
(b)      Gender;
(c)      Age and date of birth;
(d)      Weight;
(e)      Height.
2. Suspected device(s):
(a)      Name of the Device;
(b)      Indication(s) for which
suspect device was prescribed;
(c)      Device details including
model number/size/lot number, if applicable;
(d)      Starting date and time of
day;
(e)      Stopping date and time, or
duration of treatment;
3. Other treatment(s):
Provide the same
information for concomitant treatment.
4. Details of suspected
adverse device reaction(s):
(a)      Full description of
reaction(s) including body site and severity, as well as the criterion (or criteria)
for regarding the report as serious. In addition to a description of the
reported signs and symptoms, whenever possible, describe a specific diagnosis
for the reaction.
(b)      Start date (and time) of
onset of reaction.
(c)      Stop date (and time) or
duration of reaction.
(d)      Setting (e.g. hospital,
out-patient clinic, home, nursing home).
5. Outcome:
(a)      Information on recovery and
any sequel; results of specific tests and/or treatment that may have been
conducted.
(b)      For a fatal outcome, cause
of death and a comment on its possible relationship to the suspected reaction;
any post-mortem findings.
(c)      Other information: anything
relevant to facilitate assessment of the case, such as medical history
including allergy, drug or alcohol abuse; family history; findings from special
investigations, etc.
6. Details about the
Investigator:
(a)      Name;
(b)      Address;
(c)      Telephone number;
(d)      Profession (specialty);
(e)      Date of reporting the event
to Central Licensing Authority;
(f)       Date of reporting the event
to Ethics Committee overseeing the site;
(g)      Signature of the
Investigator.
Table 8 INFORMED CONSENT
FORM
Checklist for clinical
investigation Subject's informed consent documents
1.1 Essential elements:
(1)     Statement that the study
involves research and explanation of the purpose of the research.
(2)     Expected duration of the
Subject's participation.
(3)     Description of the
procedures to be followed, including all invasive procedures.
(4)     Description of any
reasonably foreseeable risks or discomforts to the Subject.
(5)     Description of any benefits
to the Subject or others reasonably expected from research. If no benefit is
expected, subject should be made aware of this.
(6)     Disclosure of specific
appropriate alternative procedures or therapies available to the Subject.
(7)     Statement describing the
extent to which confidentiality of records identifying the subject will be
maintained and who will have access to Subject's medical records.
(8)     Clinical investigation
treatment schedule(s) and the probability for random assignment to each
treatment (for randomised clinical investigation).
(9)     Statement describing the
financial compensation and medical management as under:
(a)      In case of an injury
occurring to the subject during the clinical investigation, free medical
management shall be given as long as required or till such time it is
established that the injury is not related to the clinical investigation,
whichever is earlier.
(b)      In the event of an
investigation related injury or death, the Sponsor or his representative,
whoever has obtained permission from the Central Licensing Authority for
conduct of the clinical investigation, shall provide financial compensation for
the injury or death.
(10)   An explanation about whom
to contact for clinical investigation related queries, rights of Subjects and
in the event of any injury.
(11)   The anticipated prorated
payment, if any, to the Subject for participating in the clinical
investigation.
(12)   Subject's responsibilities
on participation in the clinical investigation.
(13)   Statement that
participation is voluntary, that the Subject can withdraw from the clinical
investigation at any time and that refusal to participate will not involve any
penalty or loss of benefits to which the Subject is otherwise entitled.
(14)   Statement that there is a
possibility of failure of investigational medical device to provide intended
therapeutic effect.
(15)   Any other pertinent
information.
1.2 Additional elements,
which may be required:
(a)      Statement of foreseeable
circumstances under which the Subject's participation may be terminated by the
Investigator without the Subject's consent.
(b)      Additional costs to the
Subject that may result from participation in the clinical investigation.
(c)      The consequences of a
Subject's decision to withdraw from the investigation and procedures for
orderly termination of participation by Subject.
(d)      Statement that the Subject
or Subject's representative will be notified in a timely manner if significant
new findings are developed during the course of the investigation which may
affect the Subject's willingness to continue participation will be provided.
(e)      A statement that the
particular treatment or procedure may involve risks to the Subject (or to the
embryo or fetus, if the Subject is or may become pregnant), which are currently
unforeseeable.
(f)       Approximate number of
Subjects enrolled in the clinical investigation.
2. Format of informed
consent form for Subjects participating in a clinical investigation:
Informed Consent form to
participate in a clinical investigation
Clinical investigation
Title:
Clinical investigation
Number:
Subject's Initials:
……………………… Subject's Name: ………………………..
Date of Birth/Age:
…………………….. Gender: …………………………
Address of the Subject:
………………………………
Qualification:
……………………………………………
Occupation:
Student/Self-employed/Service/Housewife/Others (Please tick as appropriate)
Annual income of the
subject: ………………………………………
Name and address of the
nominee(s) and his relation to the subject …………………. (for the purpose of
compensation in case of clinical investigation related death).
Place initial box (Subject)
| 
   (i) I confirm that I have read and
  understood the information sheet dated …….. for the above clinical
  investigation and have had the opportunity to ask questions.  | 
  
   []  | 
 
| 
   (ii) I understand that my
  participation in the clinical investigation is voluntary and that I am free
  to withdraw at any time, without giving any reason, without my medical care
  or legal rights being affected.  | 
  
   []  | 
 
| 
   (iii) I understand that the Sponsor
  of the clinical investigation, others working on the Sponsor's behalf, the
  Ethics Committee and the regulatory authorities will not need my permission
  to look at my health records both in respect of the current clinical
  investigation and any further research that may be conducted in relation to
  it, even if I withdraw from the clinical investigation. I agree to this
  access.  | 
  
   []  | 
 
| 
   However, I understand that my
  identity will not be revealed in any information released to third parties or
  published.  | 
  
      | 
 
| 
   (iv) I agree not to restrict the use
  of any data or results that arise from this clinical investigation provided
  such a use is only for scientific purpose(s).  | 
  
   []  | 
 
| 
   (v) I agree to take part in the above
  clinical investigation.  | 
  
   []  | 
 
| 
   (vi) I understand that in case of an
  injury occurring during the clinical investigation, free medical management
  shall be given as long as required.  | 
  
      | 
 
| 
   (vii) I understand that in the event
  of an investigation related injury or death, financial compensation for such
  injury or death shall be provided in accordance with the provisions of the
  Medical Device Rules, 2017.  | 
  
      | 
 
| 
   Signature (or Thumb impression) of
  the Subject/Legally Acceptable Representative: ……………………  | 
 |
| 
   Date: …………/…………/…………  | 
 |
| 
   Signatory's Name: …………………………….  | 
 |
| 
   Signature of the Investigator:
  …………………………………….  | 
 |
| 
   Contact Details (Telephone
  Number/mobile) on which subject may contact: ……………………….  | 
 |
| 
   Date: …………/…………/…………  | 
 |
| 
   Clinical investigation Investigator's
  Name: …………………………………………  | 
 |
| 
   Signature of the Witness …………………………….
  Date: …………/…………/…………  | 
 |
| 
   Name of the Witness: …………………………………  | 
 |
| 
   Address and contact details of the
  Witness: ……………………………………………….  | 
 |
| 
   (Copy of the Patient Information
  Sheet and duly filled Informed Consent Form shall be handed over to the
  subject or his/her attendant).  | 
 |
Table 9 UNDERTAKING BY THE INVESTIGATOR
(1)     Full name, address and
title of the Principal Investigator (or Investigator(s) when there is no
Principal Investigator).
(2)     Name and address of the
medical college, hospital or other facility where the Clinical Investigation
will be conducted: Education, training & experience that qualify the
Investigator for the clinical investigation [Attach details including medical
council registration number, or any other statement(s) of qualification(s)].
(3)     Name and address of all
clinical facilities to be used in the clinical investigation.
(4)     Name and address of the
Ethics Committee that is responsible for approval and continuing review of the
clinical investigation.
(5)     Names of the other members
of the research team (Co-Investigators or sub-Investigators) who will be
assisting the Investigator in the conduct of the investigation(s).
(6)     Clinical Investigation
Plan, Title and Clinical investigation number (if any) of the clinical
investigation to be conducted by the Investigator.
(7)     Commitments:
(i)       I have reviewed the
clinical investigation plan and agree that it contains all the necessary
information to conduct the investigation. I will not begin the clinical
investigation until all necessary Ethics Committee and regulatory approvals
have been obtained.
(ii)      I agree to conduct the
investigation in accordance with the current Clinical investigation plan. I
will not implement any deviation from or changes of the Clinical investigation
plan without agreement by the Sponsor and prior review and documented
approval/favorable opinion from the Ethics Committee of the amendment, except
where necessary to eliminate an immediate hazard(s) to the clinical
investigation participant or when the change(s) involved are only logistical or
administrative in nature.
(iii)     I agree to personally
conduct and/or supervise the clinical investigation at my site.
(iv)    I agree to inform all
Subjects that the medical devices are being used for investigational purposes
and I will ensure that the requirements relating to obtaining informed consent
and Ethics Committee review and approval specified in this Schedule are met.
(v)      I agree to report to the
Sponsor all adverse experiences that occur in the course of the
investigation(s) in accordance with the regulatory and Good Clinical practice
guidelines.
(vi)    I have read and understood
the information in the Investigator's brochure, including the potential risks
and side effects of the medical device.
(vii)   I agree to ensure that all
associates, colleagues and employees assisting in the conduct of the clinical
investigation are suitably qualified and experienced and they have been
informed about their obligations in meeting their commitments in the clinical
investigation.
(viii)  I agree to maintain
adequate and accurate records and to make those records available for
audit/inspection by the Sponsor, Ethics Committee, Licensing Authority or their
authorised representatives, in accordance with regulatory and provisions of
these rules. I will fully cooperate with any clinical investigation related
audit conducted by regulatory officials or authorised representatives of the
Sponsor.
(ix)    I agree to promptly report
to the Ethics Committee all changes in the CIP activities and all unanticipated
problems involving risks to human Subjects or others.
(x)      I agree to inform all
serious adverse events to the Sponsor, Central Licensing Authority as well as
the Ethics Committee within forty eight hours of their occurrence. In case of
failure, I will submit the justification to the satisfaction of the Central
Licensing Authority. I also agree to report the serious adverse events, after
due analysis, to the Central Licensing Authority, Chairman of the Ethics
Committee and head of the institution where the investigation has been
conducted within fourteen days of the occurrence of serious adverse events.
(xi)    I will maintain
confidentiality of the identification of all participating clinical
investigation patients and assure security and confidentiality of clinical
investigation data.
(xii)   I agree to comply with all
other requirements, guidelines and statutory obligations as applicable to
clinical Investigators participating in clinical Investigations.
| 
   Date:  | 
  
   Signature of Investigator  | 
 
Table 10 CLINICAL INVESTIGATION
REPORT
(1)     General.
This table specifies the
contents of the clinical investigation report that describes the design,
execution, statistical analysis and results of a clinical investigation.
(2)     Cover page.
The page shall contain the
following information—
(a)      title of the clinical
investigation;
(b)      identification of the
investigational medical devices, including names, models, etc. as relevant for
complete identification;
(c)      if not clear from the
title, a single sentence describing the design, comparison, period, usage
method, and subject population;
(d)      name and contact details of
sponsor or sponsor's representative;
(e)      CIP identification/protocol
code;
(f)       name and department of
coordinating investigator and names of other relevant parties, e.g. experts,
biostatistician, laboratory personnel;
(g)      statement indicating
whether the clinical investigation was performed in accordance with declaration
of Helsinki, Good Clinical Practice guidelines and applicable regulations;
(h)     Brief description of
investigation design;
(i)       Start and end date of
patient accrual and names of the sponsor and the participating institutes;
(j)       author(s) of report.
(3)     Table of contents.
The table of contents may
include the following information:
(a)      the page number or locating
information of each section, including summary tables, figures, and graphs;
(b)      a list of appendices and
their location.
(4)     Summary.
The summary may contain the
following items:
(a)      the title of the clinical
investigation;
(b)      an introduction;
(c)      the purpose of the clinical
investigation;
(d)      description of the clinical
investigation population;
(e)      the clinical investigation
method used;
(f)       the results of the clinical
investigation;
(g)      the conclusion;
(h)     the date of the clinical
investigation initiation;
(i)       the completion date of the
clinical investigation or, if the clinical investigation is discontinued, the
date of premature termination.
(5)     Introduction.
The introduction may
contain a brief statement placing the clinical investigation in the context of
the development of the investigational medical device and relating the critical
features of the clinical investigation (e.g. objectives and hypotheses, target
population, treatment and follow-up duration) to that development.
(6)     Investigational medical
device and methods:
6.1 Investigational
medical device description: The description of the investigational medical
device should contain the following points:
(a)      a description of the
investigational medical device;
(b)      the intended use of the
investigational medical device(s);
(c)      previous intended uses or
indications for use, if relevant;
(d)      any changes to the
investigational medical device during the clinical investigation or any changes
from the IB, including—
(i)       raw materials;
(ii)      software;
(iii)     components;
(iv)    shelf-life;
(v)      storage conditions;
(vi)    instructions for use; and
(vii)   other changes.
6.2 Clinical
investigation plan (CIP): A summary of the CIP, including any subsequent
amendment(s) with a rational for each amendment, should be provided. The
summary will include a brief description of the following points—
(a)      the clinical investigation
objectives;
(b)      the clinical investigation
design including—
(i)       the type of clinical
investigation;
(ii)      the clinical investigation
endpoints;
(iii)     the ethical considerations;
(iv)    the data quality assurance;
(v)      the subject population for
the clinical investigation, with the—
(A)     inclusion or exclusion
criteria; and
(B)     sample size; a clear
accounting of all participant who entered the clinical investigation shall be
mentioned. Mention should also be made of all cases that were dropouts or
protocol deviations. Enumerate the patients screened, randomised, and
prematurely discontinued. State reasons for premature discontinuation of
therapy in each applicable case.
(vi)    the treatment and treatment
allocation schedule;
(vii)   any concomitant
medications/treatments;
(viii)  the duration of follow-up;
(ix)    the statistical analysis
including—
(A)     the clinical investigation
hypothesis or pass or fail criteria;
(B)     a sample size calculation;
and
(C)     statistical analysis
methods.
6.3 Ethics Committee:
This section shall document that the clinical investigation was conducted in
accordance with the ethical principles of Declaration of Helsinki. A detailed
description of the Ethics Committee constitution and date(s) of approvals of
investigation documents for each of the participating sites should be provided.
A declaration shall state that EC notifications as per Good Clinical Practice
Guidelines issued by Central Drugs Standard Control Organisation and Ethical
Guidelines for Biomedical Research on Human Subjects, issued by Indian Council
of Medical Research have been followed. The ethics report shall include the
following points:
(a)      a confirmation that the CIP
and any amendments to it were reviewed by the EC;
(b)      a list of all ECs
consulted.
6.4 Clinical
investigation team: Briefly describe the administrative structure of the
clinical investigation (Investigators, site staff, Sponsor/designates, Central
laboratory, etc.).
(7)     Results.
The results should include
the following points:
(a)      the clinical investigation
initiation date;
(b)      the clinical investigation
completion/suspension date;
(c)      the disposal of subjects
and investigational medical devices;
(d)      the subject demographics;
(e)      Clinical investigation Plan
compliance;
(f)       an analysis, which
includes—
(i)       a performance analysis as
provided in the clinical investigation plan;
(ii)      a summary of all adverse
events and adverse device effects, including a discussion of the severity,
treatment needed, resolution and relevant principal investigator's judgment
concerning the causal relationship with the investigational medical devices or
procedure;
(iii)     a table compiling all
observed device deficiencies that could have led to a serious adverse effect,
and any corrective actions taken during the clinical investigation, if any;
(iv)    any needed sub-group
analysis for special populations (i.e. gender, racial/cultural/ethnic
sub-groups), as appropriate;
(v)      an accountability of all
subjects with a description of how missing data or deviation(s) were dealt
within the analysis, including subjects—
(A)     not passing screening
tests;
(B)     lost to follow-up;
(C)     withdrawn or discontinued
from the clinical investigation and the reason.
(8)     Discussion and overall
conclusions.
The conclusions may include
the following points:
(a)      the safety and performance
results and any other endpoints;
(b)      an assessment of risks and
benefits;
(c)      a discussion of the
clinical relevance and importance of the results in the light of other existing
data;
(d)      any specific benefits or
special precautions required for individual subjects or groups considered to be
at risk;
(e)      any implications for the
conduct of future clinical investigations;
(f)       any limitations of the
clinical investigation.
(9)     Abbreviated terms and
definitions.
A list of abbreviated terms
and definitions of specialised or unusual terms should be provided.
(10)   List of appendices to the
clinical investigation report:
(a)      Protocols and amendments;
(b)      Specimen of Case Record
Form;
(c)      Investigators' name(s) with
contact addresses, phone, e-mail, etc.;
(d)      Patient data listings (e)
List of participants treated with investigational product;
(e)      Discontinued participants;
(f)       Protocol deviations;
(g)      CRFs of cases involving
death and life threatening adverse event cases;
(h)     Publications from the
investigation;
(i)       Important publications
referenced in the clinical investigation;
(j)       Audit certificate, if
available;
(k)      List of other parties involved
(e.g. core labs, contract research organisations (CROs), experts, etc.);
(l)       List of monitors involved;
(m)    Investigator's certificate
that he or she has read the report and that the report accurately describes the
conduct and the results of the clinical investigation.
EIGHTH
SCHEDULE
(See Rule
90)
Exemptions
| 
   Sl. No.  | 
  
   Class of medical devices  | 
  
   Extent and conditions of exemption  | 
 
| 
   1.  | 
  
   Custom made device.  | 
  
   All provisions of Chapter IV and Chapter V of
  these rules, subject to the condition that the device is being specifically
  made in accordance with a duly qualified medical practitioner's written
  prescription under his responsibility, in accordance with specific design,
  characteristics and the same is intended for the sole use of a particular
  patient and the label contains the words ‘custom made device’. Explanation.—Mass produced devices, which only
  need adoption to meet the specific requirement of a medical practitioner or
  any other professional user, shall not be considered as custom made device.  | 
 
| 
   2.  | 
  
   Medicated dressings and bandages for first-aid.  | 
  
   The provisions of Chapter XI of these rules which
  require them to be covered by a sale licence, subject to condition that such
  products have been manufactured by licensed manufacturers.  | 
 
| 
   3.  | 
  
   Medical devices supplied by a registered medical
  practitioner to his own patient or any medical device supplied by a
  registered medical practitioner at the request of another such practitioner
  if it is specially prepared with reference to the condition and for the use
  of an individual patient provided the registered medical practitioner is not
  (a) keeping an open shop or (b) selling across the counter, for distribution
  or sale of medical devices in India to a degree which render him liable to
  the provisions of Chapter IV of the Act and the rules made thereunder.  | 
  
   All provisions of Chapter XI of these rules which
  require them to be covered by a sale licence subject to the following
  conditions— (a) The medical devices shall be purchased only
  from a licensed manufacturer or licensed whole seller or retailer under these
  rules and records of such purchases showing the name and quantities of such
  medical devices, together with their batch numbers and names and addresses of
  the manufacturers shall be maintained. Such records shall be open to
  inspection by medical device officer appointed under this Act, who may, if
  necessary make enquiries about purchase of medical device and may also take
  samples for test. (b) Medical device shall be stored under proper
  storage conditions as specified in the label. (c) No medical device shall be sold or supplied
  or dispensed after the date of expiration recorded on its label or in
  violation of any statement or direction recorded on such label.  | 
 
| 
   4.  | 
  
   Medical devices supplied by a hospital or
  dispensary maintained or supported by Government or local body.  | 
  
   All provisions of Chapter XI of these rules which
  requires them to be covered by a sale licence subject to the following
  conditions— (a) The dispensing and supply of medical devices
  shall be carried out by or under the supervision of qualified person; (b) The premises where medical devices are
  supplied or stocked shall be open to inspection by a medical device officer
  appointed under this Act who can, if necessary, take samples for test. (c) The medical devices shall be stored under
  proper storage conditions. (d) The medical devices shall be purchased from a
  manufacturer or a whole seller or retailer licensed under these rules or
  received as transferred stocks from hospital stores for distribution. Records
  of such purchases or receipts shall be maintained. (e) No medical device shall be sold or supplied
  or dispensed after the date of expiration recorded on its label or in
  violation of any statement or direction recorded on such label.  | 
 
| 
   5.  | 
  
   Mechanical contraceptives.  | 
  
   The provisions of Chapter XI of these rules which
  require them to be covered by a sale licence subject to the condition that
  the provisions of condition that no medical device shall be sold or supplied
  or dispensed after the date of expiration recorded on its label or in
  violation of any statement or direction recorded on such label.  | 
 
| 
   6.  | 
  
   Import of small quantity of medical devices
  donated to a charitable hospital for treatment of patients free of cost by
  that hospital.  | 
  
   The provisions of Chapter V of these rules which
  require them to be covered by a licence for import provided that the Central
  Licensing Authority shall issue a No-Objection Certificate for such purpose
  to the applicant.  | 
 
| 
   [66][7.  | 
  
   All medical devices except those specified in the
  Annexure of Eighth Schedule.  | 
  
   All the provisions of these rules subject to the
  condition that such medical devices shall be registered under CHAPTER III-A
  of these rules: Provided that such exemption shall cease after a
  period of thirty months for low risk - Class A and low moderate risk - Class
  B and after a period of forty-two months for moderate high risk - Class C and
  high risk - Class D devices, respectively from the date of this notification.  | 
 
| 
   [67][8.  | 
  
   Manufacturing of Class A non-sterile and
  non-measuring medical devices  | 
  
   All provisions of Chapter IV, VII, VIII and XI of
  these rules, subject to the condition that the manufacturer shall make
  registration of such devices, under the provisions of Chapter IIIB of these
  rules.]  | 
 
| 
   [68][9.  | 
  
   Import of Class A non-sterile and non-measuring
  medical devices  | 
  
   All provisions of Chapter V, VII, VIII and XI of
  these rules, subject to the condition that the importer shall make
  registration of such devices, under the provisions of Chapter IIIB of these
  rules.]  | 
 
ANNEXURE
(See Rule
19A)
| 
   Sl. No.  | 
  
   Name of the device  | 
 
| 
   1.  | 
  
   Disposable Hypodermic Syringes  | 
 
| 
   2.  | 
  
   Disposable Hypodermic Needles  | 
 
| 
   3.  | 
  
   Disposable Perfusion Sets  | 
 
| 
   4.  | 
  
   Substances used for in vitro diagnosis including
  Blood Grouping Sera  | 
 
| 
   5.  | 
  
   Cardiac Stents  | 
 
| 
   6.  | 
  
   Drug Eluting Stents  | 
 
| 
   7.  | 
  
   Catheters  | 
 
| 
   8.  | 
  
   Intra Ocular Lenses  | 
 
| 
   9.  | 
  
   I.V. Cannulae  | 
 
| 
   10.  | 
  
   Bone Cements  | 
 
| 
   11.  | 
  
   Heart Valves  | 
 
| 
   12.  | 
  
   Scalp Vein Set  | 
 
| 
   13.  | 
  
   Orthopedic Implants  | 
 
| 
   14.  | 
  
   Internal Prosthetic Replacements  | 
 
| 
   15.  | 
  
   Ablation Devices  | 
 
| 
   16.  | 
  
   Ligatures, Sutures and Staplers  | 
 
| 
   17.  | 
  
   Intra Uterine Devices (Cu-T)  | 
 
| 
   18.  | 
  
   Condoms  | 
 
| 
   19.  | 
  
   Tubal Rings  | 
 
| 
   20.  | 
  
   Surgical Dressings  | 
 
| 
   21.  | 
  
   Umbilical tapes  | 
 
| 
   22.  | 
  
   Blood/Blood Component Bags  | 
 
| 
   23.  | 
  
   Organ Preservative Solution*  | 
 
| 
   24.  | 
  
   Nebulizer (effective from 1st Jan, 2021)  | 
 
| 
   25.  | 
  
   Blood Pressure Monitoring Device (effective from
  1st Jan, 2021)  | 
 
| 
   26.  | 
  
   Glucometer (effective from 1st Jan, 2021)  | 
 
| 
   27.  | 
  
   Digital Thermometer (effective from 1st Jan,
  2021)  | 
 
| 
   28.  | 
  
   All implantable medical devices Equipment
  (effective from 1st April, 2021)  | 
 
| 
   29.  | 
  
   CT Scan Equipment (effective from 1st April,
  2021)  | 
 
| 
   30.  | 
  
   MRI Equipment (effective from 1st April, 2021)  | 
 
| 
   31.  | 
  
   Defibrillators (effective from 1st April, 2021)  | 
 
| 
   32.  | 
  
   PET Equipment(effective from 1st April, 2021)  | 
 
| 
   33.  | 
  
   X-Ray Machine (effective from 1st April, 2021)  | 
 
| 
   34.  | 
  
   Dialysis Machine (effective from 1st April, 2021)  | 
 
| 
   35.  | 
  
   Bone marrow cell separator (effective from 1st
  April, 2021)  | 
 
| 
   36.  | 
  
   Disinfectants and insecticide specified in
  Medical Devices Rules, 2017  | 
 
| 
   37.  | 
  
   Ultrasound equipment (effective from 1st
  November, 2020)]  | 
 
APPENDIX
Form MD-1
[See sub-rule
(5) of Rule 13]
Application
for grant of Certificate of Registration of a Notified Body
| 
   1.  | 
  
   Name of Applicant:  | 
 
| 
   2.  | 
  
   Nature and constitution of Body:  | 
 
| 
      | 
  
   (i.e. proprietorship, partnership
  including Limited Liability Partnership, private or public company, society,
  trust, other to be specified)  | 
 
| 
   3.  | 
  
   Corporate/registered office address
  including telephone number, mobile number, fax number and e-mail id:  | 
 
| 
   4.  | 
  
   Details of accreditation
  (self-attested copy of certificate to be attached):  | 
 
| 
   5.  | 
  
   Standards (BIS/ISO/Others) for which
  notified body has been accredited under Rule 13:  | 
 
| 
   6.  | 
  
   Fee paid on …………… Rs ……………………
  receipt/challan/transaction id …………….  | 
 
| 
   7.  | 
  
   Documents enclosed, as specified in
  the Part I of the Third Schedule of the Medical Devices Rules, 2017, duly
  signed by me.  | 
 
| 
   8.  | 
  
   I undertake to comply with the
  provisions of the Drugs and Cosmetics Act, 1940 (23 of 1940) and the Medical
  Devices Rules, 2017 and other terms and conditions for working as a Notified
  Body as may be specified from time to time.  | 
 
| 
   Place: ……………….  | 
  
      | 
 
| 
   Date: ………………..  | 
  
   Signature of designated person in
  India  | 
 
| 
      | 
  
   (Name and designation)  | 
 
| 
      | 
  
   [To be signed digitally]  | 
 
Form MD-2
[See sub-rule
(6) of Rule 13]
Certificate
of Registration for a Notified Body under the Medical Devices Rules, 2017
Registration No.: ……………………
| 
   1.  | 
  
   M/s, ………………………………………. (Name of the
  firm) situated at …………………………….(full address with telephone and e-mail) has
  been registered as a Notified Body of following Class A [69][(other
  than non-sterile and non-measuring)]and/or Class B medical devices.  | 
 
| 
   2.  | 
  
   Details of Medical device(s):  | 
 
| 
   Sl. No.  | 
  
   Standards for which it is registered  | 
  
   Class of medical devices  | 
 
| 
      | 
  
      | 
  
      | 
 
| 
   3.  | 
  
   This Registration is subject to the
  conditions as specified in the Drugs and Cosmetics Act, 1940 (23 of 1940) and
  the Medical Devices Rules, 2017.  | 
 
| 
   Place: …………………..  | 
  
   Central Licensing Authority  | 
 
| 
   Date: ……………………  | 
  
   [To be signed digitally]  | 
 
Form MD-3
[See sub-rule
(2) of Rule 20]
Application for Grant of
Licence to Manufacture for Sale and Distribution of Class A[70][(other
than non-sterile and non-measuring)] or Class B medical device
| 
   1.  | 
  
   Name of Applicant:  | 
 
| 
   2.  | 
  
   Nature and constitution of
  manufacturer:  | 
 
| 
      | 
  
   (i.e. proprietorship, partnership
  including Limited Liability Partnership, private or public company, society,
  trust, other to be specified):  | 
 
| 
   3.  | 
  
   (i) Corporate/registered office
  address including telephone number, mobile number, fax number and e-mail id: (ii) Manufacturing site address
  including telephone number, mobile number, fax number and e-mail id: (iii) Address for correspondence: [corporate/registered
  office/manufacturing site]:  | 
 
| 
   4.  | 
  
   Details of medical device(s) to be
  manufactured [Annexed]:  | 
 
| 
   5.  | 
  
   Whether substantial equivalence to a
  predicate device is claimed: (Yes/No):  | 
 
| 
   6.  | 
  
   Fee paid on …………… Rs ……………………
  receipt/challan/transaction id …………….  | 
 
| 
   7.  | 
  
   I have enclosed the documents as
  specified in the Fourth Schedule of Medical Devices Rules, 2017.  | 
 
| 
   8.  | 
  
   I hereby state and undertake that:  | 
 
| 
      | 
  
   (i) the manufacturing site is ready
  for audit or shall be ready for audit on ……………………… in accordance with the
  requirements of Medical Devices Rules, 2017.  | 
 
| 
      | 
  
   (ii) I shall comply with all the
  provisions of the Drugs and Cosmetics Act, 1940 (23 of 1940) and the Medical
  Devices Rules, 2017.  | 
 
| 
   Place: ……………….  | 
  
      | 
 
| 
   Date: ………………..  | 
  
   Signature of designated person in
  India  | 
 
| 
      | 
  
   (Name and designation)  | 
 
| 
      | 
  
   [To be signed digitally]  | 
 
Annexure:
| 
   Sl. No.  | 
  
   Generic name  | 
  
   Model No.  | 
  
   Intended use  | 
  
   Class of medical device  | 
  
   Material of construction  | 
  
   Dimension (if any)  | 
  
   Shelf life  | 
  
   Sterile or Non-sterile  | 
  
   Brand Name (if registered under the Trade Marks
  Act, 1999)  | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
 
Form MD-4
[See sub-rule
(2) of Rule 20]
Application for Grant of
Loan Licence to Manufacture for Sale or for Distribution of Class A [71][(other
than non-sterile and non-measuring)] or Class B medical device
| 
   1.  | 
  
   Name of Applicant:  | 
 
| 
   2.  | 
  
   Nature and constitution of
  manufacturer:  | 
 
| 
      | 
  
   (i.e. proprietorship, partnership
  including Limited Liability Partnership, private or public company, society,
  trust, other to be specified):  | 
 
| 
   3.  | 
  
   (i) Corporate/registered office
  address including telephone number, mobile number, fax number and e-mail id: (ii) Name and address of
  Manufacturing site including telephone number, mobile number, fax number and
  e-mail id: (iii) Address for correspondence: [corporate/registered
  office/manufacturing site]  | 
 
| 
   4.  | 
  
   Details of medical device(s) to be
  manufactured [Annexed]:  | 
 
| 
   5.  | 
  
   Whether substantial equivalence to a
  predicate device is claimed: (Yes/No):  | 
 
| 
   6.  | 
  
   Fee paid on …………………… Rs ……………………
  receipt/challan/transaction id …………………….  | 
 
| 
   7.  | 
  
   I have enclosed the documents as
  specified in the Fourth Schedule of Medical Devices Rules, 2017.  | 
 
| 
   8.  | 
  
   I hereby state and undertake that:  | 
 
| 
      | 
  
   (i) I shall comply with all the
  provisions of the Drugs and Cosmetics Act, 1940 (23 of 1940) and the Medical
  Devices Rules, 2017.  | 
 
| 
   Place: ……………….  | 
  
      | 
 
| 
   Date: ………………..  | 
  
   Signature of designated person in
  India  | 
 
| 
      | 
  
   (Name and designation)  | 
 
| 
      | 
  
   [To be signed digitally]  | 
 
Annexure:
| 
   Sl. No.  | 
  
   Generic name  | 
  
   Model No.  | 
  
   Intended use  | 
  
   Class of medical device  | 
  
   Material of construction  | 
  
   Dimension (if any)  | 
  
   Shelf life  | 
  
   Sterile or Non-sterile  | 
  
   Brand Name (if registered under the Trade Marks
  Act, 1999)  | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
        | 
 
Form MD-5
[See sub-rule
(4) of Rule 20 and sub-rule (6) of Rule 20]
Licence
to Manufacture for Sale or for Distribution of Class A [72][(other
than non-sterile and non-measuring)] or Class B Medical Device.
Licence Number: ……….
(1)     M/s ………………………………………. (Name
and full address of manufacturer with telephone, fax and e-mail) has been
licensed to manufacture for sale or for distribution the below listed medical
device(s) at the premises situated at ……………………………………. (address of manufacturing
facility where the manufacturing will be carried out).
(2)     Details of medical
device(s) [Annexed].
(3)     This licence is subject to
the provisions of the Medical Devices Rules, 2017 and conditions prescribed
therein.
| 
   Place: ………………  | 
  
   State Licensing Authority  | 
 
| 
   Date: ………………  | 
  
   [To be signed digitally]  | 
 
Annexure:
| 
   Sl. No.  | 
  
   Generic name  | 
  
   Model No.  | 
  
   Intended use  | 
  
   Class of medical device  | 
  
   Material of construction  | 
  
   Dimension (if any)  | 
  
   Shelf life  | 
  
   Sterile or Non-sterile  | 
  
   Brand Name (if registered under the Trade Marks
  Act, 1999)  | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
 
Form MD-6
[See sub-rule
(4) of Rule 20 and sub-rule (6) of Rule 20]
Loan
Licence to Manufacture for Sale or for Distribution of Class A [73][(other
than non-sterile and non-measuring)] or Class B medical device
Loan Licence Number: ……….
(1)     M/s …………………………………. (Name
and full address of manufacturer with telephone, fax and e-mail) has been
licensed to manufacture for sale or for distribution the below listed medical
device(s) at the premises situated at ………………………………. (address of manufacturing facility
where the manufacturing will be carried out along with the licence number) C/o
………………………………… (name of manufacturing site licence holder).
(2)     Details of medical
device(s) [Annexed].
(3)     This licence is subject to
the provisions of the Medical Devices Rules, 2017 and conditions prescribed
therein.
| 
   Place: ………………..  | 
  
   State Licensing Authority  | 
 
| 
   Date: ………………  | 
  
   [To be signed digitally]  | 
 
Annexure:
| 
   Sl. No.  | 
  
   Generic name  | 
  
   Model No.  | 
  
   Intended use  | 
  
   Class of medical device  | 
  
   Material of construction  | 
  
   Dimension (if any)  | 
  
   Shelf life  | 
  
   Sterile or Non-sterile  | 
  
   Brand Name (if registered under the Trade Marks
  Act, 1999)  | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
 
Form MD-7
[See sub-rule
(1) of Rule 21 and sub-rule (2) of Rule 21]
Application
for Grant of Licence to Manufacture for Sale or for Distribution of Class C or
Class D
| 
   1.  | 
  
   Name of Applicant:  | 
 
| 
   2.  | 
  
   Nature and constitution of
  manufacturer:  | 
 
| 
      | 
  
   (i.e. proprietorship, partnership
  including Limited Liability Partnership, private or public company, society,
  trust, other to be specified):  | 
 
| 
   3.  | 
  
   (i) Corporate/registered office
  address including telephone number, mobile number, fax number and e-mail id: (ii) Manufacturing site address
  including telephone number, mobile number, fax number and e-mail id: (iii) Address for correspondence: [corporate/registered
  office/manufacturing site]:  | 
 
| 
   4.  | 
  
   Details of medical device(s) to be
  manufactured [Annexed]:  | 
 
| 
   5.  | 
  
   Whether substantial equivalence to a
  predicate device is claimed: (Yes/No):  | 
 
| 
   6.  | 
  
   Fee paid on …………………… Rs ……………………
  receipt/challan/transaction id …………………….  | 
 
| 
   7.  | 
  
   I have enclosed the documents as
  specified in the Fourth Schedule of Medical Devices Rules, 2017.  | 
 
| 
   8.  | 
  
   I hereby state and undertake that:  | 
 
| 
      | 
  
   (i) the manufacturing site is ready
  for audit or shall be ready for audit on ……………………… in accordance with the
  requirements of Medical Devices Rules, 2017. (ii) I shall comply with all the
  provisions of the Drugs and Cosmetics Act, 1940 (23 of 1940) and the Medical
  Devices Rules, 2017.  | 
 
| 
   Place: ……………….  | 
  
   Signature  | 
 
| 
   Date: ………………..  | 
  
   (Name and designation)  | 
 
| 
      | 
  
   [To be signed digitally]  | 
 
Annexure:
| 
   Sl. No.  | 
  
   Generic name  | 
  
   Model No.  | 
  
   Intended use  | 
  
   Class of medical device  | 
  
   Material of construction  | 
  
   Dimension (if any)  | 
  
   Shelf life  | 
  
   Sterile or Non-sterile  | 
  
   Brand Name (if registered under the Trade Marks
  Act, 1999)  | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
 
Form MD-8
[See sub-rule
(1) of Rule 21 and sub-rule (2) of Rule 21]
Application
for Grant of Loan Licence to Manufacture for Sale or for Distribution of Class
C or Class D
| 
   1.  | 
  
   Name of Applicant:  | 
 
| 
   2.  | 
  
   Nature and constitution of
  manufacturer:  | 
 
| 
      | 
  
   (i.e. proprietorship, partnership
  including Limited Liability Partnership, private or public company, society,
  trust, other to be specified):  | 
 
| 
   3.  | 
  
   (i) Corporate/registered office
  address including telephone number, mobile number, fax number and e-mail id: (ii) Manufacturing site address
  including telephone number, mobile number, fax number and e-mail id: (iii) Address for correspondence: [corporate/manufacturing site]:  | 
 
| 
   4.  | 
  
   Details of medical device(s) to be
  manufactured [Annexed]:  | 
 
| 
   5.  | 
  
   Whether substantial equivalence to a
  predicate device is claimed: (Yes/No):  | 
 
| 
   6.  | 
  
   Fee paid on …………………… Rs ……………………
  receipt/challan/transaction id …………………….  | 
 
| 
   7.  | 
  
   I have enclosed the documents as
  specified in the Fourth Schedule of Medical Devices Rules, 2017.  | 
 
| 
   8.  | 
  
   I hereby state and undertake that:  | 
 
| 
      | 
  
   (i) the manufacturing site is ready
  for audit or shall be ready for audit on ……………………… in accordance with the
  requirements of Medical Devices Rules, 2017. (ii) I shall comply with all the
  provisions of the Drugs and Cosmetics Act, 1940 (23 of 1940) and the Medical
  Devices Rules, 2017.  | 
 
| 
   Place: ……………….  | 
  
   Signature  | 
 
| 
   Date: ………………..  | 
  
   (Name and designation)  | 
 
| 
      | 
  
   [To be signed digitally]  | 
 
Annexure:
| 
   Sl. No.  | 
  
   Generic name  | 
  
   Model No.  | 
  
   Intended use  | 
  
   Class of medical device  | 
  
   Material of construction  | 
  
   Dimension (if any)  | 
  
   Shelf life  | 
  
   Sterile or Non-sterile  | 
  
   Brand Name (if registered under the Trade Marks
  Act, 1999)  | 
 
Form MD-9
[See sub-rule
(1) Rule 25]
Licence
to Manufacture for Sale or for Distribution of Class C or Class D
Licence Number: ……….
(1)     M/s ………………………………………. (Name
and full address of manufacturer with telephone, fax and e-mail) has been
licensed to manufacture for sale or for distribution the below listed medical
device(s) at the premises situated at ……………………………………. (address of manufacturing
facility where the manufacturing will be carried out).
(2)     Details of medical
device(s) [Annexed].
(3)     The names, qualifications
and experience of the competent technical staff responsible for the manufacture
and testing of the abovementioned medical device(s).
(4)     This licence is subject to
the provisions of the Medical Devices Rules, 2017 and conditions prescribed
therein.
| 
   Place: …………………  | 
  
   State Licensing Authority  | 
 
| 
   Date: …………………  | 
  
   [To be signed digitally]  | 
 
Annexure:
| 
   Sl. No.  | 
  
   Generic name  | 
  
   Model No.  | 
  
   Intended use  | 
  
   Class of medical device  | 
  
   Material of construction  | 
  
   Dimension (if any)  | 
  
   Shelf life  | 
  
   Sterile or Non-sterile  | 
  
   Brand Name (if registered under the Trade Marks
  Act, 1999)  | 
 
Form MD-10
[See sub-rule
(1) Rule 25]
Loan
Licence to Manufacture for Sale or for Distribution of Class C or Class D
medical device
Loan Licence Number: ……….
(1)     M/s ………………………………………. (Name
and full address of manufacturer with telephone, fax and e-mail) has been
licensed to manufacture for sale or for distribution the below listed medical
device(s) at the premises situated at ……………………………………….. (address of
manufacturing facility where the manufacturing will be carried out along with
the licence number) C/o …………………………… (name of manufacturing site licence holder).
(2)     Details of medical
device(s) [Annexed].
(3)     The names, qualifications
and experience of competent technical staff responsible for the manufacture and
testing of the abovementioned medical device.
(4)     This licence is subject to
the provisions of the Medical Devices Rules, 2017 and conditions prescribed
therein.
| 
   Place: ………………  | 
  
   [74][Central Licensing
  Authority]  | 
 
| 
   Date: ………………  | 
  
   To be signed digitally]  | 
 
Annexure:
| 
   Sl. No.  | 
  
   Generic name  | 
  
   Model No.  | 
  
   Intended use  | 
  
   Class of medical device  | 
  
   Material of construction  | 
  
   Dimension (if any)  | 
  
   Shelf life  | 
  
   Sterile or Non-sterile  | 
  
   Brand Name (if registered under the Trade Marks
  Act, 1999)  | 
 
Form MD-11
[See clause
(vii) of Rule 26]
Form
in which the Audit or Inspection Book shall be maintained
| 
   (A)  | 
  
   The cover of the audit or inspection
  book shall contain the following particulars, namely—  | 
 
| 
      | 
  
   1. The name and address of the
  licensee ………………  | 
 
| 
      | 
  
   2. Licence Number …………………  | 
 
| 
   (B)  | 
  
   (i) The pages of the audit or
  inspection book shall be serially numbered and duly stamped by the Central
  Licensing Authority*/State Licensing Authority*. The pages, other than the
  first and the last pages, shall have the following particulars—  | 
 
| 
      | 
  
   Name and designation of the auditor
  or medical device officer who audited or inspected the premises:  | 
 
| 
      | 
  
   Date of audit or inspection
  ………………………………………………………  | 
 
| 
      | 
  
   Observations of the auditor or
  medical device officer …………………………  | 
 
| 
      | 
  
   Signature of the auditor or medical
  device officer  | 
 
| 
      | 
  
   (ii) The first and last pages of the
  audit or inspection book shall be endorsed by the Central Licensing
  Authority*/State Licensing Authority* with the following words, namely—  | 
 
| 
      | 
  
   Audit or inspection book maintained
  by M/s ……………………………… situated at ……………………………. for licence number ………………. in
  Form ……………………… under the Medical Devices Rules, 2017.  | 
 
| 
      | 
  
   [75]Central Licensing
  Authority/ [76]State Licensing
  Authority [To be signed digitally]  | 
 
Notes:
(i)       Printed copy of the
Inspection Book may be obtained by the licensee from the Licensing Authority on
payment of fee as may be specified by the concerned Licensing Authority from
time to time.
(ii)      The audit or inspection
book shall be maintained at the premises of the licensee.
(iii)     The original copy of
observations made by the auditor or medical device officer shall be maintained
in the premises of the licensee and duplicate copy shall be sent to the Central
Licensing Authority/State Licensing Authority. The triplicate copy shall be
taken as record by the auditor or medical device officer.
Form MD-12
[See sub-rule
(1) of Rule 31]
Application
for licence to manufacture medical device for purpose of clinical
investigations, test, evaluation, examination, demonstration or training
| 
   1.  | 
  
   Name of Applicant:  | 
 
| 
   2.  | 
  
   Nature and constitution of
  manufacturer:  | 
 
| 
      | 
  
   (i.e. proprietorship, partnership
  including Limited Liability Partnership, private or public company, society,
  trust, other to be specified):  | 
 
| 
   3.  | 
  
   (i) Corporate/registered office
  address including telephone number, mobile number, fax number and e-mail id: (ii) Testing or evaluation site
  address including telephone number, mobile number, fax number and e-mail id: (iii) Address for correspondence: [corporate office/testing site]:  | 
 
| 
   4.  | 
  
   Details of medical device(s) to be
  manufactured [Annexed]:  | 
 
| 
   5.  | 
  
   Fee paid on …………… Rs ……………………
  receipt/challan/transaction id …………….  | 
 
| 
   6.  | 
  
   I hereby state and undertake that I
  shall comply with all applicable provisions of the Drugs and Cosmetics Act,
  1940 (23 of 1940) and the Medical Devices Rules, 2017.  | 
 
| 
   Place: ……………….  | 
  
   Signature  | 
 
| 
   Date: ………………..  | 
  
   (Name and designation)  | 
 
| 
      | 
  
   [To be signed digitally]  | 
 
Annexure:
| 
   Sl. No.  | 
  
   Generic name  | 
  
   Class of medical device  | 
  
   Quantity proposed to be manufactured  | 
 
Form MD-13
[See sub-rule
(3) of Rule 31]
Licence
to Manufacture Medical Devices for the Purposes of Clinical Investigations or
Test or Evaluation or Demonstration or Training
(1)     M/s ……………………., of……………, is
hereby licensed to manufacture the medical device(s) specified below for the
purposes of clinical investigations or test or evaluation or demonstration or
training at …………………… (address of the premise).
| 
   Sl. No.  | 
  
   Generic name  | 
  
   Class of medical device  | 
  
   Quantity permitted to be manufactured  | 
 
| 
          | 
  
      | 
  
      | 
  
      | 
 
(2)     This licence is subject to
the provisions of the Medical Devices Rules, 2017 and conditions prescribed
therein.
(3)     This licence shall be in
force for a period of three year from the date specified below.
| 
   Place: ……………………  | 
  
   Central Licensing Authority  | 
 
| 
   Date: ……………………  | 
  
   [To be signed digitally]  | 
 
Form MD-14
[See sub-rule
(1) of Rule 34]
Application
for issue of import licence to import medical device
| 
   1.  | 
  
   Name of Authorised agent:  | 
 
| 
   2.  | 
  
   Nature and constitution of Authorised
  agent:  | 
 
| 
      | 
  
   (i.e. proprietorship, partnership
  including Limited Liability Partnership, private or public company, society,
  trust, other to be specified):  | 
 
| 
   3.  | 
  
   (i) Corporate/registered office
  address including telephone number, mobile number, fax number and e-mail id: (ii) Authorised Agent address
  including telephone number, mobile number, fax number and e-mail id as per
  wholesale licence or manufacturing licence [77][or
  registration certificate]: (iii) Address for correspondence: [corporate/registered
  office/authorised agent]:  | 
 
| 
   4.  | 
  
   Particulars of overseas Manufacturer,
  Manufacturing site(s):  | 
 
| 
   Sl. No.  | 
  
   Name and address of manufacturer (full address
  with telephone, fax and e-mail address of the manufacturer)  | 
  
   Name and address of manufacturing site (full
  address with telephone, fax and e-mail address of the manufacturing site)  | 
 
| 
      | 
  
      | 
  
      | 
 
| 
   5.  | 
  
   Details of medical device(s) to be
  imported [Annexed]:  | 
 
| 
   6.  | 
  
   Whether substantial equivalence to a
  predicate device is claimed: (Yes/No):  | 
 
| 
   7.  | 
  
   Fee paid on …………………… Rs …………………… receipt/challan/transaction
  id …………………….  | 
 
| 
   8.  | 
  
   I have enclosed the documents as
  specified in the Fourth Schedule for grant of licence to import medical
  device(s).  | 
 
| 
   8.  | 
  
   I hereby state and undertake that:  | 
 
| 
      | 
  
   (i) I shall comply with applicable
  provisions of the Drugs and Cosmetics Act, 1940 (23 of 1940) and the Medical
  Devices Rules, 2017.  | 
 
| 
   Place: ……………….  | 
  
   Signature  | 
 
| 
   Date: ………………..  | 
  
   (Name and designation)  | 
 
| 
      | 
  
   [To be signed digitally]  | 
 
Annexure:
| 
   Sl. No.  | 
  
   Generic name  | 
  
   Model No.  | 
  
   Intended use  | 
  
   Class of medical device  | 
  
   Material of construction  | 
  
   Dimension (if any)  | 
  
   Shelf life  | 
  
   Sterile or Non-sterile  | 
  
   Brand Name (if registered under the Trade Marks
  Act, 1999)  | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
              | 
 
Form MD-15
[See sub-rule
(1) of Rule 36]
Licence
to Import Medical Device
Licence No.: …………………..
| 
   1.  | 
  
   M/s ………………………………… (Name, full
  address, as per wholesale licence/manufacturing licence [78][or
  registration certificate], of authorised agent with telephone and e-mail
  address) is hereby licensed to import the medical device(s) manufactured by
  overseas manufacturer having manufacturing site as specified below.  | 
 
| 
   2.  | 
  
   Details of overseas manufacturer and
  manufacturing site under this licence.  | 
 
| 
   Sl. No.  | 
  
   Name and address of overseas manufacturer (full
  address with telephone, fax and e-mail address of the manufacturer)  | 
  
   Name and address of overseas manufacturing site
  (full address with telephone, fax and e-mail address of the manufacturing
  site)  | 
 
| 
              | 
  
      | 
  
      | 
 
| 
   3.  | 
  
   Details of medical device(s)
  [Annexed].  | 
 
| 
   4.  | 
  
   The authorised agent M/s …………………………….
  will be responsible for the business activities of the overseas manufacturer,
  in India in all respects.  | 
 
| 
   5.  | 
  
   This licence is subject to the
  provisions of the Medical Devices Rules, 2017 and conditions prescribed
  therein.  | 
 
| 
   Place: ……………….  | 
  
   Central Licensing Authority  | 
 
| 
   Date: ………………..  | 
  
   Seal or Stamp  | 
 
Annexure:
| 
   Sl. No.  | 
  
   Generic name  | 
  
   Model No.  | 
  
   Intended use  | 
  
   Class of medical device  | 
  
   Material of construction  | 
  
   Dimension (if any)  | 
  
   Shelf life  | 
  
   Sterile or Non-sterile  | 
  
   Brand Name (if registered under the Trade Marks
  Act, 1999)  | 
 
| 
              | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
 
Form MD-16
[See sub-rule
(2) of Rule 40]
Application
for Licence to Import Medical Devices for the Purposes of Clinical
Investigations or Test or Evaluation or Demonstration or Training
| 
   1.  | 
  
   Name of applicant:  | 
 
| 
   2.  | 
  
   Address of applicant including
  telephone number, mobile number, fax number and e-mail id:  | 
 
| 
   3.  | 
  
   Name and Address of device
  Manufacturer:  | 
 
| 
   4.  | 
  
   Name and Address of site(s) where
  test or evaluation is proposed to be conducted:  | 
 
| 
   5.  | 
  
   Details of medical device(s) to be
  imported [Annexed]:  | 
 
| 
   6.  | 
  
   Brief description of the medical
  device:  | 
 
| 
   7.  | 
  
   Purpose of import:  | 
 
| 
   8.  | 
  
   Justification for quantity to be
  imported:  | 
 
| 
   9.  | 
  
   An undertaking stating that required
  facilities including equipment, instrument and personnel have been provided
  to test or evaluate medical device:  | 
 
| 
   10.  | 
  
   An undertaking stating that the
  medical device proposed to be imported to be used exclusively for purpose
  specified at Serial Number 7 and shall not be used for commercial purpose:  | 
 
| 
   11.  | 
  
   Fee paid on ……………………… Rs …………
  receipt/challan/transaction id …………………………………  | 
 
| 
   12.  | 
  
   I hereby state and undertake that, I
  shall comply with all applicable provisions of the Drugs and Cosmetics Act,
  1940 (23 of 1940) and the Medical Devices Rules, 2017.  | 
 
| 
      | 
  
      | 
 
| 
   Place: ……………….  | 
  
   Signature  | 
 
| 
   Date: ………………..  | 
  
   (Name and designation)  | 
 
| 
      | 
  
   (To be signed digitally)  | 
 
Annexure:
| 
   Sl. No.  | 
  
   Name of medical device (Generic and brand)  | 
  
   Model No.  | 
  
   Intended use  | 
  
   Class of medical device  | 
  
   Material of construction  | 
  
   Dimension (if any)  | 
  
   Shelf life  | 
  
   Sterile or Non-sterile  | 
  
   Quantity to be imported  | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
 
Form MD-17
[See sub-rule
(1) of Rule 41]
Licence
to Import Medical Devices for the Purposes of Clinical Investigations or Test
or Evaluation or Demonstration or Training
| 
   1.  | 
  
   M/s ……………………………………… is hereby
  licensed to import the medical device specified below from M/s
  ………………………………….. (Name and full address of overseas manufacturer) for the
  purposes of clinical investigations or test or evaluation or demonstration or
  training at …………………… (Name and address, where clinical investigations or test
  or evaluation or is to be carried out).  | 
 
| 
   Sl. No.  | 
  
   Generic Name  | 
  
   Class of medical device  | 
  
   Quantity permitted to be imported  | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
 
| 
   2.  | 
  
   This licence is subject to conditions
  prescribed under the Medical Devices Rules, 2017.  | 
 
| 
   3.  | 
  
   This licence shall, unless previously
  suspended or revoked, be in force for a period of three year from the date
  specified below—  | 
 
| 
   Place: …………………….  | 
  
   Central Licensing Authority  | 
 
| 
   Date: ……………………..  | 
  
   [To be signed digitally]  | 
 
Form MD-18
[See sub-rule
(1) of Rule 42]
Application
for licence to import investigational medical devices for the purposes by a
government hospital or statutory medical institution for the treatment of
patients
| 
   1.  | 
  
   I ……………………… (Name and designation)
  ………………………. of …………… (Name of the Government Hospital or Statutory Medical
  Institution) hereby apply for a licence to import small quantities of
  investigational medical device specified below manufactured by M/s.…………………
  (Name and full address of overseas manufacturer) for the purpose of treatment
  of patients for the disease ……………………………… (Name of the disease)…………………………….
  at…………………………… (name and address of the hospital).  | 
 
| 
   2.  | 
  
   Details of medical device to be
  imported:  | 
 
| 
   Name of the investigational Medical device  | 
  
   Name and address of the manufacturer  | 
  
   Quantities which may be imported  | 
 
| 
      | 
  
      | 
  
      | 
 
| 
   3.  | 
  
   I shall comply with the provisions of
  the Drugs and Cosmetics Act, 1940 (23 of 1940) and the Medical Devices Rules,
  2017.  | 
 
| 
   4.  | 
  
   A fee of Rs ………………… has been credited
  to the Government under the Head through Challan/receipt No. ………………
  dated…………………… (copy attached).  | 
 
| 
   Place: …………………..  | 
  
   Signature …………………………..  | 
 
| 
   Date: ……………………  | 
  
   Name ………………………………..  | 
 
| 
      | 
  
   Seal or Stamp …………………….  | 
 
Certificate
Certified that the
investigational medical device specified above for import are urgently required
for the treatment of patients suffering from …………………………… and that the said
medical device is not available in India.
| 
   Place: ………………..  | 
  
   Signature …………………………..  | 
 
| 
   Date: ………………….  | 
  
   Medical Superintendent of the
  Government Hospital/Head of Statutory Medical Institution Seal or Stamp  | 
 
Form MD-19
[See sub-rule
(2) of Rule 42]
Licence
to import investigational medical device by a government hospital or statutory
medical institution for the treatment of patients
| 
   Licence No. ………………….  | 
 |
| 
   1.  | 
  
   Dr ………………………………… (Name and
  designation) of …………………….. (Name of Hospital or Statutory Medical
  Institution) here by grant licence to import from M/s ……………………….. (Name and
  full address of manufacturer) the medical devices specified below for the
  purpose of treatment of patients for the disease …………….. (name of the
  disease) at …………………………..(name and address of the hospital).  | 
 
| 
   2.  | 
  
   Details of medical device to be
  imported:  | 
 
| 
   Name of medical device  | 
  
   Quantities which may be imported  | 
 
| 
      | 
  
      | 
 
| 
      | 
  
      | 
 
| 
      | 
  
      | 
 
| 
      | 
  
      | 
 
| 
      | 
  
      | 
 
| 
   3.  | 
  
   This licence shall, unless previously
  suspended or revoked, be in force for a period of one year from the date of
  issue specified above.  | 
 
| 
   Place: ……………………..  | 
  
   Central Licensing Authority  | 
 
| 
   Date: ………………………  | 
  
   Seal or Stamp  | 
 
Form MD-20
[See sub-rule
(2) of Rule 43]
Application
for permission to import small quantity of medical devices for personal use
To
The Central Licensing
Authority,
………………………………………..
………………………………………..
Sir/Madam,
| 
   1.  | 
  
   I ………………………………… resident of………………… by
  occupation……………………….. hereby apply for a permission to import the medical
  device specified below for personal use manufactured by ………………. (Name and
  full address of manufacturer) for the treatment of …………………… (name of the
  disease).  | 
 
| 
   Name of medical device  | 
  
   Quantity which may be imported  | 
 
| 
      | 
  
      | 
 
| 
   2.  | 
  
   The prescription from a registered
  medical practitioner prescribing the need for the said medical device is
  attached.  | 
 
| 
   3.  | 
  
   The particular of the patients is
  specified below.  | 
 
| 
   Name  | 
  
   Age  | 
  
   Gender  | 
  
   Complete Address  | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
 
| 
   Place: ……………………..  | 
  
      | 
 
| 
   Date: ……………………….  | 
  
   Signature of applicant  | 
 
Form MD-21
[See sub-rule
(3) of Rule 43]
Permission
to import of small quantity of medical devices for personal use
| 
   Permit No. ………………………  | 
  
   Date …………………..  | 
 
| 
   1.  | 
  
   ………………………..is hereby permitted to
  import the medical device manufactured by………………………… (Name and full address of
  manufacturer) specified below for personal use.  | 
 
| 
   Name of the medical device  | 
  
   Quantity  | 
 
| 
      | 
  
      | 
 
| 
   2.  | 
  
   This licence is subject to conditions
  prescribed in the Medical Devices Rules, 2017.  | 
 
| 
   3.  | 
  
   This licence shall, unless previously
  suspended or revoked, be in force for a period of one hundred and eighty days
  from the date of issue specified above.  | 
 
| 
   Central Licensing Authority Seal or Stamp  | 
 |
Form MD-22
[See sub-rule
(1) of Rule 51]
Application
for Grant of permission to conduct clinical investigation of an investigational
medical device
| 
   1.  | 
  
   Name of Applicant:  | 
 
| 
   2.  | 
  
   Nature and constitution of applicant:  | 
 
| 
      | 
  
   (i.e. proprietorship, partnership
  including Limited Liability Partnership, company, society, trust, other to be
  specified)  | 
 
| 
   3.  | 
  
   (i) Sponsor address including
  telephone number, mobile number, fax number and e-mail id:  | 
 
| 
      | 
  
   (ii) Clinical investigation site
  address including telephone number, mobile number, fax number and e-mail id:  | 
 
| 
      | 
  
   (iii) Address for correspondence:  | 
 
| 
   4.  | 
  
   Details of investigational medical
  device(s) and Clinical investigation site [Annexed].  | 
 
| 
   5.  | 
  
   Clinical investigation plan number
  with date:  | 
 
| 
   6.  | 
  
   Fee paid on ……………………… Rs ………………………
  receipt/challan/transaction id ……………………  | 
 
| 
   7.  | 
  
   I have enclosed the documents as
  specified in the Seventh Schedule of Medical Devices Rules, 2017.  | 
 
| 
   8.  | 
  
   I hereby state and undertake that:  | 
 
| 
      | 
  
   (i) I shall comply with all the
  provisions of the Drugs and Cosmetics Act, 1940 and the Medical Devices
  Rules, 2017.  | 
 
| 
   Place: …………………..  | 
  
   Central Licensing Authority  | 
 
| 
   Date: ……………………  | 
  
   [To be signed digitally]  | 
 
Annexure:
Details of investigational
medical device(s):
| 
   Sl. No.  | 
  
   Generic name  | 
  
   Intended use  | 
  
   Class of medical device  | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
 
 
| 
   Sl. No.  | 
  
   Name and address of site(s)  | 
  
   Ethics Committee details  | 
  
   Name of Principle Investigator  | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
 
Form MD-23
[See clause
(i) of Rule 52]
Permission
to conduct Clinical Investigation
Permission No. ………………….
| 
   1.  | 
  
   M/s ………………………………… (Name and full
  address) is hereby granted permission to conduct clinical investigation for
  following investigational medical device as per clinical investigation plan
  ……………………….. dated …………………. in the below mentioned clinical investigation
  sites.  | 
 
| 
   2.  | 
  
   Details of investigational medical
  device(s) and Clinical investigation site [Annexed].  | 
 
| 
   3.  | 
  
   This permission is subject to
  conditions as prescribed under Medical Devices Rules, 2017.  | 
 
| 
   Place: …………………..  | 
  
   Central Licensing Authority  | 
 
| 
   Date: ……………………  | 
  
   [To be signed digitally]  | 
 
Details
of investigational medical device(s)
Annexure:
| 
   Sl. No.  | 
  
   Generic name  | 
  
   Intended use  | 
  
   Class of medical device  | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
 
Details of Clinical
investigation site:
| 
   Sl. No.  | 
  
   Name and address of site(s)  | 
  
   Ethics Committee details  | 
  
   Name of Principle Investigator  | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
 
Form MD-24
[See sub-rule
(2) of Rule 59]
Application
for grant of permission to conduct clinical performance evaluation of new in
vitro diagnostic medical device
| 
   1.  | 
  
   Name of Applicant:  | 
 
| 
   2.  | 
  
   Nature and constitution of applicant:  | 
 
| 
      | 
  
   (i.e. proprietorship, partnership
  including Limited Liability Partnership, company, society, trust, other to be
  specified)  | 
 
| 
   3.  | 
  
   (i) Sponsor address including
  telephone number, mobile number, fax number and e-mail id:  | 
 
| 
      | 
  
   (ii) Laboratory(s) or institution(s)
  address including telephone number, mobile number, fax number and e-mail id:  | 
 
| 
      | 
  
   (iii) Address for correspondence:  | 
 
| 
   4.  | 
  
   Details of new in vitro diagnostic
  medical device and laboratory(s) or institution(s) [Annexed].  | 
 
| 
   5.  | 
  
   Clinical performance evaluation plan number
  with date:  | 
 
| 
   6.  | 
  
   Fee paid on ……………………… Rs ………………………
  receipt/challan/transaction id ……………………  | 
 
| 
   7.  | 
  
   I have enclosed the documents as
  specified in the sub-rule (3) of Rule 59 of Medical Devices Rules, 2017.  | 
 
| 
   8.  | 
  
   I hereby state and undertake that:  | 
 
| 
      | 
  
   (i) I shall comply with all the
  provisions of the Drugs and Cosmetics Act, 1940 and the Medical Devices
  Rules, 2017.  | 
 
| 
   Place: …………………..  | 
  
   Signature  | 
 
| 
   Date: ……………………  | 
  
   (Name and designation)  | 
 
| 
      | 
  
   [To be signed digitally]  | 
 
Annexure:
Details
of new in vitro diagnostic medical device
| 
   Sl. No.  | 
  
   Generic name  | 
  
   Intended use  | 
  
   Class of medical device  | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
 
Details
of laboratory(s)/institution(s) involved
| 
   Sl. No.  | 
  
   Name and address of laboratory(s)/institution(s)  | 
  
   Ethics Committee details  | 
  
   Name of Principle Investigator  | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
 
Form MD-25
[See sub-rule
(5) of Rule 59]
Permission
to conduct clinical performance evaluation of new in vitro diagnostic medical
device
Permission No. …………………..
| 
   1.  | 
  
   M/s …………………………………(Name and full
  address of manufacturer with telephone and e-mail) is hereby granted
  permission to conduct clinical performance evaluation of following new in
  vitro diagnostic device as per clinical performance evaluation plan ……………
  dated ……………….on the below mentioned laboratory(s) or institution(s) involved.  | 
 
| 
   2.  | 
  
   Details of new in vitro diagnostic
  medical device and laboratory(s) or institution(s) [Annexed].  | 
 
| 
   3.  | 
  
   This permission is subject to
  conditions as prescribed under Medical Devices Rules, 2017.  | 
 
| 
   Place: …………………..  | 
  
   Central Licensing Authority  | 
 
| 
   Date: ……………………  | 
  
   [To be signed digitally]  | 
 
Annexure:
Details of new in vitro
diagnostic medical device:
| 
   Sl. No.  | 
  
   Generic name  | 
  
   Intended use  | 
  
   Class of medical device  | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
 
Details
of laboratory(s)/institution(s) involved
| 
   Sl. No.  | 
  
   Name and address of laboratory(s)/institution(s)  | 
  
   Ethics Committee details  | 
  
   Name of Principle Investigator  | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
 
Form MD-26
[See sub-rule
(1) of Rule 63]
Application
for grant of permission to import/manufacture for sale or for distribution of
medical device which does not have predicate medical device
| 
   1.  | 
  
   Name of Applicant:  | 
 
| 
   2.  | 
  
   Nature and constitution of applicant:  | 
 
| 
      | 
  
   (i.e. proprietorship, partnership
  including Limited Liability Partnership, company, society, trust, other to be
  specified)  | 
 
| 
   3.  | 
  
   (i) Corporate/registered office
  address including telephone number, mobile number, fax number and e-mail id:  | 
 
| 
      | 
  
   (ii) Manufacturing site/Authorised
  Agent address including telephone number, mobile number, fax number and
  e-mail id as per wholesale licence or manufacturing licence [79][or
  registration certificate]:  | 
 
| 
      | 
  
   (iii) Address for correspondence:  | 
 
| 
      | 
  
   [Corporate/registered
  office/Manufacturing site/authorised agent]  | 
 
| 
   4.  | 
  
   Particulars of Manufacturer,
  Manufacturing site(s):  | 
 
| 
   Sl. No.  | 
  
   Name and address of manufacturer (full address
  with telephone, fax and e-mail address of the manufacturer)  | 
  
   Name and address of manufacturing site (full
  address with telephone, fax and e-mail address of the manufacturing site)  | 
 
| 
          | 
  
      | 
  
      | 
 
| 
   5.  | 
  
   Details of medical device(s) to be
  imported or manufactured [Annexed].  | 
 
| 
   6.  | 
  
   Fee paid on ……………… Rs ……………..
  receipt/challan/transaction id …………………….. .  | 
 
| 
   7.  | 
  
   I have enclosed the documents as
  specified in the Part IV of the Fourth Schedule to the Medical Devices Rules,
  2017.  | 
 
| 
   Place: …………………..  | 
  
   Signature  | 
 
| 
   Date: ……………………  | 
  
   (Name and designation)  | 
 
| 
      | 
  
   [To be signed digitally]  | 
 
Annexure:
| 
   Sl. No.  | 
  
   Generic name  | 
  
   Model No.  | 
  
   Intended use  | 
  
   Class of medical device  | 
  
   Material of construction  | 
  
   Dimension (if any)  | 
  
   Shelf life  | 
  
   Sterile or Non-sterile  | 
 
| 
            | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
 
Form MD-27
[See sub-rule
(2) of Rule 63]
Permission
to import or manufacture for sale or for distribution of medical device which
does not have predicate medical device
Permission No. …………………..
| 
   1.  | 
  
   M/s …………………………………(Name and full
  address of manufacturer with telephone and e-mail) having manufacturing site
  ……………….. (address of manufacturing site), is hereby permitted to
  import/manufacture for sale or for distribution of following medical devices.  | 
 
| 
   2.  | 
  
   Details of medical device(s) to be
  imported or manufactured [Annexed].  | 
 
| 
   3.  | 
  
   This permission is subject to
  conditions as specified in the Drugs and Cosmetics Act (23 of 1940) and the
  Medical Devices Rules, 2017  | 
 
| 
   Place: …………………..  | 
  
   Central Licensing Authority  | 
 
| 
   Date: ……………………  | 
  
   [To be signed digitally]  | 
 
Annexure:
| 
   Sl. No.  | 
  
   Generic name  | 
  
   Brand name  | 
  
   Model No.  | 
  
   Dimension  | 
  
   Intended use  | 
  
   Shelf life  | 
  
   Sterile/Non-sterile  | 
  
   Class of medical device  | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
 
Form MD-28
[See sub-rule
(1) of Rule 64]
Application
for grant of permission to Import or Manufacture for sale or for distribution
of new in vitro diagnostic medical device
| 
   1.  | 
  
   Name of Applicant:  | 
 
| 
   2.  | 
  
   Nature and constitution of applicant:  | 
 
| 
      | 
  
   (i.e. proprietorship, partnership
  including Limited Liability Partnership, company, society, trust, other to be
  specified)  | 
 
| 
   3.  | 
  
   (i) Corporate/registered office
  address including telephone number, mobile number, fax number and e-mail id:  | 
 
| 
      | 
  
   (ii) Manufacturing site/Authorised
  Agent address including telephone number, mobile number, fax number and
  e-mail id as per wholesale licence or manufacturing licence [80][or
  registration certificate]:  | 
 
| 
      | 
  
   (iii) Address for correspondence:  | 
 
| 
      | 
  
   [Corporate/registered
  office/Manufacturing site/authorised agent]  | 
 
| 
   4.  | 
  
   Particulars of Manufacturer,
  Manufacturing site(s):  | 
 
| 
   Sl. No.  | 
  
   Name and address of manufacturer (full address
  with telephone, fax and e-mail address of the manufacturer)  | 
  
   Name and address of manufacturing site (full
  address with telephone, fax and e-mail address of the manufacturing site)  | 
 
| 
      | 
  
      | 
  
      | 
 
| 
   5.  | 
  
   Details of new in vitro diagnostic
  medical device to be imported or manufactured [Annexed].  | 
 
| 
   6.  | 
  
   Fee paid on ………… Rs ………
  receipt/challan/transaction id…………  | 
 
| 
   7.  | 
  
   I have enclosed the documents as
  specified in the Part IV of the Fourth Schedule Medical Devices Rules, 2017.  | 
 
| 
   Place: …………………..  | 
  
   Signature  | 
 
| 
   Date: ……………………  | 
  
   (Name and designation)  | 
 
| 
      | 
  
   [To be signed digitally]  | 
 
Annexure:
| 
   Sl. No.  | 
  
   Generic name  | 
  
   Model No.  | 
  
   Intended use  | 
  
   Class of medical device  | 
  
   Material of construction  | 
  
   Dimension (if any)  | 
  
   Shelf life  | 
  
   Sterile or Non-sterile  | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
 
Form MD-29
[See sub-rule
(2) of Rule 64]
Permission
to Import or Manufacture New In Vitro Diagnostic Medical Device
Permission No. …………………..
| 
   1.  | 
  
   The new in vitro diagnostic medical
  device(s) specified below manufactured by M/s. ………………. (Name and full address
  of manufacturer with telephone, and e-mail) having manufacturing site…………………
  (address of manufacturing site), is hereby permitted to import or
  manufacture.  | 
 
| 
   2.  | 
  
   Details of new in vitro diagnostic
  medical device to be imported or manufactured [Annexed].  | 
 
| 
   3.  | 
  
   This permission is subject to
  conditions as specified in the Drugs and Cosmetics Act, 1940 (23 of 1940) and
  the Medical Devices Rules, 2017.  | 
 
| 
   Place: …………………….  | 
  
   Central Licensing Authority  | 
 
| 
   Date: …………………….  | 
  
   [To be signed digitally]  | 
 
Annexure:
| 
   Sl. No.  | 
  
   Generic name  | 
  
   Brand Name  | 
  
   Model No.  | 
  
   Dimension  | 
  
   Intended Use  | 
  
   Shelf life  | 
  
   Sterile/Non-sterile  | 
  
   Class of medical device  | 
 
| 
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
  
      | 
 
Form MD-30
[See sub-rule
(1) of Rule 67]
Memorandum
to the Central Medical Device Testing Laboratory
Serial Number …………………….
To
The Director,
Central Medical Device
Testing Laboratory,
…………………………………………………..
…………………………………………………..
From ………………………………………….
| 
   1.  | 
  
   I send herewith, under the provisions
  of sub-section (4) of Section 25 of the Drugs and Cosmetics Act, 1940 (23 of
  1940), sample(s) of a medical device purporting to be …………………………….. for test
  or evaluation and request that a report of the result of the test or
  evaluation may be supplied to this Court.  | 
 
| 
   2.  | 
  
   The distinguishing number on the
  packet is …………………………………………..  | 
 
| 
   3.  | 
  
   Particulars of offence alleged
  ………………………  | 
 
| 
   4.  | 
  
   Matter on which opinion is required
  …………………………  | 
 
| 
   5.  | 
  
   A fee of Rs.……………….. has been
  deposited in Court.  | 
 
| 
   Date: ……………………  | 
  
   ……………………  | 
 
| 
      | 
  
   Magistrate  | 
 
Form MD-31
[See sub-rule
(4) of Rule 67]
Certificate
of test or evaluation by the Central Medical Device Testing Laboratory
| 
   1.  | 
  
   Certified that the samples, bearing
  number ………………. purporting to be a sample of ……………………. received on ………………….
  with memorandum No. …………….. dated …………. from…………… has been tested/evaluated
  and that the result of such test/evaluation is as stated below.  | 
 
| 
   2.  | 
  
   The condition of the seals on the
  packet on receipt was as follow.  | 
 
| 
   [81]3.  | 
  
   In the opinion of the undersigned the
  sample is of standard quality/not of standard quality as defined in the Drugs
  and Cosmetics Act, 1940 (23 of 1940) and Medical Devices Rules, 2017 for the
  reasons given below.  | 
 
| 
   Date: …………………….  | 
  
   Director of Central Medical Device
  Testing Laboratory/other Authorised Officer  | 
 
Details
of results of testing or evaluation with protocols of test or evaluation
applied
| 
   Date: …………………….  | 
  
   Director of Central Medical Device
  Testing Laboratory/other Authorised Officer  | 
 
Form MD-32
[See sub-rule
(2) of Rule 68]
Report
of Test or Evaluation of Medical Devices by Medical Device Testing Officer
| 
   1.  | 
  
   It is certified that the samples
  having serial number of memorandum or receipt number ………………. dated …………………
  purporting to be sample of ……………………… received on …………………………… from ……………… has
  been tested or evaluated and the results of tests or evaluation is as stated
  below.  | 
 
| 
   2.  | 
  
   The conditions of seals on the packet
  or on portion of sample or container were as follows ………………  | 
 
| 
   3.  | 
  
   Based upon the test or evaluation and
  in the opinion of undersigned the sample is of standard quality/not of
  standard quality/adulterated/misbranded/spurious, as defined in the Drugs and
  Cosmetics Act, 1940 (23 of 1940) for the reasons given below—  | 
 
| 
   Date: …………………  | 
  
   Medical Device Testing Officer  | 
 
| 
      | 
  
   Seal or Stamp  | 
 
Form MD-33
[See Rule
69]
Application
from a purchaser for test or evaluation of a Medical Device under Section 26 of
the Drugs and Cosmetics Act, 1940 (23 of 1940)
To
The Central Licensing
Authority,
……………………………………………..
……………………………………………..
Sir/Madam,
| 
   1.  | 
  
   Full name and address of the
  applicant ……………………  | 
 
| 
   2.  | 
  
   Occupation ………………………  | 
 
| 
   3.  | 
  
   Name of medical device purporting to
  be contained in the sample ………………………………………………….  | 
 
| 
   4.  | 
  
   Name and full address of the pharmacy
  or concern where the medical device was purchased.  | 
 
| 
   5.  | 
  
   Date on which purchased
  …………………………………………………………………. (invoice attached)  | 
 
| 
   6.  | 
  
   Reasons why the medical device is
  being submitted for test or evaluation ………………………………………  | 
 
| 
   7.  | 
  
   A fee of rupees ……………………….. as
  charged by medical device testing laboratory has been paid under receipt
  number ……………………… dated: ………………  | 
 
 
I hereby declare that the
medical device being submitted for test or evaluation was purchased by or for
me. I further declare that the sample of the medical device being sent for test
or evaluation is exactly as it was purchased and has not been tampered with in
any way to reduce its potency.
| 
   Date: ………………….  | 
  
   Signature Seal or Stamp  | 
 
Form MD-34
[See Rule
72]
Order
under clause (c) of sub-section (1) of section of the Drugs and Cosmetics Act,
1940, (23 of 1940) requiring a person not to dispose of stock in his possession
Whereas, I have reason to
believe that the stocks of medical devices in your possession, detailed below
contravenes the provisions of Section 18 of the Drugs and Cosmetics Act, 1940
(23 of 1940);
Now, therefore, I hereby
require you under clause (c) of sub-section (1) of Section 22 of the said Act,
not to dispose of the said stock for a period of…………………days from the date of
this order.
| 
   Date: ……………….  | 
  
   Medical Device Officer Seal or Stamp  | 
 
Details
of stock of medical devices.
| 
   Date: ……………….  | 
  
   Medical Device Officer Seal or Stamp  | 
 
Form MD-35
[See Rule
74]
Receipt for stock of
medical devices for record, register, document or material object seized under
clause (c) or clause (cc) of sub-section (1) of Section 22 of the Drugs and
Cosmetics Act (23 of 1940)
The stock of medical
devices or records, registers, documents or material objects, detailed below
has/have this day been seized by me under the provisions of clause (c) or
clause (cc) of sub-section (1) of Section 22 of the Drugs and Cosmetics Act,
1940 (23 of 1940), from the premises of ………………………………………….. situated at
……………………………………
| 
   Date: ……………….  | 
  
   Medical Device Officer Seal or Stamp  | 
 
Details
of stock of medical devices or records, registers, documents or material
objects seized.
| 
   Date: ……………….  | 
  
   Medical Device Officer Seal or Stamp  | 
 
Form MD-36
[See Rule
76]
Intimation
of Person from Whom Sample is taken
To
…………………..
I have this day taken from
the premises of ………………. situated at ………………. samples of medical devices
specified below for the purpose of test or evaluation.
| 
   Date: ……………….  | 
  
   Medical Device Officer Seal or Stamp  | 
 
Details
of sample of medical devices.
| 
   Date: ……………….  | 
  
   Medical Device Officer Seal or Stamp  | 
 
Form MD-37
[See Rule
77]
Receipt
for Sample of medical device(s) taken where fair price tendered thereof under
sub-section (1) of Section 23 of the Drugs and Cosmetics Act, 1940 is refused
To
………………….
Whereas I, this …………….. day
of ……………., have taken from the premises situated at ……. samples of medical
devices as specified below:
Details of samples—
And whereas I had offered
to you rupees ………….. as the fair price of the samples of aforesaid medical
devices taken:
And whereas, you have
refused to accept the fair price tendered thereof;
Now, therefore, I give you
this receipt as the fair price tendered for the samples of the medical devices
taken by me.
| 
   Date: ………………..  | 
  
   Medical Device Officer Seal or Stamp  | 
 
Form MD-38
[See sub-rule
(1) Rule 78]
Memorandum
to Medical Device Testing Officer
Serial No. of Memorandum
……………
From
To
The Medical Device Testing
Officer
The sample of medical
device described below is enclosed for test or evaluation under the provisions
of clause (i) of sub-section (4) of Section 23 of the Drugs and Cosmetics Act,
1940 (23 of 1940).
The sample of medical
device has been marked by me with following mark.
……………………………………………..
Details of sample of medical
device with name of medical device which it purports to contain—
……………………………………………..
| 
   Date: ………………..  | 
  
   Medical Device Officer Seal or Stamp  | 
 
Form MD-39
[See sub-rule
(1) of Rule 81]
Application
for grant of registration to Medical Device Testing Laboratory for carry out
Test or Evaluation of a medical device on behalf of manufacturer
| 
   1.  | 
  
   Name of Applicant:  | 
 
| 
   2.  | 
  
   Nature and constitution of applicant:  | 
 
| 
      | 
  
   (i.e. proprietorship, partnership
  including Limited Liability Partnership, private or public company, society,
  trust, other to be specified)  | 
 
| 
   3.  | 
  
   (i) Corporate/registered office
  address including telephone number, mobile number, fax number and e-mail id:  | 
 
| 
      | 
  
   (ii) Testing laboratory address
  including telephone number, mobile number, fax number and e-mail id:  | 
 
| 
      | 
  
   (iii) Address for correspondence:  | 
 
| 
      | 
  
   [corporate office/testing laboratory]  | 
 
| 
   4.  | 
  
   Details of medical device(s) to be
  tested or evaluated [Annexed].  | 
 
| 
   5.  | 
  
   Fee paid on ……………………… Rs ……………………..
  receipt/challan/transaction id …………………. .  | 
 
| 
   6.  | 
  
   I have enclosed the documents as
  specified in sub-rule (2) of Rule 82 of Medical Devices Rules, 2017.  | 
 
| 
   7.  | 
  
   I hereby state and undertake that:  | 
 
| 
      | 
  
   (i) the testing laboratory is ready
  for inspection or shall be ready for inspection on………………………….. in accordance
  with the requirements of Medical Devices Rules, 2017. (ii) I shall comply with the
  applicable provisions of the Drugs and Cosmetics Act, 1940 (23 of 1940), and
  the Medical Devices Rules, 2017.  | 
 
| 
   Place: …………………..  | 
  
   Signature  | 
 
| 
   Date: ……………………  | 
  
   (Name and designation)  | 
 
| 
      | 
  
   [To be signed digitally]  | 
 
Annexure:
| 
   Sl. No.  | 
  
   Generic name  | 
  
   Class of medical devices  | 
 
| 
      | 
  
      | 
  
      | 
 
Form MD-40
[See sub-rule
(3) of Rule 83]
Certificate
of registration to Medical Device Testing Laboratory for carry out Test or
Evaluation of a medical device on behalf of manufacturer
Registration No.: ……………….
| 
   1.  | 
  
   M/s …………………………………………………………… (Name of
  the firm) situated at ……………………………………………(full address with telephone and e-mail)
  has been registered as a Medical Device Testing Laboratory for carry out Test
  or Evaluation of a medical device on behalf of manufacturer under the Medical
  Devices Rules, 2017.  | 
 
| 
   2.  | 
  
   Details of medical device(s) to be
  tested or evaluated [Annexed].  | 
 
| 
   3.  | 
  
   This Registration is subject to the
  conditions as specified in the Drugs and Cosmetics Act, 1940 (23 of 1940) and
  the Medical Devices Rules, 2017.  | 
 
| 
   Place: …………….. Date: ………………  | 
  
   Central Licensing Authority [To be signed digitally]  | 
 
Annexure:
| 
   Sl. No.  | 
  
   Generic name  | 
  
   Class of medical devices  | 
 
| 
      | 
  
      | 
  
      | 
 
[82][Form MD-41
[See sub-rule
(2) of Rule 87A]
Application
for Grant of Registration Certificate to sell, stock, exhibit or offer for sale
or distribute a Medical Device including in Vitro Diagnostic Medical Device
(1)     Name of applicant:
(2)     Address of the premises to
be registered:
(3)     Contact details of
applicant including telephone number, mobile number, fax number and e-mail id:
(4)     Nature and constitution of
applicant: (i.e. proprietorship, partnership including Limited Liability
Partnership, private or public company, society, trust, other to be specified)
(5)     Name, qualification and
experience of competent person appointed:
(6)     Fee paid on …………………
Rs………………………… receipt/challan/transaction Id…………….
(7)     I have enclosed the
documents as specified in the sub-rule (3) of Rule 87A of the Medical Devices
Rules, 2017.
Place: ……………
Date: ……………
Name, designation &
signature of Director/Proprietor/Partner]
[83][Form MD-42
[See sub-rule
(4) of Rule 87A and sub-rule (1) of Rule 87C]
Registration
Certificate to Sell, Stock, Exhibit or Offer for Sale or Distribute a Medical
Device including in Vitro Diagnostic Medical Device
Registration No.: ……………….
(1)     M/s,
……………………………………………………………………………(Name of the firm) situated at
……………………………………………………………(full address with telephone and e-mail) has been
registered to sell, stock, exhibit or offer for sale or distribute a medical
device including in vitro diagnostic medical device under the Medical Devices
Rules, 2017.
(2)     Name and qualification of
competent person:
(3)     This registration is
subject to the conditions as specified in the Drugs and Cosmetics Act, 1940 (23
of 1940) and the Medical Devices Rules, 2017.
Place: ……………
Date: ……………
State Licensing Authority]
[84][Form MD-43
[See sub-rule
(8) of Rule 87B]
Form
in which the Inspection Book shall be maintained
(A)     The cover of the inspection
book shall contain the following particulars, namely—
(1)     The name and address of the
registration certificate holder ……………………
(2)     Registration certificate
number……………………………………….
(B)     (i) The pages of the
inspection book shall be serially numbered and duly stamped by the State
Licensing Authority*. The pages, other than the first and the last pages, shall
have the following particulars—
Name and designation of the
Medical Device Officer who inspected the premises:
Date of inspection
………………………………………………………
Observations of the Medical
Device Officer ……………………………
Signature of the Medical
Device Officer
(ii) The first and last
pages of the inspection book shall be endorsed by the State Licensing Authority
with the following words, namely—
Inspection book maintained
by M/s ……………………………situated at ………………… for Registration number ………………… in Form
MD-43………………… under the Medical Devices Rules, 2017.
State Licensing Authority
Notes:
(i)       Printed copy of the
inspection book may be obtained by the licencee from the Licensing Authority on
payment of fee as may be specified by the concerned Licensing Authority from
time-to-time.
(ii)      The inspection book shall
be maintained at the premises of the licencee.
(iii)     The original copy of
observations made by the Medical Device Officer shall be maintained in the
premises of the licencee and duplicate copy shall be sent to the State
Licensing Authority. The triplicate copy shall be taken as record by the
Medical Device Officer.]
[1] Ministry of Health and
Family Welfare (Deptt. of Health and Family Welfare), Noti. No. G.S.R. 78(E),
dated January 31, 2017, published in the Gazette of India, Extra., Part II,
Section 3(i), dated 31st January, 2017, pp. 143-250, No. 70
[2] Ins. by G.S.R. 30(E),
dt. 15-1-2019 (w.e.f. 15-1-2019).
[3] Ins. by G.S.R.
777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[4] Ins. by G.S.R.
777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[5] Ins. by G.S.R.
777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[6] Ins. by G.S.R.
787(E), dt. 16-10-2019 (w.e.f. 16-10-2019).
[7] Ins. by G.S.R.
102(E), dt. 11-2-2020 (w.e.f. 1-4-2020).
[8] Ins. by G.S.R. 19(E),
dt. 18-1-2022 (w.e.f. 18-1-2022).
[9] Subs. for “shall
mention the registration number” by G.S.R. 19(E), dt. 18-1-2022 (w.e.f.
18-1-2022).
[10] Ins. by G.S.R. 19(E),
dt. 18-1-2022 (w.e.f. 18-1-2022).
[11] Subs. for “shall
mention the registration number” by G.S.R. 19(E), dt. 18-1-2022 (w.e.f.
18-1-2022).
[12] Ins. by G.S.R.
777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[13] Ins. by G.S.R.
777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[14] Ins. by G.S.R.
777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[15] Ins. by G.S.R.
777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[16] Ins. by G.S.R.
777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[17] Ins. by G.S.R.
777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[18] Ins. by G.S.R.
777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[19] Ins. by G.S.R.
777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[20] Ins. by G.S.R.
777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[21] Ins. by G.S.R.
777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[22] Ins. by G.S.R.
777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[23] Ins. by G.S.R.
777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[24] Ins. by G.S.R.
777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[25] Ins. by G.S.R.
754(E), dt. 30-9-2022 (w.e.f. 30-9-2022).
[26] Subs. for “or the
United States of America” by G.S.R. 174(E), dt. 4-3-2022 (w.e.f. 4-3-2022).
[27] Ins. by G.S.R.
777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[28] Subs. by G.S.R.
918(E), dt. 31-12-2021 (w.e.f. 31-12-2021). Prior to substitution it read as:
“46. Unique
device identification of the medical device.—With effect from 1st day of
January, 2022, a medical device, approved for manufacture for sale or
distribution or import, shall bear unique device identification which shall
contain device identifier and production identifier.
Explanation.—For
the purposes of this rule,—
(i)
“device identifier” means a global trade item number;.
(ii)
“production identifier” means a serial number, lot or batch number, software as
a medical device version, manufacturing and or expiration date.”
[29] Ins. by G.S.R. 30(E),
dt. 15-1-2019 (w.e.f. 15-1-2019).
[30] Ins. by G.S.R.
754(E), dt. 30-9-2022 (w.e.f. 30-9-2022).
[31] Ins. by G.S.R.
754(E), dt. 30-9-2022 (w.e.f. 30-9-2022).
[32] Ins. by G.S.R.
754(E), dt. 30-9-2022 (w.e.f. 30-9-2022).
[33] Ins. by G.S.R.
754(E), dt. 30-9-2022 (w.e.f. 30-9-2022).
[34] Ins. by G.S.R.
754(E), dt. 30-9-2022 (w.e.f. 30-9-2022).
[35] Ins. by G.S.R.
754(E), dt. 30-9-2022 (w.e.f. 30-9-2022).
[36] Ins. by G.S.R.
318(E), dt. 18-4-2019 (w.e.f. 22-4-2019).
[37] Ins. by G.S.R.
754(E), dt. 30-9-2022 (w.e.f. 30-9-2022).
[38] Subs. for “Class A”
by G.S.R. 777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[39] Subs. for “Class A”
by G.S.R. 777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[40] Subs. for “Class A”
by G.S.R. 777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[41] Subs. for “Class A”
by G.S.R. 777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[42] Subs. for “Class A”
by G.S.R. 777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[43] Subs. for “Class A”
by G.S.R. 777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[44] Subs. for “Class A”
by G.S.R. 777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[45] Subs. for “Class A”
by G.S.R. 777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[46] Subs. for “Class A”
by G.S.R. 777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[47] Subs. for “distinct”
by G.S.R. 318(E), dt. 18-4-2019 (w.e.f. 22-4-2019).
[48] Ins. by G.S.R.
754(E), dt. 30-9-2022 (w.e.f. 30-9-2022).
[49] Ins. by G.S.R.
777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[50] Ins. by G.S.R.
754(E), dt. 30-9-2022 (w.e.f. 30-9-2022).
[51] Ins. by G.S.R.
777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[52] Subs. for “IFU” by
G.S.R. 30(E), dt. 15-1-2019 (w.e.f. 15-1-2019).
[53] Subs. for “IFU” by
G.S.R. 30(E), dt. 15-1-2019 (w.e.f. 15-1-2019).
[54] Subs. by G.S.R.
729(E), dated 1-8-2018 (w.e.f. 1-8-2018). Prior to substitution it read as:
“(h)
In case of in vitro diagnostic medical devices, a copy of performance
evaluation report issued by the central medical device testing laboratory or
medical device testing laboratory registered under sub-rule (3) of Rule 83.”
[55] Optional
[56] Subs. for
“instructions for use” by G.S.R. 30(E), dt. 15-1-2019 (w.e.f. 15-1-2019).
[57] Subs. by G.S.R.
30(E), dt. 15-1-2019 (w.e.f. 15-1-2019). Prior to substitution it read as:
“(b)
Instructions for use (Prescriber's manual);”
[58] Ins. by G.S.R.
450(E), dt. 15-6-2022 (w.e.f. 15-6-2022).
[59] Subs. by G.S.R.
30(E), dt. 15-1-2019 (w.e.f. 15-1-2019). Prior to substitution it read as:
“8.
Proposed Instruction for use and labels.”
[60] Subs. by G.S.R.
30(E), dt. 15-1-2019 (w.e.f. 15-1-2019). Prior to substitution it read as:
“5.
Proposed Instruction for use and labels.”
[61] Subs. by G.S.R.
224(E), dt. 18-3-2019 (w.e.f. 19-3-2019).
[62] Subs. by G.S.R.
224(E), dt. 18-3-2019 (w.e.f. 19-3-2019).
[63] Subs. for
“instructions for use” by G.S.R. 30(E), dt. 15-1-2019 (w.e.f. 15-1-2019).
[64] Subs. for
“instructions for use” by G.S.R. 30(E), dt. 15-1-2019 (w.e.f. 15-1-2019).
[65] Subs. by G.S.R.
30(E), dt. 15-1-2019 (w.e.f. 15-1-2019). Prior to substitution it read as:
“11.
Proposed Instructions for use and labels.”
[66] Ins. by G.S.R.
102(E), dt. 11-2-2020 (w.e.f. 1-4-2020).
[67] Ins. by G.S.R.
777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[68] Ins. by G.S.R.
777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[69] Ins. by G.S.R.
777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[70][70] Ins. by G.S.R.
777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[71] Ins. by G.S.R.
777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[72] Ins. by G.S.R.
777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[73] Ins. by G.S.R.
777(E), dt. 14-10-2022 (w.e.f. 14-10-2022).
[74] Subs. for “State
Licensing Authority” by G.S.R. 652(E), dt. 13-9-2019 (w.e.f. 13-9-2019).
[75] Delete whichever is not
applicable.
[76] Delete whichever is not
applicable.
[77] Ins. by G.S.R.
754(E), dt. 30-9-2022 (w.e.f. 30-9-2022).
[78] Ins. by G.S.R.
754(E), dt. 30-9-2022 (w.e.f. 30-9-2022).
[79] Ins. by G.S.R.
754(E), dt. 30-9-2022 (w.e.f. 30-9-2022).
[80] Ins. by G.S.R.
754(E), dt. 30-9-2022 (w.e.f. 30-9-2022).
[81] If opinion is required on
any other matter, the paragraph should be suitably amended.
[82] Ins. by G.S.R.
754(E), dt. 30-9-2022 (w.e.f. 30-9-2022).
[83] Ins. by G.S.R.
754(E), dt. 30-9-2022 (w.e.f. 30-9-2022).
[84] Ins. by G.S.R.
754(E), dt. 30-9-2022 (w.e.f. 30-9-2022).